Literature DB >> 20717078

Inverse planning optimization for hybrid prostate permanent-seed implant brachytherapy plans using two source strengths.

J Adam M Cunha1, Barby Pickett, Jean Pouliot.   

Abstract

The purpose is to demonstrate the ability to generate clinically acceptable prostate permanent seed implant plans using two seed types which are identical except for their activity. The IPSA inverse planning algorithms were modified to include multiple dose matrices for the calculation of dose from different sources, and a selection algorithm was implemented to allow for the swapping of source type at any given source position. Five previously treated patients with a range of prostate volumes from 20-48 cm3 were re-optimized under two hybrid scenarios: (1) using 0.32 and 0.51 mGy m2 / h 125I, and (2) using 0.64 and 0.76 mGy m2 / h 125I. Isodose lines were generated and dosimetric indices , V150Prostate, D90Prostate, V150Urethra, V125Urethra, V120Urethra,V100Urethra, and D10Urethra were calculated. The algorithm allows for the generation of single-isotope, multi-activity hybrid brachytherapy plans. By dealing with only one radionuclide, but of different activity, the biology is unchanged from a standard plan. All V100Prostate were within 2.3 percentage points for every plan and always above the clinically desirable 95%. All V150Urethra were identically zero, and V120Urethra is always below the clinically acceptable value of 1.0 cm3. Clinical optimization times for the hybrid plans are still under one minute, for most cases. It is possible to generate clinically advantageous brachytherapy plans (i.e. obtain the same quality dose distribution as a standard single-activity plan) while incorporating leftover seeds from a previous patient treatment. This method will allow a clinic to continue to provide excellent patient care, but at a reduced cost. Multi-activity hybrid plans were equal in quality (as measured by the standard dosimetric indices) to plans with seeds of a single activity. Despite the expanded search space, optimization times for these studies were still under two minutes on a modern day laptop and can be reduced to below one minute in a clinical setting. With the typical cost of a set of PPI seeds on the order of thousands of dollars, it is possible to reduce the cost of brachytherapy treatments by allowing for easier use of seeds left over from a previous patient or unused due to a cancelled treatment.

Entities:  

Mesh:

Year:  2010        PMID: 20717078      PMCID: PMC5720434          DOI: 10.1120/jacmp.v11i3.3096

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


I. INTRODUCTION

Current permanent prostate implant brachytherapy (PPI) dose plans use one type of radioactive seed per treatment plan, all with the same activity. Radioactive seeds must be ordered from a vendor and used before the radioactivity dissipates. The half‐life of one of the more common radionuclides, , is approximately 60 days; therefore, if the seeds are not used within one week, the activity will have decreased by approximately 10%. For seeds like and , with half lives of 17 and 10 days respectively, the decrease is even greater (Table 1). Thus, if a treatment is missed, seeds — and therefore money — are wasted. These seeds are not entirely unusable and may be used for a different patient by generating a plan using the partially‐decayed activity ; however, this can only be done if there are enough seeds for an entire plan.
Table 1

Half‐lives for commonly‐used PPI seeds: after two weeks.

Seed Type Half Life Activity after
7 days 14 days
Iodine‐12559.9 days92%85%
Palladium‐10317.0 days75%57%
Cesium‐1319.7 days62%38%
Half‐lives for commonly‐used PPI seeds: after two weeks. In addition, it has been reported that it is important to account for differences in the seeds supplied by different manufacturers when generating treatment plans and calculating dosimetry. Identical plans using the same radionuclide but from different vendors will have different dosimetry due to different dose rate constants, anisotropy factors, and radial dose functions of each vendor's seeds.( , , ) There has been recent interest in generating dose plans with two different radionuclides. In their 2002 paper on the feasibility of seeds for PPI, Glasgow et al.( ) proposed that combinations of and seeds can provide adequate coverage of the prostate and sparing of the organs at risk. In this case, the argument is that and have similar half lives (73.83 days verses 59.40 days) and similar absorbed doses delivered in the permanent implant. Therefore, the radiobiological effects of using two different radionuclides is negligible for cases which used up to . In 2007, Chaswal et al.( ) expanded on the forward planning‐based work of Glasgow and used a Greedy Heuristic algorithm to optimize the treatment plans. They showed that there can be dosimetric and trauma‐reducing benefits to using plans with a combination of and . These studies are promising in that they show a clinical benefit can be derived from including more than one radionuclide in a given brachytherapy plan. However, the results are dependent on knowing and understanding the biology of the superposition of two different radiation delivery mechanisms. There has been progress made recently in our understanding of the biological responses to different dose delivery characteristics of different radionuclides. In fact, some recent studies have shown no significant difference between biochemical failure and control as a function of biological equivalent dose (BED), equivalent uniform dose (EUD), or tumor control probability (TCP) when comparing plans using and .( ) However, even setting aside the uncertainty in CT‐reconstructed target volumes, there is a lack of consensus about the actual value of the biological parameters that govern the definition of the BED, EUD, and TCP, with values of α/β differing by more than a factor of 2.( ) The uncertainty in the models that predict the biological effect of the specific characteristics naturally propagates to the dosimetry of brachytherapy plans that incorporate multiple radionuclides. However, assuming the uncertainty in the biology is acceptable, there does appear to be a benefit to using multiple types of seeds (each with a different dosimetry) in one brachytherapy plan.( , ) In this work, we aim to avoid the biology‐based uncertainties attached to the use of two different radionuclides. It would be beneficial to be able to generate clinically‐acceptable hybrid plans that incorporate seeds of the same radionuclide, but with different characteristics — be they different activity or slightly different dosimetric geometry. In contrast to plans using multiple radionuclides, single radionuclide hybrid plans are free from the biological uncertainties associated with the efficacy combining characteristically different radiation. We present an algorithm that allows for the generation of single‐isotope, multi‐activity hybrid brachytherapy plans. By dealing with only one radionuclide, but of different activity, it is possible to generate clinically advantageous brachytherapy plans: i.e. obtain the same quality dose distribution as a standard single‐activity plan while incorporating left over seeds from a previous patient treatment. This method will allow a clinic to continue to provide excellent patient care, but at a reduced cost.

