Breast MRI as an adjunct to mammography: Does it really suffer from low specificity? A retrospective analysis stratified by mammographic BI-RADS classes.

Matthias Benndorf1, Pascal A T Baltzer, Tibor Vag, Mieczyslaw Gajda, Ingo B Runnebaum, Werner A Kaiser.   

Abstract

BACKGROUND: Reports on the specificity of breast MRI are heterogeneous, depending on the respective setting of the performed study.
PURPOSE: To retrospectively estimate the sensitivity and especially the specificity of breast MRI in the non-screening setting as an adjunct to mammography sorted by breast density and to estimate the accuracy of breast MRI in cases rated BI-RADS 0 and 3 mammographically.
MATERIAL AND METHODS: A total of 216 consecutive patients with referral to breast MRI and previously acquired mammography were enrolled in this analysis. Negative findings were followed up with a mean time of 26.7 months. The loss to follow-up was 10.8%. The single breast was regarded as the study subject (n=399, 364 cases were eligible for calculation of diagnostic accuracy). BI-RADS 1 and 2 were rated as benign, 4 and 5 as malignant. BI-RADS 0 and 3 were analyzed separately. The 95% confidence intervals (CIs) were calculated from the normally approximated binomial distribution and taken to represent significant differences for the two imaging modalities if they did not overlap.
RESULTS: Among the study population, 62 malignant neoplasms were detected. For cases rated BI-RADS 1, 2, 4, and 5 (n=251), the sensitivity of breast MRI was 95.7% (95% CI 89.9-100.0%) and 74.5% (95% CI 62.0-87.0%) for mammography, respectively. The specificity of breast MRI was 96.1% (95% CI 93.4-98.8%) and 92.2% (95% CI 88.5-95.9%) for mammography, respectively. The diagnostic accuracy of breast MRI did not depend on breast density. In cases rated BI-RADS 0, n=57 (3, n=56), breast MRI achieved a sensitivity of 100% (90.9%) and a specificity of 98.1% (88.9%). There was a significant (P< 0.01) accumulation of dense breast tissue (ACR IV) in breasts rated BI-RADS 0 in mammography. Breast MRI missed three malignant lesions, two of them being smaller than 3 mm.
CONCLUSION: There is no rationale to criticize the low specificity of breast MRI when used as an adjunct to mammography. The independency of the diagnostic accuracy of breast MRI from breast density makes it a worthwhile choice in mammographic BI-RADS 0 cases.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20707656     DOI: 10.3109/02841851.2010.497164

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  15 in total

1.  Diffusion-weighted imaging of breast tumours at 3 Tesla and 7 Tesla: a comparison.

Authors:  S Gruber; L Minarikova; K Pinker; O Zaric; M Chmelik; B Strasser; P Baltzer; T Helbich; S Trattnig; W Bogner
Journal:  Eur Radiol       Date:  2015-08-27       Impact factor: 5.315

2.  Current Status and New Developments in Breast MRI.

Authors:  Katja C Siegmann; Bernhard Krämer; Claus Claussen
Journal:  Breast Care (Basel)       Date:  2011-04-29       Impact factor: 2.860

3.  Fat saturation in dynamic breast MRI at 3 Tesla: is the Dixon technique superior to spectral fat saturation? A visual grading characteristics study.

Authors:  P Clauser; K Pinker; T H Helbich; P Kapetas; M Bernathova; P A T Baltzer
Journal:  Eur Radiol       Date:  2014-05-04       Impact factor: 5.315

4.  A method for limiting pitfalls in the production of enhancement kinetic curves in 3T dynamic magnetic resonance mammography.

Authors:  Eleftherios Lavdas; Panayiotis Mavroidis; Violeta Roka; Nikolaos Arikidis; Dimitrios L Arvanitis; Ioannis V Fezoulidis; Katerina Vassiou
Journal:  J Thorac Dis       Date:  2012-08       Impact factor: 2.895

5.  DCE-MRI of the breast in a stand-alone setting outside a complementary strategy - results of the TK-study.

Authors:  Clemens G Kaiser; C Reich; M Dietzel; P A T Baltzer; J Krammer; K Wasser; S O Schoenberg; W A Kaiser
Journal:  Eur Radiol       Date:  2015-01-11       Impact factor: 5.315

6.  MR mammography using diffusion-weighted imaging in evaluating breast cancer: a correlation with proliferation index.

Authors:  Cristina Molinari; Paola Clauser; Rossano Girometti; Anna Linda; Elisa Cimino; Fabio Puglisi; Chiara Zuiani; Massimo Bazzocchi
Journal:  Radiol Med       Date:  2015-03-17       Impact factor: 3.469

7.  Economic potential of abbreviated breast MRI for screening women with dense breast tissue for breast cancer.

Authors:  Fabian Tollens; Pascal A T Baltzer; Matthias Dietzel; Moritz L Schnitzer; Vincent Schwarze; Wolfgang G Kunz; Johann Rink; Johannes Rübenthaler; Matthias F Froelich; Stefan O Schönberg; Clemens G Kaiser
Journal:  Eur Radiol       Date:  2022-04-28       Impact factor: 5.315

8.  MRI Evaluation of the Contralateral Breast in Women with Recently Diagnosed Breast Cancer: 2-Year Follow-up.

Authors:  Nanette D Debruhl; Su-Ju Lee; Mary C Mahoney; Lucy Hanna; Catherine Tuite; Constantine A Gatsonis; Constance Lehman
Journal:  J Breast Imaging       Date:  2019-11-08

9.  Value of the BI-RADS classification in MR-Mammography for diagnosis of benign and malignant breast tumors.

Authors:  Christian Sohns; Martin Scherrer; Wieland Staab; Silvia Obenauer
Journal:  Eur Radiol       Date:  2011-07-31       Impact factor: 5.315

10.  AutoDensity: an automated method to measure mammographic breast density that predicts breast cancer risk and screening outcomes.

Authors:  Carolyn Nickson; Yulia Arzhaeva; Zoe Aitken; Tarek Elgindy; Mitchell Buckley; Min Li; Dallas R English; Anne M Kavanagh
Journal:  Breast Cancer Res       Date:  2013       Impact factor: 6.466

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.