Literature DB >> 20627771

Effect of unilateral misfit on preload of retention screws of implant-supported prostheses submitted to mechanical cycling.

Wirley Gonçalves Assunção1, Valentim Adelino Ricardo Barão, Juliana Aparecida Delben, Érica Alves Gomes, Idelmo Rangel Garcia.   

Abstract

PURPOSE: The aim of this study was to evaluate the effect of different levels of unilateral angular misfit on preload maintenance of retention screws of single implant-supported prostheses submitted to mechanical cycling.
MATERIALS AND METHODS: Premachined UCLA abutments were cast with cobalt-chromium alloy to obtain 48 crowns divided into four groups (n=12). The crowns presented no misfit in Group A (control group) and unilateral misfits of 50 μm, 100 μm and 200 μm in the groups B, C and D, respectively. The crowns were attached to external hexagon implants with a titanium retention screw with torque of 30 N/cm. Oblique loading of 130 N at 2 Hz was applied on each replica, totalizing 5×10⁴ and 1×10⁶ cycles. Detorque values were measured initially and after each cycling period. Data were evaluated by analysis of variance and Tukey's HSD test (p<0.05).
RESULTS: All groups presented reduced initial detorque values (p<0.05) in comparison to the insertion torque (30±0.5 N/cm) and Group A (25.18 N/cm) exhibited the lowest reduction. After mechanical cycling, all groups presented detorque values from 19.5 N/cm to 22.38 N/cm and the mechanical cycling did not statistically influence the detorque values regardless the misfit level of the replicas.
CONCLUSION: The unilateral misfit influenced the preload maintenance only before mechanical cycling. The mechanical cycling did not influence the torque reduction.
Copyright © 2010 Japan Prosthodontic Society. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20627771     DOI: 10.1016/j.jpor.2010.05.002

Source DB:  PubMed          Journal:  J Prosthodont Res        ISSN: 1883-1958            Impact factor:   4.642


  7 in total

1.  Wear of Morse taper and external hexagon implant joints after abutment removal.

Authors:  Abraão M Prado; Jorge Pereira; Filipe S Silva; Bruno Henriques; Rubens M Nascimento; Cesar A M Benfatti; José López-López; Júlio C M Souza
Journal:  J Mater Sci Mater Med       Date:  2017-03-20       Impact factor: 3.896

2.  Stability of external and internal implant connections after a fatigue test.

Authors:  Paulo Cesar Pinheiro Feitosa; Ana Paula Barbosa de Lima; Laís Regiane Silva-Concílio; William Cunha Brandt; Ana Christina Claro Neves
Journal:  Eur J Dent       Date:  2013-07

3.  Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force.

Authors:  Yuseung Yi; Seong-Joo Heo; Jai-Young Koak; Seong-Kyun Kim
Journal:  J Adv Prosthodont       Date:  2019-12-18       Impact factor: 1.904

4.  Analysis of Torque Maintenance and Fracture Resistance after Fatigue in Retention Screws Made of Different Metals for Screw-Retained Implant-Borne Prosthesis Joints.

Authors:  Maria Beatriz Bello Taborda; Gabriela Sumie Yaguinuma Gonçalves; Cecília Alves de Sousa; Wirley Gonçalves Assunção
Journal:  Int J Dent       Date:  2021-11-18

5.  Implant-Abutment Misfit after Cyclic Loading: An In Vitro Experimental Study.

Authors:  John Eversong Lucena de Vasconcelos; Jefferson David Melo de Matos; Daher Antonio Queiroz; Guilherme da Rocha Scalzer Lopes; Bruna Caroline Gonçalves Vasconcelos de Lacerda; Marco Antonio Bottino; Cecilia Pedroso Turssi; Roberta Tarkany Basting; Flávia Lucisano Botelho do Amaral; Fabiana Mantovani Gomes França
Journal:  Materials (Basel)       Date:  2022-08-03       Impact factor: 3.748

6.  Influence of Manufacturing Methods of Implant-Supported Crowns on External and Internal Marginal Fit: A Micro-CT Analysis.

Authors:  Izabela C M Moris; Silas Borges Monteiro; Raíssa Martins; Ricardo Faria Ribeiro; Erica A Gomes
Journal:  Biomed Res Int       Date:  2018-01-23       Impact factor: 3.411

7.  Comparison of the Effect of Four Different Abutment Screw Torques on Screw Loosening in Single Implant-Supported Prosthesis after the Application of Mechanical Loading.

Authors:  Behnaz Ebadian; Amirhossein Fathi; Saba Khodadad
Journal:  Int J Dent       Date:  2021-07-19
  7 in total