II. MATERIALS AND METHODS

A. Code development rational

The algorithms developed for this work augment the IPSA inverse planning algorithms.( , , , ) While the core optimization engine remains unchanged, the code was modified in three ways: (1) more than one radionuclide source can be specified using the TG‐43 formalism;( ) (2) the system for evaluating the dose delivered to the target and surrounding organs was expanded to incorporate two separate radionuclide dose profiles; and (3) the optimization engine can switch the seed type at each source position at specific moments during the iteration process. As in the original algorithm, to determine the dose, , delivered to unit volume (voxel) j of an organ, the contributions from the source position, , are summed: is the number of source positions. The total dose, D, to an organ is the sum of the dose at each voxel within that organ, is the number of voxels in the organ. However, in the new algorithm, the dose matrix () used depends on which seed is present at position i. During the initialization phase of the optimization process, the algorithm computes a separate dose matrix for each seed type. Then an initial configuration of seeds is placed in the target. This configuration is random, but the user has the ability to specify a desired number of each seed type. The user also has the ability to permanently set the seed type at any source position. The algorithm then proceeds to the iteration phase, during which the search space is probed by placing or removing seeds of either type at randomly selected source positions. The new configuration is then (1) evaluated to determine the value of the objective function , (2) compared to the best configuration yet attained, and (3) kept or rejected based on the result of No. 2. In order to maximize clinical relevance, the optimization parameters were chosen from the class solution developed in Lessard et al.( ) and Pouliot.( ) This class solution was shown to mimic an experienced dosimetrist by consistently producing dose plans equivalent (as measured by common dosimetric indices) to those created by a dosimetrist in our clinic. This allows for better isolation of the variables which control the use of partial‐activity seeds in the optimization. Two limits on the number of needles allowed per plan are included: a soft penalty and an upper limit. The soft penalty is a weight multiplied by the total number of needles and then added into the objective function of the optimization. The upper limit restricts the number of needles allowed in one implant. Throughout the studies in this paper, the upper limit is set to 30. This is the limit our clinic imposes, since our clinical experience has led to the conventional wisdom that, as the number of needles in one implant approaches 30, the complications due to edema, erectile dysfunction and other trauma begin to have unacceptably deleterious effects on the post‐treatment quality of life of the patient. On average our clinic uses 25 needles per treatment plan for a prostate gland.

B. Dosimetric analysis

The quality of the plans generated in this analysis will be graded using the standard dosimetric indices developed by the American Association of Physicists in Medicine (AAPM), the American Brachytherapy Society (ABS), and our clinic. For brachytherapy treatment of prostate cancer, Stock et al.( ) have demonstrated that the minimum dose delivered to at least 90% of the gland () had a critical impact on the subsequent risk of prostate‐specific antigen‐diagnosed recurrence. The American Brachytherapy Society has published general guidelines for different anatomical sites( ) that consist of a set of dose limit specifications. Our clinic incorporates these recommendations and imposes even stricter limits for the target coverage and urethra dose,( ) which have yielded reproducible results: mature five‐year biochemical control rates of 96% (median follow‐up 63 months) reported for 118 consecutive patients.( ) Plan quality was evaluated by the UCSF clinical physicist with experience generating over 1500 plans. For the purposes of this study, the single‐activity plans were considered the control sample against which the dosimetry of the multiple‐activity plans was judged. The validity of the standard (single‐activity) plans was examined and established in previous work.( )

