Measurement of the linear attenuation coefficients of breast tissues by synchrotron radiation computed tomography.

R C Chen1, R Longo, L Rigon, F Zanconati, A De Pellegrin, F Arfelli, D Dreossi, R-H Menk, E Vallazza, T Q Xiao, E Castelli.   

Abstract

The measurement of the linear attenuation coefficients of breast tissues is of fundamental importance in the field of breast x-ray diagnostic imaging. Different groups have evaluated the linear attenuation coefficients of breast tissues by carrying out direct attenuation measurements in which the specimens were thin and selected as homogeneous as possible. Here, we use monochromatic and high-intensity synchrotron radiation computed tomography (SR CT) to evaluate the linear attenuation coefficients of surgical breast tissues in the energy range from 15 to 26.5 keV. X-ray detection is performed by a custom digital silicon micro-strip device, developed in the framework of the PICASSO INFN experiment. Twenty-three human surgical breast samples were selected for SR CT and histological study. Six of them underwent CT, both as fresh tissue and after formalin fixation, while the remaining 17 were imaged only as formalin-fixed tissues. Our results for fat and fibrous tissues are in good agreement with the published values. However, in contrast to the published data, our measurements show no significant differences between fibrous and tumor tissues. Moreover, our results for fresh and formalin-fixed tissues demonstrate a reduction of the linear attenuation coefficient for fibrous and tumor tissues after fixation.

Entities:  

Mesh:

Year:  2010        PMID: 20702925     DOI: 10.1088/0031-9155/55/17/008

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  10 in total

1.  Monochromatic breast computed tomography with synchrotron radiation: phase-contrast and phase-retrieved image comparison and full-volume reconstruction.

Authors:  Luca Brombal; Bruno Golosio; Fulvia Arfelli; Deborah Bonazza; Adriano Contillo; Pasquale Delogu; Sandro Donato; Giovanni Mettivier; Piernicola Oliva; Luigi Rigon; Angelo Taibi; Giuliana Tromba; Fabrizio Zanconati; Renata Longo
Journal:  J Med Imaging (Bellingham)       Date:  2018-11-27

2.  Dose and diagnostic performance comparison between phase-contrast mammography with synchrotron radiation and digital mammography: a clinical study report.

Authors:  Christian Fedon; Luigi Rigon; Fulvia Arfelli; Diego Dreossi; Elisa Quai; Maura Tonutti; Giuliana Tromba; Maria Assunta Cova; Renata Longo
Journal:  J Med Imaging (Bellingham)       Date:  2018-02-01

3.  Radiochromic film dosimetry in synchrotron radiation breast computed tomography: a phantom study.

Authors:  Giovanni Mettivier; Marica Masi; Fulvia Arfelli; Luca Brombal; Pasquale Delogu; Francesca Di Lillo; Sandro Donato; Christian Fedon; Bruno Golosio; Piernicola Oliva; Luigi Rigon; Antonio Sarno; Angelo Taibi; Paolo Russo
Journal:  J Synchrotron Radiat       Date:  2020-04-22       Impact factor: 2.616

4.  Breast tissue characterization with photon-counting spectral CT imaging: a postmortem breast study.

Authors:  Huanjun Ding; Michael J Klopfer; Justin L Ducote; Fumitaro Masaki; Sabee Molloi
Journal:  Radiology       Date:  2014-05-07       Impact factor: 11.105

5.  Dosimetry in x-ray-based breast imaging.

Authors:  David R Dance; Ioannis Sechopoulos
Journal:  Phys Med Biol       Date:  2016-09-12       Impact factor: 3.609

6.  Development and preclinical evaluation of a patient-specific high energy x-ray phase sensitive breast tomosynthesis system.

Authors:  Muhammad U Ghani; Xizeng Wu; Laurie L Fajardo; Zhengxue Jing; Molly D Wong; Bin Zheng; Farid Omoumi; Yuhua Li; Aimin Yan; Peter Jenkins; Stephen L Hillis; Laura Linstroth; Hong Liu
Journal:  Med Phys       Date:  2021-04-01       Impact factor: 4.071

7.  Phantom-Based Feasibility Studies on Phase-Contrast Mammography at Indian Synchrotron Facility Indus-2.

Authors:  Reena Sharma; S D Sharma; P S Sarkar; B Singh; A K Agrawal; D Datta
Journal:  J Med Phys       Date:  2019 Jan-Mar

8.  Experimental optimization of the energy for breast-CT with synchrotron radiation.

Authors:  Piernicola Oliva; Vittorio Di Trapani; Fulvia Arfelli; Luca Brombal; Sandro Donato; Bruno Golosio; Renata Longo; Giovanni Mettivier; Luigi Rigon; Angelo Taibi; Giuliana Tromba; Fabrizio Zanconati; Pasquale Delogu
Journal:  Sci Rep       Date:  2020-10-15       Impact factor: 4.379

9.  Hybrid Detection of Breast Abnormalities Based on Contrast Agents: Introducing a Proof of Concept from a Physics Perspective.

Authors:  Panagiotis Liaparinos
Journal:  Sensors (Basel)       Date:  2022-10-03       Impact factor: 3.847

10.  Three-layer heterogeneous mammographic phantoms for Monte Carlo simulation of normalized glandular dose coefficients in mammography.

Authors:  Tien-Yu Chang; Kuan-Jen Lai; Chun-Yuan Tu; Jay Wu
Journal:  Sci Rep       Date:  2020-02-10       Impact factor: 4.379

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.