Literature DB >> 2061742

Rapid and specific cyclosporine assay for the Du Pont aca discrete clinical analyzer performed directly on whole blood.

J B Hansen1, H P Lau, C J Janes, W K Miller, D A Wiebe.   

Abstract

A rapid, sensitive, and specific enzyme immunometric assay for cyclosporine in whole blood has been developed for the aca analyzer, using a monoclonal antibody from Sandoz, Ltd. Between-run CVs ranged from 6.5 to 7.6% for samples containing between 60 and 400 ng/mL cyclosporine. Sensitivity was better than 25 ng/mL in the assay, which has an effective upper range of greater than 600 ng/mL. Two correlation studies compared cyclosporine values from the Du Pont method to those determined by HPLC procedures in three hospital laboratories. The results from a total of 120 whole blood samples with CsA between 20 and 800 ng/mL showed excellent correlation between the methodologies. HPLC and Du Pont CsA values from 10 day serial studies also correlated well for samples from a kidney, kidney-pancreas, heart, and two liver transplant patients. We conclude that the Du Pont CSA assay provides accurate and reproducible results in a convenient format in less than 30 minutes.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 2061742     DOI: 10.1002/jcla.1860050307

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  1 in total

1.  Determination of benzodiazepines in clinical serum samples: comparative evaluation of REMEDi system, aca analyzer, and conventional HPLC performance.

Authors:  T L Chang; K W Chen; Y D Lee; K Fan
Journal:  J Clin Lab Anal       Date:  1999       Impact factor: 2.352

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.