B.1 Target dose coverage

The target coverage was compared across plans by examining volume of the prostate receiving at least the prescription dose (), at least 150% of the prescription dose (), and the minimum dose that covers 90% of the prostate (). In general, we strive to achieve and . A above 95% ensures complete coverage of the target, but will encourage a value of 10%–20% above the prescription dose. Because this is a not a post‐implant dosimetry analysis, but rather a pre‐implant dosimetry analysis, it is important to see the higher than the prescription dose by approximately this amount,( , ) since the dosimetry generally sees a decrease in post‐implant with respect to the pre‐implant .

B.2 Urethral dosimetry

One of the most common side effects of prostate brachytherapy (urethral stricture) is caused by excessive dose delivered to the urethra. The urethral volume receiving at least 100%, 120%, 125%, and 150% of the presection dose (, , , and , respectively) as well as the minimum dose delivered to 10% of the urethra (), can be used to monitor the dose delivered to the urethra. To adequately protect the urethra, we strive to obtain and .

C. Procedure

Five previously treated patients with a range of prostate volumes from 20 to were chosen and reoptimized using the hybrid‐activity optimization. Different clinics use different seed activities — 0.3–0.5 mCi and up to 0.7 mCi are common.( ) Since our clinic uses 0.39 mCi, we performed this work using two different sets of seed activity. Four studies were organized into two scenarios. Of course the work presented here should not be considered indicative of the clinical work done at any institution; since many factors go into generating plans, these studies were kept as general as possible. In Scenario I, we generate hybrid plans using 0.32 and seeds . In Scenario II, we generate hybrid plans using 0.64 and seeds . Table 2 outlines the parameters of the studies performed.
Table 2

A summary of the studies performed.

Study Air Kerma Rate 125I(U) Per‐needle Penalty Results Table
Scenario I; Study 10.32 & 0.51On Table 3
Scenario I; Study 20.32 & 0.51Off Table 4
Scenario II; Study 10.64 & 0.76On Table 5
Scenario II; Study 20.64 & 0.76Off Table 6

For each study five plan types were generated

A summary of the studies performed. For each study five plan types were generated Scenario I; Study 1. Data showing five separate plans generated for each patient case. Plans were generated with: (1) only one seed activity available to the optimization, 0.51 U ; (2) a requested 30 seeds of 0.32 U activity plus any number of seeds of 0.51 U ; (3) a requested 60 seeds of 0.32 U plus any number of seeds of 0.51 U ; (4) only 0.32 U seeds available to the optimization; (5) an unrestricted hybrid mix of 0.32 and 0.51 U . Per‐needle penalty is on. Scenario I; Study 2. Data showing five separate plans generated for each patient case. Plans were generated with: (1) only one seed activity available to the optimization, 0.51 U ; (2) a requested 30 seeds of 0.32 U activity plus any number of seeds of 0.51 U ; (3) a requested 60 seeds of 0.32 U plus any number of seeds of 0.51 U ; (4) only one seed activity available, 0.32 U ; (5) unrestricted hybrid mix of 0.32 U and 0.51 U . Per‐needle penalty is off. Scenario II; Study 1. Data showing five separate plans generated for each patient case. Plans were generated with: (1) only one seed activity available to the optimization, 0.76 U ; (2) a requested 30 seeds of 0.64 U activity plus any number of seeds of 0.60 mCi ; (3) a requested 60 seeds of 0.64 U plus any number of seeds of 0.76 U ; (4) only 0.64 U seeds available to the optimization; (5) an unrestricted hybrid mix of 0.64 and 0.76 U . Per‐needle penalty is on. Scenario II; Study 2. Study of 0.635 U vs. 0.6 mCi and hybrid plans. Data showing five separate plans generated for each patient case – Prostate dosimetric indices. Plans were generated with: (1) only one seed activity available to the optimization, 0.76 U ; (2) a requested 30 seeds of 0.64 U activity plus any number of seeds of 0.60 mCi ; (3) a requested 60 seeds of 0.64 U plus any number of seeds of 0.76 U ; (4) only 0.64 U seeds available to the optimization; (5) an unrestricted hybrid mix of 0.64 and 0.76 U . Per‐needle penalty off. For each study, five plan types were generated with a prescription dose of 144 Gy: Uni‐activity–full activity: Only one seed type (the full activity) is available to theoptimization 30 partial activity: The optimization is requested to use 30 of the partial activity seeds plus any number of the full activity seeds 60 partial activity: The optimization is requested to use 60 of the partial activity seeds plus any number of the full activity seeds Uni‐activity–partial activity: Only one seed (the partial activity) available to the optimization Unrestricted mix (pure hybrid): Allow the optimization to employ any number of each seed type The soft needle penalty encourages the optimization to minimize the number of needles in addition to optimizing the dose distribution. For Study 1 in Scenario I and II, this needle penalty is set to the value used in our clinic. For Study 2 in Scenario I and II, this was removed and the optimization was run with no per‐needle penalty.

