Literature DB >> 20601772

Water calibration for CT scanners with tube voltage modulation.

Ludwig Ritschl1, Frank Bergner, Christof Fleischmann, Marc Kachelriess.   

Abstract

X-ray CT measures the attenuation of polychromatic x-rays through an object of interest. The CT data acquired are the negative logarithm of the relative x-ray intensity after absorption. These data must undergo water precorrection to linearize the measured data and convert them into line integrals through the patient that can be reconstructed to yield the final CT image. The function to linearize the measured projection data depends on the tube voltage U. In most circumstances, CT scans are carried out with a constant tube voltage. For those cases there are dozens of different techniques to carry out water precorrection. In our case the tube voltage is rather modulated as a function of the object. We propose an empirical cupping correction (ECCU) algorithm to correct for CT cupping artifacts that are induced by nonlinearities in the projection data. The method is rawdata based, empirical and requires neither knowledge of the x-ray spectrum nor of the attenuation coefficients. It aims at linearizing the attenuation data using a precorrection function of polynomial form in the polychromatic attenuation data q and in the tube voltage U. The coefficients of the polynomial are determined once using a calibration scan of a homogeneous phantom. The coefficients are computed in the image domain by fitting a series of basis images to a template image. The template image is obtained directly from the uncorrected phantom image and no assumptions on the phantom size or of its positioning are made. Rawdata are precorrected by passing them through the once-determined polynomial. Numerical examples are shown to demonstrate the quality of the precorrection. ECCU is achieved to remove the cupping artifacts and to obtain well-calibrated CT values. A combination of ECCU with analytical techniques yielding a hybrid cupping correction method is possible and allows for channel-dependent correction functions.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20601772     DOI: 10.1088/0031-9155/55/14/010

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Varying kVp as a means of reducing CT breast dose to pediatric patients.

Authors:  K B Mathieu; A C Turner; M Khatonabadi; M F McNitt-Gray; C H Cagnon; D D Cody
Journal:  Phys Med Biol       Date:  2013-06-12       Impact factor: 3.609

2.  An Investigation of Calibration Phantoms for CT Scanners with Tube Voltage Modulation.

Authors:  Jing Zou; Xiaodong Hu; Hanyu Lv; Xiaotang Hu
Journal:  Int J Biomed Imaging       Date:  2013-12-25
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.