Prostate brachytherapy: comparison of dose distribution with different 125I source designs.

D C Beyer1, F Puente, K L Rogers, E M Gurgoze.   

Abstract

PURPOSE: To evaluate the interchangeability of various commercially available iodine 125 ((125)I) sources and to assess the dosimetric effect of a change in source.
MATERIALS AND METHODS: A modified peripherally loading prostate brachytherapy plan to deliver 145 Gy was devised by using a model (125)I source, which until recently was the only available (125)I source. A dose-volume histogram was generated. By using the available radial dose functions and anisotropy distributions for eight other currently commercially available sources, the same implant placement was planned and dose-volume histogram distributions tabulated. This exercise was performed for 15-, 45-, and 60-cm(3) glands. No implants were placed, and no physical radiation measurements were made. Dose calculations were theoretic: They were generated by using a widely available treatment planning system.
RESULTS: There was little difference in dose distribution to the volume receiving 100% of the prescribed dose (<6%); only one source showed a difference greater than 2%. Large differences, up to -40% to +60%, were seen in the volume of tissue encompassed within internal high-dose regions receiving 150% or 200% of the prescribed dose. These findings held true, irrespective of gland size, within a clinically relevant range (15-60 cm(3)) and for a uniformly loaded radionuclide distribution.
CONCLUSION: Reviewing only peripheral dose at or near the prescription dose of 145 Gy revealed little difference in doses for various source designs. Marked differences in high-dose regions were seen and may affect the dose received by internal sites. Effects of these changes on cure and/or complications remain speculative.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11719655     DOI: 10.1148/radiol.2213010130

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  2 in total

1.  A directional 103Pd brachytherapy device: Dosimetric characterization and practical aspects for clinical use.

Authors:  Mark J Rivard
Journal:  Brachytherapy       Date:  2016-12-28       Impact factor: 2.362

2.  Inverse planning optimization for hybrid prostate permanent-seed implant brachytherapy plans using two source strengths.

Authors:  J Adam M Cunha; Barby Pickett; Jean Pouliot
Journal:  J Appl Clin Med Phys       Date:  2010-06-03       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.