III. RESULTS

The dosimetric results for Scenario I are presented in Tables 3 and 4, and a set of isodose lines are shown in Fig. 1. The same are presented for Scenario II in Tables 5 & 6 and Fig. 2. The number of seeds of each activity is listed along with the standard dosimetric indices for the prostate (, , and ) and for the urethra (, , , and ). is identically zero for all cases and is, therefore, not listed in the Tables.
Table 3

Scenario I; Study 1. Data showing five separate plans generated for each patient case. Plans were generated with: (1) only one seed activity available to the optimization, 0.51 U ; (2) a requested 30 seeds of 0.32 U activity plus any number of seeds of 0.51 U ; (3) a requested 60 seeds of 0.32 U plus any number of seeds of 0.51 U ; (4) only 0.32 U seeds available to the optimization; (5) an unrestricted hybrid mix of 0.32 and 0.51 U . Per‐needle penalty is on.

Target Volume (cm3) Plan Type # of Needles Total Seeds Low Act. #(%) V100Prostate (%) V150Prostate (%) D90Prostate (Gy) D100Urethra (cm3) V120Urethra (cm3) V125Urethra (cm3) D10Urethra (Gy)
48Only 0.51 U26850(0)96.964.01780.60.40.1184
4830 0.32 U299428(30)99.256.41740.60.10.1184
4860 0.32 U2910557(54)98.259.61780.60.20.1186
48Only 0.32 U30130130(100)98.657.21780.60.30.1186
48Pure hybrid299424(26)98.257.31800.60.10.0177
47Only 0.51 U23890(0)99.461.41790.80.10.0174
4730 0.32 U2610030(30)99.659.51730.80.10.0171
4760 0.32 U2610958(53)99.259.31750.80.10.0175
47Only 0.32 U30137137(100)98.861.91750.90.10.0175
47Pure hybrid2510032(32)97.660.21730.80.10.0173
36Only 0.51 U28730(0)96.663.91730.50.10.0177
3630 0.32 U308530(35)97.363.71810.50.10.0177
3660 0.32 U309458(62)98.262.01760.50.00.0173
36Only 0.32 U39111111(100)97.659.81760.60.00.0173
36Pure hybrid288225(30)97.161.61780.50.10.0178
29Only 0.51 U22630(0)98.262.71780.90.20.0178
2930 0.32 U257429(39)98.061.51800.80.20.0177
2960 0.32 U308559(69)97.961.61760.80.20.0177
29Only 0.32 U309797(100)97.456.71760.80.00.0177
29Pure hybrid257122(31)97.858.71770.90.00.0174
20Only 0.51 U18470(0)96.964.71750.40.20.2189
2030 0.32 U185729(51)96.662.31740.40.20.1183
2060 0.32 U226657(86)96.764.11730.40.20.1188
20Only 0.32 U227070(100)97.362.71730.40.20.1188
20Pure hybrid195112(24)96.760.41750.40.20.1184
Table 4

Scenario I; Study 2. Data showing five separate plans generated for each patient case. Plans were generated with: (1) only one seed activity available to the optimization, 0.51 U ; (2) a requested 30 seeds of 0.32 U activity plus any number of seeds of 0.51 U ; (3) a requested 60 seeds of 0.32 U plus any number of seeds of 0.51 U ; (4) only one seed activity available, 0.32 U ; (5) unrestricted hybrid mix of 0.32 U and 0.51 U . Per‐needle penalty is off.

