Comparison of high definition oscillometric and Doppler ultrasonic devices for measuring blood pressure in anaesthetised cats.

Aleksandra Domanjko Petric1, Zrimsek Petra, Sredensek Jerneja, Seliskar Alenka.   

Abstract

Indirect blood pressure measurements using high definition oscillometric (HDO) and Doppler devices were compared in 50 anaesthetised client-owned cats presented for various surgical procedures. Sites of cuff placement for Doppler were identified as forelimb and hindlimb and for HDO as forelimb and tail. Oscillometric and Doppler readings were obtained in 90.05% and 100% of attempts, respectively. Both devices enabled precise measurement of systolic arterial pressure (SAP), although the Doppler device gave higher precision. In the low pressure group (SAP<100 mmHg; n=30), 66.7% biases were within 10 mmHg of discrepancy, but in groups of normal (100 mmHg ≤ SAP ≤ 150 mmHg; n=120) and high measurements (SAP>150 mmHg; n=62), 86.7% and 75.0% of discrepancy, respectively, were lower than 10 mmHg. Frequency of discrepancy at the range of 15 mmHg showed similar differences between pressure groups. There were significantly higher discrepancies when the cuff was positioned on the tail rather than on the leg. The SAP value obtained by HDO can be calculated from the Doppler measurement from SAP (HDO)=0.8515 × SAP (Doppler)+19.221 mmHg. Compared to Doppler, HDO overestimated low pressure and underestimated high pressure values.
Copyright © 2010 ISFM and AAFP. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20615734     DOI: 10.1016/j.jfms.2010.04.007

Source DB:  PubMed          Journal:  J Feline Med Surg        ISSN: 1098-612X            Impact factor:   2.015


  4 in total

1.  Blood Pressure Reference Intervals for Ketamine-sedated Rhesus Macaques (Macaca mulatta).

Authors:  Rachel D Brownlee; Philip H Kass; Rebecca L Sammak
Journal:  J Am Assoc Lab Anim Sci       Date:  2020-01-02       Impact factor: 1.232

2.  Comparison of Doppler ultrasonic and oscillometric devices (with or without proprietary optimisations) for non-invasive blood pressure measurement in conscious cats.

Authors:  Petra Cerna; Panos E Archontakis; Hester Ok Cheuk; Danièlle A Gunn-Moore
Journal:  J Feline Med Surg       Date:  2020-06-25       Impact factor: 2.015

3.  'The Mercury Challenge': feline systolic blood pressure in primary care practice - a European survey.

Authors:  Andrew Sparkes; Catherine Garelli-Paar; Thomas Blondel; Emilie Guillot
Journal:  J Feline Med Surg       Date:  2022-06-27       Impact factor: 1.971

4.  Indirect arterial blood pressure measurement in healthy anesthetized cats using a device that combines oscillometry with photoplethysmography.

Authors:  Yasuhiro Heishima; Yasutomo Hori; Seishiro Chikazawa; Kazutaka Kanai; Fumio Hoshi; Naoyuki Itoh
Journal:  J Vet Med Sci       Date:  2016-03-18       Impact factor: 1.267

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.