Accuracy of the Precision® point-of-care ketone test examined by liquid chromatography tandem-mass spectrometry (LC-MS/MS) in the same fingerstick sample.

Marcel J W Janssen1, Ben H E Hendrickx, Carin D Habets-van der Poel, Joop P W van den Bergh, Anton A M Haagen, Jaap A Bakker.   

Abstract

BACKGROUND: The Precision(®) (Abbott Diabetes Care) point-of-care biosensor test strips are widely used by patients with diabetes and clinical laboratories for measurement of plasma β-hydroxybutyrate (β-HB) concentrations in capillary blood samples obtained by fingerstick. In the literature, this procedure has been validated only against the enzymatic determination of β-HB in venous plasma, i.e., the method to which the Precision(®) has been calibrated.
METHODS: In this study, the Precision(®) Xceed was compared to a methodologically different and superior procedure: determination of β-HB by liquid chromatography tandem-mass spectrometry (LC-MS/MS) in capillary blood spots. Blood spots were obtained from the same fingerstick sample from out of which Precision(®) measurements were performed. Linearity was tested by adding varying amounts of standard to an EDTA venous whole blood matrix.
RESULTS: The Precision(®) was in good agreement with LC-MS/MS within the measuring range of 0.0-6.0 mmol/L (Passing and Bablok regression: slope=1.20 and no significant intercept, R=0.97, n=59). Surprisingly, the Precision(®) showed non-linearity and full saturation at concentrations above 6.0 mmol/L, which were confirmed by a standard addition experiment. Results obtained at the saturation level varied between 3.0 and 6.5 mmol/L.
CONCLUSIONS: The Precision(®) β-HB test strips demonstrate good comparison with LC-MS/MS. Inter-individual variation around the saturation level, however, is large. Therefore, we advise reporting readings above 3.0 as >3.0 mmol/L. The test is valid for use in the clinically relevant range of 0.0-3.0 mmol/L.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20731618     DOI: 10.1515/CCLM.2010.351

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

1.  Evaluation of Nova StatStrip and FreeStyle Precision Pro blood ketone tests using 3-hydroxybutyrate doped samples.

Authors:  Johanna Helmersson-Karlqvist; Kristina Höög Hammarström; Kicki Palmberg; Carolina Backman-Johansson
Journal:  J Clin Lab Anal       Date:  2019-02-27       Impact factor: 2.352

2.  Comparative performance assessment of point-of-care testing devices for measuring glucose and ketones at the patient bedside.

Authors:  Ferruccio Ceriotti; Ewa Kaczmarek; Elena Guerra; Fabrizio Mastrantonio; Fausto Lucarelli; Francesco Valgimigli; Andrea Mosca
Journal:  J Diabetes Sci Technol       Date:  2014-12-17

3.  An Intermittent Fasting Mimicking Nutrition Bar Extends Physiologic Ketosis in Time Restricted Eating: A Randomized, Controlled, Parallel-Arm Study.

Authors:  Angie W Huang; Min Wei; Sara Caputo; Melissa L Wilson; Joseph Antoun; William C Hsu
Journal:  Nutrients       Date:  2021-04-30       Impact factor: 5.717

4.  Reliability and diagnostic performance of a new blood ketone and glucose meter in humans.

Authors:  Andrew Ray Moore; Angelia Maleah Holland-Winkler; Jenna Kate Ansley; Eric Deiondre Hunter Boone; Megahn Kimberanne O'Reilly Schulte
Journal:  J Int Soc Sports Nutr       Date:  2021-01-07       Impact factor: 5.150

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.