Target Volume (cm3) Plan Type # of Needles Total Seeds Low Act. #(%) V100Prostate (%) V150Prostate (%) D90Prostate (Gy) D100Urethra (cm3) V120Urethra (cm3) V125Urethra (cm3) D10Urethra (Gy)
48Only 0.51 U30860(0)97.963.91790.60.30.1183
4830 0.32 U309730(31)99.056.21810.60.10.0176
4860 0.32 U3010558(55)99.058.61790.60.20.1181
48Only 0.32 U30129129(100)99.154.11790.60.20.0181
48Pure hybrid309629(30)99.356.51800.60.10.0178
47Only 0.51 U29880(0)99.462.41780.80.10.0176
4730 0.32 U3010030(30)99.660.51730.80.00.0173
4760 0.32 U3011060(55)98.558.81710.80.10.0175
47Only 0.32 U30136136(100)99.361.01710.90.10.0175
47Pure hybrid2910029(29)99.057.11760.80.10.0174
36Only 0.51 U30740(0)96.963.51730.60.00.0174
3630 0.32 U308529(34)97.561.41790.50.10.0176
3660 0.32 U309559(62)98.261.21760.60.00.0174
36Only 0.32 U30111111(100)98.259.31760.60.10.0174
36Pure hybrid308531(36)96.760.81780.50.10.0175
29Only 0.51 U29640(0)97.762.71780.80.10.0176
2930 0.32 U307429(39)97.959.61790.80.10.0173
2960 0.32 U308560(71)98.962.31750.90.10.0173
29Only 0.32 U309696(100)97.759.01750.90.00.0173
29Pure hybrid297531(41)97.959.41760.80.00.0172
20Only 0.51 U19470(0)97.666.61820.40.20.2194
2030 0.32 U226830(52)97.663.71790.40.20.1183
2060 0.32 U246657(86)96.964.71750.40.20.2189
20Only 0.32 U257171(100)96.963.81750.40.20.1189
20Pure hybrid215316(30)96.462.11770.40.20.0180
Figure 1

Scenario I, Study 1: Isodose lines for one slice of the prostrate. Isodose lines (outside in) are 50% (purple), 100% (red), 125% (orange), 150% (yellow), 150% (black).

Table 5

Scenario II; Study 1. Data showing five separate plans generated for each patient case. Plans were generated with: (1) only one seed activity available to the optimization, 0.76 U ; (2) a requested 30 seeds of 0.64 U activity plus any number of seeds of 0.60 mCi ; (3) a requested 60 seeds of 0.64 U plus any number of seeds of 0.76 U ; (4) only 0.64 U seeds available to the optimization; (5) an unrestricted hybrid mix of 0.64 and 0.76 U . Per‐needle penalty is on.

Target Volume (cm3) Plan Type # of Needles Total Seeds Low Act. #(%) V100Prostate (%) V150Prostate (%) D90Prostate (Gy) D100Urethra (cm3) V120Urethra (cm3) V125Urethra (cm3) D10Urethra (Gy)
48Only 0.76 U22580(0)98.650.31740.60.20.0179
4830 0.64 U236229(47)99.350.91720.60.10.1181
4860 0.64 U236860(88)98.453.81780.70.20.1183
48Only 0.64 U256969(100)99.657.01780.70.10.1183
48Pure hybrid236333(52)99.156.01770.60.10.1181
47Only 0.76 U28610(0)98.759.91690.90.10.0173
4730 0.64 U306530(46)98.056.81750.80.10.0175
4760 0.64 U297059(84)98.562.41750.80.10.0176
47Only 0.64 U287272(100)96.956.91750.80.10.0176
47Pure hybrid286533(51)98.257.11710.80.00.0172
36Only 0.76 U28510(0)96.957.71760.50.20.2186
3630 0.64 U295529(53)95.756.41780.50.10.0178
3660 0.64 U286059(98)96.160.11760.50.10.0175
36Only 0.64 U286060(100)97.058.41760.50.20.0175
36Pure hybrid305528(51)96.358.01800.50.20.0179
29Only 0.76 U26430(0)97.357.61770.80.20.0177
2930 0.64 U254829(60)96.553.91740.70.00.0171
2960 0.64 U265252(100)97.658.51790.70.10.0176
29Only 0.64 U285252(100)97.057.11790.90.00.0176
29Pure hybrid274719(40)97.953.31790.80.00.0172
20Only 0.76 U17320(0)94.762.01720.40.20.1185
2030 0.64 U183729(78)97.358.81730.40.20.0181
2060 0.64 U183939(100)94.762.01720.40.20.1185
20Only 0.64 U183939(100)93.759.61720.40.20.1185
20Pure hybrid183517(49)95.156.11710.40.20.0181
Table 6

Scenario II; Study 2. Study of 0.635 U vs. 0.6 mCi and hybrid plans. Data showing five separate plans generated for each patient case – Prostate dosimetric indices. Plans were generated with: (1) only one seed activity available to the optimization, 0.76 U ; (2) a requested 30 seeds of 0.64 U activity plus any number of seeds of 0.60 mCi ; (3) a requested 60 seeds of 0.64 U plus any number of seeds of 0.76 U ; (4) only 0.64 U seeds available to the optimization; (5) an unrestricted hybrid mix of 0.64 and 0.76 U . Per‐needle penalty off.

