Comparing measured total carbon dioxide and calculated bicarbonate.

Nadzimah Mohd Nasir1, Pavai Sthaneshwar, Putri Junaidah Megat Yunus, Sook-Fan Yap.   

Abstract

INTRODUCTION: The objective of the study is to determine the level of agreement between measured total carbon dioxide (TCO2) and calculated bicarbonate (HCO3-) in our laboratory.
MATERIALS AND METHODS: TCO2 and HCO3- values of 1820 samples drawn at the same time from the patient were compared. TCO2 from venous samples was measured on Dimension RxL while HCO3- was obtained from arterial blood gas samples analyzed on Radiometer ABL 700.
RESULTS: The TCO2 and HCO3- values correlated well (r = 0.977, p < 0.001), with the correlation given by the equation, y = 0.986x - 0.5335. Using Bland-Altman analysis, the bias was 0.87 mmol/L (SD 1.42 mmol/L), and the limits of agreement (LOA) were -1.92 to 3.67 mmol/L. Story and Poustie's criteria were applied to study the agreement between these two methods. Based on the first criterion that the bias between TCO2 and HCO3- should be less than +/- 1 mmol/L, the results for the two methods appear to be in good agreement. The second criterion requires that the LOA between the two methods should range between a bias of +/- 2 mmol/L or a total span of 4 mmol/L; the LOA was exceeded in our study. Using the total allowable error in the Bland Altman plot also showed that the two values cannot be used interchangeably especially at the lower values.
CONCLUSIONS: TCO2 did not show good agreement with HCO3-. Clinicians should be aware of this discrepancy and hence should be cautious when using HCO3- for management of acid-base disorders.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20614722

Source DB:  PubMed          Journal:  Malays J Pathol        ISSN: 0126-8635            Impact factor:   0.656


  6 in total

Review 1.  Anesthetic Management of a Child With Unspecified Mitochondrial Disease in an Outpatient Dental Setting.

Authors:  Taylor R Gordon; Richard J Montandon
Journal:  Anesth Prog       Date:  2017

2.  Serum Bicarbonate as a Surrogate for pH in Hemodialysis: A Pilot Study.

Authors:  Tyler B Woodell; Dena E Rifkin; David H Ellison; Dan Negoianu; Jessica W Weiss
Journal:  Kidney Med       Date:  2019-10-22

3.  Comparison of values of traditionally measured venous bicarbonate with calculated arterial bicarbonate in intensive care unit patients of a hospital in a third-world country.

Authors:  Waheeda Nargis; Akm Shafiqur Rahman; Borhan Uddin Ahamed; Md Zakir Hossain
Journal:  Niger Med J       Date:  2014-07

Review 4.  Sodium bicarbonate therapy in patients with metabolic acidosis.

Authors:  María M Adeva-Andany; Carlos Fernández-Fernández; David Mouriño-Bayolo; Elvira Castro-Quintela; Alberto Domínguez-Montero
Journal:  ScientificWorldJournal       Date:  2014-10-21

5.  Correlation between peripheral venous and arterial blood gas measurements in patients admitted to the intensive care unit: A single-center study.

Authors:  Bo Ra Kim; Sae Jin Park; Ho Sik Shin; Yeon Soon Jung; Hark Rim
Journal:  Kidney Res Clin Pract       Date:  2013-01-29

6.  Comparison of bicarbonate values from venous blood gas and chemistry panels measured at the time of diagnosis and resolution of diabetes ketoacidosis.

Authors:  Poorani Nallam Goundan; Devina Luhur Willard; Ayse Sahin-Efe; Shu-Ling Liang Fan; Sara Michelle Alexanian
Journal:  J Clin Transl Endocrinol       Date:  2019-08-16
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.