Factors influencing resonance frequency analysis assessed by Osstell mentor during implant tissue integration: I. Instrument positioning, bone structure, implant length.

Christina P C Sim1, Niklaus P Lang.   

Abstract

AIM: To monitor longitudinally the development of implant stability of SLA Straumann tissue-level implants using resonance frequency analysis (RFA) and to determine the influence of instrument positioning, bone structure and implant length on the assessment of RFA.
MATERIAL AND METHODS: Thirty-two healthy adult patients received either 8 mm, v4.1 mm Straumann Standard Plus tissue-level implants (n=16: Group A) or 10 mm, v4.1 mm Straumann Standard Plus tissue-level implants (n=16: Group B). During healing, RFA was performed on Weeks 0,1, 2, 3, 4, 5, 6, 8 and 12. The implants were restored after 10 weeks (impression taking) and 12 weeks. In addition, probing depth, presence of plaque and bleeding on probing were assessed. Implant stability quotient (ISQ) values of Groups A and B were compared using unpaired t-tests and longitudinally applying paired t-tests between Week 0 and the subsequent time points.
RESULTS: Positioning of the Osstell mentor device did not affect the ISQ values. Generally, ISQ values increased continuously during healing from a mean of 65.1 (SD 16.97) to 74.7 (SD 5.17) (significantly from Week 0 to Weeks 6, 8 and 12). Lower bone density (Type III or IV) resulted in significantly lower ISQ values up to Week 8. Implant length affected the increase in ISQ values over time. While no significant increase was observed with 10 mm implants, ISQ values of 8 mm implants increased significantly from Week 0 to Weeks 6, 8 and 12.
CONCLUSIONS: Using Osstell mentor, ISQ values are reproducible irrespective of instrument positioning. ISQ values are affected by the bone structure and implant length. Hence, no predictive values can be attributed to implant stability.

Entities:  

Mesh:

Year:  2010        PMID: 20666786     DOI: 10.1111/j.1600-0501.2009.01878.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  28 in total

1.  Effect of Magnetic Field on Bone Healing around Endosseous Implants - An In-vivo Study.

Authors:  Mahalakshmi Gujjalapudi; Chandrasekar Anam; Praveen Mamidi; Radha Chiluka; A Gautam Kumar; Ragini Bibinagar
Journal:  J Clin Diagn Res       Date:  2016-10-01

2.  Influence of cortical bone anchorage on the primary stability of dental implants.

Authors:  Anne Kelly de Oliveira Nicolau Mantovani; Ivete Aparecida de Mattias Sartori; Luciana Reis Azevedo-Alanis; Rodrigo Tiossi; Flávia Noemy Gasparini Kiatake Fontão
Journal:  Oral Maxillofac Surg       Date:  2018-06-06

3.  Evaluation of the correlation between insertion torque and primary stability of dental implants using a block bone test.

Authors:  Dorjpalam Bayarchimeg; Hee Namgoong; Byung Kook Kim; Myung Duk Kim; Sungtae Kim; Tae-Il Kim; Yang Jo Seol; Yong Moo Lee; Young Ku; In-Chul Rhyu; Eun Hee Lee; Ki-Tae Koo
Journal:  J Periodontal Implant Sci       Date:  2013-02-28       Impact factor: 2.614

4.  Mini-implant stability at the initial healing period: a clinical pilot study.

Authors:  Manuel Nienkemper; Benedict Wilmes; Alexander Pauls; Dieter Drescher
Journal:  Angle Orthod       Date:  2013-07-24       Impact factor: 2.079

5.  A Radiographic and Clinical Comparison of Immediate vs. Early Loading (4 Weeks) of Implants with a New Thermo-Chemically Treated Surface: A Randomized Clinical Trial.

Authors:  Matteo Albertini; Federico Herrero-Climent; Carmen María Díaz-Castro; Jose Nart; Ana Fernández-Palacín; José Vicente Ríos-Santos; Mariano Herrero-Climent
Journal:  Int J Environ Res Public Health       Date:  2021-01-29       Impact factor: 3.390

6.  Effect of bone quality and implant surgical technique on implant stability quotient (ISQ) value.

Authors:  Hong-Gi Yoon; Seong-Joo Heo; Jai-Young Koak; Seong-Kyun Kim; Su-Young Lee
Journal:  J Adv Prosthodont       Date:  2011-03-31       Impact factor: 1.904

7.  The relationship between initial implant stability quotient values and bone-to-implant contact ratio in the rabbit tibia.

Authors:  In-Phill Park; Seong-Kyun Kim; Shin-Jae Lee; Joo-Hee Lee
Journal:  J Adv Prosthodont       Date:  2011-06-30       Impact factor: 1.904

8.  In vitro evaluation of resonance frequency analysis values to different implant contact ratio and stiffness of surrounding material.

Authors:  Mu-Seung Kwak; Seok-Gyu Kim
Journal:  J Adv Prosthodont       Date:  2013-11-28       Impact factor: 1.904

9.  Impact of mini-implant length on stability at the initial healing period: a controlled clinical study.

Authors:  Manuel Nienkemper; Benedict Wilmes; Alexander Pauls; Dieter Drescher
Journal:  Head Face Med       Date:  2013-10-20       Impact factor: 2.151

10.  Stability Testing of a Wide Bone-Anchored Device after Surgery without Skin Thinning.

Authors:  Malou Hultcrantz
Journal:  Biomed Res Int       Date:  2015-07-05       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.