Caroline E Wright1, Jennifer Jackson1, Rachel L Thompson1, Alyn H Morice1.
Abstract
UNLABELLED: Protocols measuring cough sensitivity can vary in terms of nebuliser, tussive agent, single and dose response. A definitive method for measuring cough sensitivity needs to be established.The ERS guidelines recommend the KoKo DigiDoser (KD) delivery system. Study aim, was to compare the reproducibility of this citric acid (CA) cough challenge and previously established Mefar dosimeter (MD) protocol.39 (female 26) volunteers mean age (40.4 yrs) were randomised to either KD or MD. Intra-day and inter-day reproducibility was compared.We calculated the concentration of citric acid evoking 2 coughs (C2).The geometric mean C2 (95%CI) was similar for both KD and MD, of 263 (200,339) mM and 209 (151,288) mM respectively.The mean KD C2 was not significantly different. (F = 0.807, p = 0.93) from baseline over 1, 2, and 4 hrs however, the MD demonstrated significant variability (F = 7.85, P < 0.001)Measuring mean log C2 at baseline and at 2 weeks, the KD demonstrated a stronger intraclass correlation of log C2 at baseline with 2 week log C2, ICC = 0.70 than was shown with the Mefar, ICC = 0.41Administering CA from KD offers a reproducible cough challenge in healthy volunteers. The results correlate well with the MD challenge but offer greater intra-day and inter-day reproducibility. TRIAL REGISTRATION: Current controlled trials ISRCTN98385033.Entities:
Year: 2010 PMID: 20698995 PMCID: PMC2925329 DOI: 10.1186/1745-9974-6-8
Source DB: PubMed Journal: Cough ISSN: 1745-9974
IOS indices Pre and Post Citric acid Cough challenge (n = 39).
| IOS Variable | CA cough challenge with KoKo DigiDoser | CA Challenge with Mefar dosimeter | P value |
|---|---|---|---|
| R5 (kPa/L/sec) | 0.331 (0.30,0.36) | 0.334(0.30,0.37) | 0.95 |
| R20 (kPa/L/sec) | 0.281(0.26,0.31) | 0.282 (0.25,0.31) | 0.99 |
| R5-R20 (kPa/L/sec) | 0.050 (0.04,0.60) | 0.052 (0.40,0.64) | 0.89 |
| X | -0.09 (-0.96,-0.8) | -0.09 (-0.10,-0.08) | 0.77 |
| ΔR5(%) | -2.13(-5.52,1.25) | -2.51(-7.72,2.70) | 0.75 |
| ΔR20(%) | -2.27(-5.48,0.94) | -0.12(-4.30,4.08) | 0.53 |
| ΔR5-R20(%) | -5.78(-21.7,10.17) | -34.08(-68.8,0.67) | 0.17 |
| ΔX(%) | 3.51(-0.64,7.65) | -6.89(-15.21,1.42) | 0.03* |
R5 resistance at 5 Hz; R10 resistance at 10 hz; × reactance; Δ, change in measurement after citric acid. Data represented as mean, with 95% confidence interval.
Comparison of mean C2 measured by KoKo DigiDoser and Mefar dosimeter.
| Time point | KoKo DigiDoser (n = 39) | Mefar Dosimeter (n = 39) | P value |
|---|---|---|---|
| Baseline | 2.42 ± 0.35 (2.30-2.53) | 2.32 ± 0.43 (2.18-2.46) | 0.205 |
| 1 hr | 2.45 ± 0.38 (2.32-2.57) | 2.50 ± 0.39 (2.37-2.63) | 0.455 |
| 2 hr | 2.48 ± 0.37 (2.36-2.60) | 2.51 ± 0.42 (2.38-2.66) | 0.556 |
| 4 hr | 2.45 ± 0.33 (2.34-2.55) | 2.53 ± 0.41 (2.40-2.66) | 0.174 |
| 2 wk | 2.37 ± 0.39 (2.24-2.50) | 2.45 ± 0.39 (2.32,2.58) | 0.179 |
P value based on paired ttest
Figure 1Bland and Altman plot for cough C2 thresholds measured by KoKo DigiDoser and Mefar dosimeter Sample figure title.
% Change in Mean logC2 from baseline measured at 1,2,4 hrs and 2 wks following citric acid inhalation challenge with the KoKo DigiDoser versus the Mefar Dosimeter.
| Time point | KoKo DigiDoser (n = 39) | Pearson Correlation | Mefar Dosimeter (n = 39) | Pearson Correlation |
|---|---|---|---|---|
| %Δ Mean C2 log mM | %ΔMean C2 log mM | |||
| 1 hr | -1.57 (-5.01-1.86) | 0.76 | -9.79 (-0.28 - -0.08) | 0.74 |
| 2 hr | -3.14 (-7.50-1.23) | 0.67 | -10.70 (-18.0 - -3.41) | 0.63 |
| 4 hr | -2.08 (-6.12-1.96) | 0.75 | -11.69 (-19.09 - -4.30) | 0.60 |
| 2 wk | 1.36 (-3.04-5.75) | 0.67 | -8.24 (-15.71 - -0.78) | 0.56 |
Figure 2Change in log C2 from baseline at 1,2,4 hrs and 2 wks. Percentage change in log C2 from baseline in repeated cough challenge at 1,2,4 hrs and 2 wks comparing KoKo DigiDoser with the Mefar dosimeter
Figure 3(A) Bland Altman plot of inter-day (baseline -1 hr) repeatability of C2 measurement with KoKo Digidoser N = 39. (B) Bland Altman plot of inter-day (baseline -1 hr) repeatability of C2 measurement with Mefar dosimeter N = 39. Intra-individual differences (n = 38) between mean log C2 measured at baseline and 1 hr post baseline plotted against the mean of the sum scores. On each plot, the central line represents the mean of the intra-individual differences, and the flanking lines represent the 95% limits of agreement.
Within day and between day repeatability of Mean log C2 measurements from baseline measured at 1,2,4 hrs and 2 wks following citric acid inhalation challenge with the KoKo DigiDoser versus the Mefar Dosimeter.
| Method | Within day (n = 39) base 1,2,4 hrs | Between day (n = 39) baseline, 2 wks | ||||||
|---|---|---|---|---|---|---|---|---|
| ICC (95%CI) | SEM | CV(%) | LOA | ICC (95%CI) | SEM | CV(%) | LOA | |
| KoKo Digidoser | 0.68(0.55-0.80 | 0.04 | -2.2 | -3.38, 3.01 | 0.70(0.55-0.86) | 0.05 | 4.2 | -3.34, 4.46 |
| Mefar Dosimeter | 0.67(0.54-0.80) | 0.05 | -14.8 | -5.89, 2.57 | 0.41(0.14-0.67) | 0.06 | -10.7 | -7.80, 4.36 |
ICC: intraclass correlation, CI: confidence interval, LOA: limits of agreement, CV: coefficient of variation.
Figure 4(A) Bland Altman plot of intra-day (baseline -2 wk) repeatability of C2 measurement with KoKo DigiDoser N = 39. (B) Bland Altman plot of intra-day (baseline -2 wk) repeatability of C2 measurement with Mefar dosimeter N = 39. Intra-individual differences (n = 38) between mean log C2 measured at baseline and 1 hr post baseline plotted against the mean of the sum scores. On each plot, the central line represents the mean of the intra-individual differences, and the flanking lines represent the 95% limits of agreement.