Target Volume (cm3) Plan Type # of Needles Total Seeds Low Act. #(%) V100Prostate (%) V150Prostate (%) D90Prostate (Gy) D100Urethra (cm3) V120Urethra (cm3) V125Urethra (cm3) D10Urethra (Gy)
48Only 0.76 U30570(0)99.352.61770.60.10.0181
4830 0.64 U306229(47)99.346.21730.60.10.1181
4860 0.64 U306860(88)99.056.71800.60.10.0177
48Only 0.64 U306969(100)99.350.11800.60.10.1177
48Mix306436(56)99.547.61740.60.10.1182
47Only 0.76 U28620(0)98.058.71740.80.20.1176
4730 0.64 U306630(45)98.157.21650.80.10.0175
4760 0.64 U297059(84)98.055.51670.90.10.0175
47Only 0.64 U287373(100)98.452.81670.80.10.0175
47Mix286529(45)97.955.71710.90.10.0173
36Only 0.76 U28510(0)96.257.31760.50.10.0180
3630 0.64 U295529(53)97.357.41750.50.20.1184
3660 0.64 U286058(97)96.461.71760.50.10.0180
36Only 0.64 U286060(100)96.556.51760.50.20.1180
36Mix305526(47)96.759.81800.50.10.0180
29Only 0.76 U26440(0)96.056.81700.40.20.0177
2930 0.64 U254929(59)96.360.71700.40.20.1183
2960 0.64 U265252(100)96.056.81700.40.20.0177
29Only 0.64 U285353(100)98.364.61700.40.20.1177
29Mix274720(43)96.760.91680.40.20.1182
20Only 0.76 U17330(0)98.357.41810.80.10.0176
2030 0.64 U183830(79)96.758.51810.80.00.0173
2060 0.64 U183838(100)98.158.71760.80.00.0174
20Only 0.64 U183838(100)97.354.21760.80.10.0174
20Mix183618(50)97.060.01810.80.10.0176
Figure 2

Scenario II, Study 1: Isodose lines for one slice of the prostate. Isodose lines (outside in) are 50% (purple), 100% (red), 125% (orange), 150% (yellow), 150% (black).

Scenario I, Study 1: Isodose lines for one slice of the prostrate. Isodose lines (outside in) are 50% (purple), 100% (red), 125% (orange), 150% (yellow), 150% (black). Scenario II, Study 1: Isodose lines for one slice of the prostate. Isodose lines (outside in) are 50% (purple), 100% (red), 125% (orange), 150% (yellow), 150% (black).

A. Activities 0.32 U and 0.51 U

Note that for any one case in Scenario I, Study 1 (Table 3), the target volume receiving 100% of the prescribed dose () is within 2.3 percentage points for every plan and always above the clinically desirable 95%. The for all plans hovers around 120% of the prescription dose. This is to be expected when evaluating the preplan dosimetry as opposed to the post‐plan dosimetry.( , ) Also note that the is always below the clinically acceptable value of . Study 2 under this scenario shows the effect of removing the per‐needle penalty compared to Study 1, which preserves it. Note that the effect of removing the needle penalty is to increase the number of needles and obtain a slight increase in target coverage. Comparing Table 3 with Table 4 (and Table 5 with Table 6 in Scenario II) shows the effect of the per‐needle penalty. In general, this penalty allows for a decrease in the number of needles needed independent of the hybrid status of the plan. It is interesting to note that, even in the case where the per‐needle penalty is turned on (Tables 3 and 5), the plans generated with an unrestricted mix of seeds still preferred using some partial‐activity seeds. For the studies with the per‐needle penalty turned on, the number of needles recommended for each implant is consistent with the number of needles routinely used for actual implants performed at our clinic.

B. Activities 0.64 U and 0.76 U

The results are similar in Scenario II for both Study 1 (Table 5) and Study 2 (Table 6) across all plans. It is clear from these results that it is possible to achieve clinically acceptable dose distributions with hybrid plans that incorporate different‐activity seeds. Across all four studies of each of the five patients, there is no evidence that the prostate size (which ranges from 20 to 48 cc) has an effect on the dosimetry of the hybrid plans. Studies were performed on the geometric distribution of each seed type. The density of seeds as a function of distance from the center of mass of the prostate was examined. The prostate was divided in octants with the origin placed at the center of the prostate , and the density of seeds in each octant was evaluated. No correlation was found between seed location and seed type. A final study was performed in order to determine whether the optimized configuration of seed types is independent of the initial configuration. The initial configuration of seed types was set to be all of one activity or the other (as opposed to randomly chosen for each seed position, as was done in the main studies). The plans were then optimized requesting an unrestricted mix of the two seed types. The final configuration was found to have no correlation to the initial configuration of seed types.

