Determination of benzodiazepines in clinical serum samples: comparative evaluation of REMEDi system, aca analyzer, and conventional HPLC performance.

T L Chang1, K W Chen, Y D Lee, K Fan.   

Abstract

Emergency toxicology or drug screening in clinical settings requires rapid qualitative and quantitative analysis with acceptable levels of sensitivity and specificity. The aim of this study was to comparatively evaluate the performance of the multi-column HPLC-based REMEDi drug profiling system (Bio-Rad), the aca analyzer (Du Pont), and the bench standard conventional HPLC method in the identification of 12 clinically important benzodiazepines. In this study, the presence of benzodiazepines in 133 patients' serum samples were qualitatively and comparatively analyzed by these three procedures. Among these methods, 120 of 133 samples were identified as benzodiazepine-positive by conventional HPLC (90%); 127 by aca analyzer (95%); and 84 by REMEDi (63%). Detection sensitivity of REMEDi for most of the benzodiazepines was found satisfactory when concentrations were greater than 1.0 microg/mL. When benzodiazepine concentrations were in the ranges of 0.3-1.0 microg/mL, detection sensitivity became varied among the benzodiazepine family of drugs by REMEDi method. REMEDi procedure should not be considered as the method of choice for detection of benzodiazepines if expected concentration levels are below 0.3 microg/mL, with the exception of bromazepam. Conventional HPLC displayed the highest sensitivity and specificity for the detection of benzodiazepines. In our studies, 36 REMEDi-negative samples were positive by HPLC, although in 16 of the 36 REMEDi negative samples (13.3%), the presence of benzodiazepines was detected but only listed as candidates without positive identification of the individual compounds by REMEDi. In our series, however, there were no false positives by the REMEDi method whereas aca procedure showed 13 false positive results (9%) and 6 cases of false negative (4%). Our data indicate that the REMEDi procedure can be used on serum samples for rapid qualitative screening of clinically important high levels of benzodiazepines with high specificity. However, due to the relatively low sensitivity of REMEDi in samples with low benzodiazepine levels and relatively low specificity by aca, all samples should be further confirmed by conventional HPLC procedure.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10323474      PMCID: PMC6807818     

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  8 in total

1.  REMEDi drug profiling system readily distinguishes between cyclobenzaprine and amitriptyline in emergency toxicology urine specimens.

Authors:  A Poklis; L E Edinboro
Journal:  Clin Chem       Date:  1992-11       Impact factor: 8.327

2.  Single procedure for detection, confirmation, and quantification of benzodiazepines in serum by liquid chromatography with photodiode-array detection.

Authors:  P R Puopolo; M E Pothier; S A Volpicelli; J G Flood
Journal:  Clin Chem       Date:  1991-05       Impact factor: 8.327

3.  Identification of drugs in physiological fluids following on-line liquid chromatographic purification and analysis.

Authors:  A K Adams; H Essien; S R Binder
Journal:  Ann Biol Clin (Paris)       Date:  1991       Impact factor: 0.459

4.  Automated liquid-chromatographic analyzer used for toxicology screening in a general hospital: 12 months' experience.

Authors:  N Sadeg; G François; B Petit; H Dutertre-Catella; M Dumontet
Journal:  Clin Chem       Date:  1997-03       Impact factor: 8.327

5.  Automated liquid chromatographic analysis of drugs in urine by on-line sample cleanup and isocratic multi-column separation.

Authors:  S R Binder; M Regalia; M Biaggi-McEachern; M Mazhar
Journal:  J Chromatogr       Date:  1989-07-07

6.  The development of a broad-spectrum toxicology screening program in Taiwan.

Authors:  J S Chen; K J Chang; R C Charng; S J Lai; S R Binder; H Essien
Journal:  J Toxicol Clin Toxicol       Date:  1995

7.  Evaluation of the REMEDi drug profiling system.

Authors:  P Demedts; A Wauters; F Franck; H Neels
Journal:  Eur J Clin Chem Clin Biochem       Date:  1994-05

8.  Rapid and specific cyclosporine assay for the Du Pont aca discrete clinical analyzer performed directly on whole blood.

Authors:  J B Hansen; H P Lau; C J Janes; W K Miller; D A Wiebe
Journal:  J Clin Lab Anal       Date:  1991       Impact factor: 2.352

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.