Ventricular asystole due to inhibition of ventricular backup pacing by a dual chamber pacemaker: possible pitfalls of default settings.

F M van de Sandt1, R Jansen, G P Kimman, J H Ruiter.   

Abstract

A 62-year-old patient presented with dizzy spells after her dual chamber pacemaker (Medtronic Enrhythm P1501DR), implanted for complete AV block, had been reprogrammed to deliver antitachycardia therapy (ATP) for paroxysmal atrial tachycardia. Her symptoms were caused by inhibition of ventricular backup pacing during ATP, leading to ventricular asystoles. Inhibition was the result of premature ventricular beats occurring prior to ATP: when ventricular backup pacing is left in the default setting, this pacemaker withholds backup pacing if any of the four preceding events is a sensed event. This case illustrates the possibly hazardous effects of default pacemaker settings, especially in pacemaker-dependent patients. (Neth Heart J 2010;18:323-6.).

Entities:  

Keywords:  Arrhythmias; Cardiac Pacing; Pacemaker; Ventricular Function

Year:  2010        PMID: 20657679      PMCID: PMC2881350          DOI: 10.1007/BF03091784

Source DB:  PubMed          Journal:  Neth Heart J        ISSN: 1568-5888            Impact factor:   2.380


  4 in total

1.  Ventricular asystole during atrial antitachycardia pacing by implanted cardioverter-defibrillator.

Authors:  Bengt Herweg; S Serge Barold
Journal:  Pacing Clin Electrophysiol       Date:  2005-05       Impact factor: 1.976

2.  Failure to pace: pacemaker malfunction?

Authors:  Antonio Navarrete
Journal:  J Cardiovasc Electrophysiol       Date:  2007-10-05

3.  Inhibition of a dual chamber pacemaker by oversensing of far-field atrial depolarisation in a patient after His-bundle ablation.

Authors:  J H Ruiter; M J Barrett; L Weteling; R Jansen
Journal:  Neth Heart J       Date:  2004-12       Impact factor: 2.380

4.  Ventricular asystole due to inhibition of ventricular backup pacing by a dual chamber pacemaker: possible pitfalls of default settings.

Authors:  F M van de Sandt; R Jansen; G P Kimman; J H Ruiter
Journal:  Neth Heart J       Date:  2010-06       Impact factor: 2.380

  4 in total
  1 in total

1.  Ventricular asystole due to inhibition of ventricular backup pacing by a dual chamber pacemaker: possible pitfalls of default settings.

Authors:  F M van de Sandt; R Jansen; G P Kimman; J H Ruiter
Journal:  Neth Heart J       Date:  2010-06       Impact factor: 2.380

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.