C. CPU time

Table 7 shows the time needed to process the optimization in Scenario I Study 1 (all studies had similar results). The baseline IPSA( , , , ) optimization for a single‐activity varies from 51–155 s. The hybrid optimization incorporates an increase in the degrees of freedom that must be probed, so an increase in optimization time is expected. The increase in optimization time varies from 17%–35%; however, this translates into only a 16–37 s increase in real time. Thus, the optimization time is still on the order of two minutes for all cases. Our clinical protocol for the baseline optimization uses a factor of five fewer iterations than used in this study. The optimizations were stable after a two‐fold increase in the number of iterations over the clinical baseline; so the times listed here should be considered an upper limit which can be reduced without any clinically noticeable loss in plan quality by a factor of 2.
Table 7

Scenario I; Study 1. CPU time required for optimization using a MacBook Pro running Mac OS 10.5.5 with a 2.33 GHz Intel Core 2 Duo processor and 3 GB RAM. On average there is a 25% increase in CPU time needed for the hybrid plan optimization; however, in real time this only amounts to at most about an 26‐second increase in optimization time. This can be further reduced by a factor of two when implemented on a clinical system where fewer iterations are needed than in this proof of hypothesis study.

Plan Type CPU Time for Each Optimization
48cm3 47cm3 36cm3 29cm3 20cm3
Only 0.51 U90 s118 s100 s85 s51 s
30 0.32 U104 s151 s135 s110 s68 s
60 0.32 U106 s155 s131 s112 s67 s
Only 0.32 U97 s125 s104 s90 s51 s
Pure Hybrid100 s144 s122 s102 s61 s
Max Increase (s)16 s37 s35 s27 s17 s
Max Increase (%)17%24%35%24%25%
Scenario I; Study 1. CPU time required for optimization using a MacBook Pro running Mac OS 10.5.5 with a 2.33 GHz Intel Core 2 Duo processor and 3 GB RAM. On average there is a 25% increase in CPU time needed for the hybrid plan optimization; however, in real time this only amounts to at most about an 26‐second increase in optimization time. This can be further reduced by a factor of two when implemented on a clinical system where fewer iterations are needed than in this proof of hypothesis study.

IV. DISCUSSION

As noted in the Introduction, the effect of the radioactive decay half‐life can have an impact on the usability of seeds more than a week old. With activity reductions on the order of 40% after one week (Table 1), it is necessary to incorporate this change in activity into the planning workflow. In and of itself, this is not a problem since it is relatively simple to calculate the current activity of the seeds available and build a treatment plan accordingly. But this only works if there are enough seeds to accommodate an entire implant. If there are fewer partial‐activity seeds than is required for the implant, the partial‐activity seeds are useless; the plan must be discarded and a new plan using full‐activity seeds must be generated. First and foremost, it is clear from Tables 3–6 that there is no degradation of plan quality when using partial activity seeds — both coverage of the target and sparing of the urethra are stable across plans. Given that plan quality is preserved, this algorithm would save both time (by eliminating the possibility that a plan with too many partial‐activity seeds must be discarded) and money (by allowing for the use of any inventory of partial‐activity seeds). Since the algorithm includes input parameters to allow the user to chose a specific number of partial activity seeds, this method will work for any number of partial activity seeds.

A. Post‐implant dosimetry

From a clinical perspective, a challenging element of implementing the method presented in this work is calculating the real dose delivered to the patient due to the actual (rather than planned) implant. Since there is inherent seed placement error for PPI procedures, a post‐implant dose calculation is used to assess the potential efficacy of the implant. It has been standard practice via the recommendations of the Radiation Therapy Oncology Group (RTOG) for this procedure( , ) to obtain a one‐month, post‐implant CT of the patient, digitize the seeds, contour the organs and calculate the actual dose delivered. This may not be sufficient in hybrid cases since there are visually identical seeds with different activities — these must be identified. Thus it is necessary to create a one‐to‐one map between the planned seed positions and the actual positions. This procedure may be complicated by changes in the target volume and shape during the intervening one month between delivery and post‐implant imaging. However, a recent trend( ) in PPI procedure is to obtain the post‐implant CT immediately after the implant procedure (rather than after one month). This would mitigate the difficulties in seed matching which stem from debulking and edema. Obtaining the post‐implant CT immediately after the procedure still would not account for the uncertainty inherent in placing seeds, but recent advances in post‐implant seed identification show that a one‐to‐one map can be generated. Brunet‐Benkhoucha et al.( ) have shown that with only seven X‐ray images taken with a cone‐beam system subtending a 60° arc, seed detection rates of 96.7%, false negative rates of 3.3%, and false positive rates of 2.7% can be achieved. This is done by locating seeds to in 3D space, and resulted in uncertainties in and of 1.5% and 0.3%, respectively.

