Literature DB >> 18682674

Review of failed CT phantom image evaluations in 2005 and 2006 by the CT accreditation program of the Korean Institute for Accreditation of Medical Image.

Hye Jung Park1, Seung Eun Jung, Young Joon Lee, Woo Il Cho, Kyung Hyun Do, Seung Hoon Kim, Dong Gyu Na.   

Abstract

OBJECTIVE: The CT accreditation program was established in 2004 by the Korean Institute for Accreditation of Medical Image (KIAMI) to confirm that there was proper quality assurance of computed tomography (CT) images. We reviewed all the failed CT phantom image evaluations performed in 2005 and 2006.
MATERIALS AND METHODS: We analyzed 604 failed CT phantom image evaluations according to the type of evaluation, the size of the medical institution, the parameters of the phantom image testing and the manufacturing date of the CT scanners.
RESULTS: The failure rates were 10.5% and 21.6% in 2005 and 2006, respectively. Spatial resolution was the most frequently failed parameter for the CT phantom image evaluations in both years (50.5% and 49%, respectively). The proportion of cases with artifacts increased in 2006 (from 4.5% to 37.8%). The failed cases in terms of image uniformity and the CT number of water decreased in 2006. The failure rate in general hospitals was lower than at other sites. In 2006, the proportion of CT scanners manufactured before 1995 decreased (from 12.9% to 9.3%).
CONCLUSION: The continued progress in the CT accreditation program may achieve improved image quality and thereby improve the national health of Korea.

Entities:  

Mesh:

Year:  2008        PMID: 18682674      PMCID: PMC2627278          DOI: 10.3348/kjr.2008.9.4.354

Source DB:  PubMed          Journal:  Korean J Radiol        ISSN: 1229-6929            Impact factor:   3.500


  1 in total

1.  The phantom portion of the American College of Radiology (ACR) computed tomography (CT) accreditation program: practical tips, artifact examples, and pitfalls to avoid.

Authors:  Cynthia H McCollough; Michael R Bruesewitz; Michael F McNitt-Gray; Krista Bush; Thomas Ruckdeschel; J Thomas Payne; James A Brink; Robert K Zeman
Journal:  Med Phys       Date:  2004-09       Impact factor: 4.071

  1 in total
  3 in total

1.  Image quality improvement after implementation of a CT accreditation program.

Authors:  You Sung Kim; Seung Eun Jung; Byung Gil Choi; Yu Ri Shin; Seong Su Hwang; Young Mi Ku; Yeon Soo Lim; Jae Mun Lee
Journal:  Korean J Radiol       Date:  2010-08-27       Impact factor: 3.500

2.  The relationship between subjective and objective parameters in CT phantom image evaluation.

Authors:  Hye Jung Park; Seung Eun Jung; Young Joon Lee; Woo Il Cho; Kyung Hyun Do; Seung Hyup Kim; Ki Hwang Kim
Journal:  Korean J Radiol       Date:  2009-08-25       Impact factor: 3.500

3.  Intra- and interobserver reliability of gray scale/dynamic range evaluation of ultrasonography using a standardized phantom.

Authors:  Song Lee; Joon-Il Choi; Michael Yong Park; Dong Myung Yeo; Jae Young Byun; Seung Eun Jung; Sung Eun Rha; Soon Nam Oh; Young Joon Lee
Journal:  Ultrasonography       Date:  2014-02-26
  3 in total