V. CONCLUSIONS

An optimization algorithm that can generate hybrid brachytherapy plans was developed. Five previously‐treated PPI patients with a range of prostate volumes from 20 to were chosen and reoptimized using the hybrid‐activity PPI optimization. These multi‐activity hybrid plans were equal in quality (as measured by the standard dosimetric indices) to plans with seeds of a single activity. Potential gains achievable by using different radionuclides have recently been explored in the literature; but, since the efficacy of incorporating the BED is still hotly debated in the community, we have focused on the multi‐activity hybrid plans. Despite the expanded search space, optimization times for these studies were still under two minutes on a modern day laptop and can be reduced to below one minute in a clinical setting. With the typical cost of a set of PPI seeds on the order of thousands of dollars, it is possible to reduce the cost of brachytherapy treatments by allowing for easier use of seeds left over from a previous patient or unused due to a cancelled or postponed treatment.
  19 in total

1.  Prostate brachytherapy: comparison of dose distribution with different 125I source designs.

Authors:  D C Beyer; F Puente; K L Rogers; E M Gurgoze
Journal:  Radiology       Date:  2001-12       Impact factor: 11.105

Review 2.  The American Brachytherapy Society recommendations for permanent prostate brachytherapy postimplant dosimetric analysis.

Authors:  S Nag; W Bice; K DeWyngaert; B Prestidge; R Stock; Y Yu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-01-01       Impact factor: 7.038

3.  Inverse planning for HDR prostate brachytherapy used to boost dominant intraprostatic lesions defined by magnetic resonance spectroscopy imaging.

Authors:  Jean Pouliot; Yongbok Kim; Etienne Lessard; I-Chow Hsu; Daniel B Vigneron; John Kurhanewicz
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-07-15       Impact factor: 7.038

4.  Multi-species prostate implant treatment plans incorporating 192Ir and 125I using a Greedy Heuristic based 3D optimization algorithm.

Authors:  V Chaswal; S Yoo; B R Thomadsen; D L Henderson
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

Review 5.  Permanent prostate seed brachytherapy: a current perspective on the evolution of the technique and its application.

Authors:  Arjun Sahgal; Mack Roach
Journal:  Nat Clin Pract Urol       Date:  2007-12

6.  Clinical implementation of a digital tomosynthesis-based seed reconstruction algorithm for intraoperative postimplant dose evaluation in low dose rate prostate brachytherapy.

Authors:  Malik Brunet-Benkhoucha; Frank Verhaegen; Stéphanie Lassalle; Dominic Béliveau-Nadeau; Brigitte Reniers; David Donath; Daniel Taussky; Jean-François Carrier
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

Review 7.  American Brachytherapy Society (ABS) recommendations for transperineal permanent brachytherapy of prostate cancer.

Authors:  S Nag; D Beyer; J Friedland; P Grimm; R Nath
Journal:  Int J Radiat Oncol Biol Phys       Date:  1999-07-01       Impact factor: 7.038

8.  Class solution for inversely planned permanent prostate implants to mimic an experienced dosimetrist.

Authors:  Etienne Lessard; Stefan L S Kwa; Barby Pickett; Mach Roach; Jean Pouliot
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

9.  Results of a phase II trial of transrectal ultrasound-guided permanent radioactive implantation of the prostate for definitive management of localized adenocarcinoma of the prostate (radiation therapy oncology group 98-05).

Authors:  Colleen A Lawton; Michelle DeSilvio; W Robert Lee; Leonard Gomella; David Grignon; Michael Gillin; Gerard Morton; Thomas Pisansky; Howard Sandler
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-11-02       Impact factor: 7.038

10.  A detailed radiobiological and dosimetric analysis of biochemical outcomes in a case-control study of permanent prostate brachytherapy patients.

Authors:  Wayne M Butler; Renee R Stewart; Gregory S Merrick
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

View more
  2 in total

1.  Application programming in C# environment with recorded user software interactions and its application in autopilot of VMAT/IMRT treatment planning.

Authors:  Henry Wang; Lei Xing
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

2.  Hybrid optimization based on non-coplanar needles for brachytherapy dose planning.

Authors:  Xiaodong Ma; Zhiyong Yang; Shan Jiang; Guobin Zhang; Bin Huo; Shude Chai
Journal:  J Contemp Brachytherapy       Date:  2019-06-28
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.