Literature DB >> 20577755

Corneal flap assessment with Rondo microkeratome in laser in situ keratomileusis.

Eleftherios I Paschalis1, Antonis P Aristeidou, Nikitas C Foudoulakis, Lambros A Razis.   

Abstract

PURPOSE: To assess the accuracy of flap thickness in laser in situ keratomileusis (LASIK) with Rondo, Wavelight AG microkeratome and to examine factors that can influence flap thickness (FT).
SETTING: The study took place at the Laser & Ophthalmos Eye Clinic, Thessaloniki, Greece.
MATERIAL AND METHODS: Three hundred and sixty eyes from 180 patients underwent LASIK with Rondo microkeratome. Three surgeons (A, B and C) performed all surgeries with no previous experience of Rondo microkeratome. All patients were treated with the 130 μm plate. Central corneal thickness (CCT) and stromal bed thickness were measured by Scheimpflug and ultrasound pachymetry. Right eye (OD) was treated first.
RESULTS: Mean FT for OD: 120 ± 19 μm (range 69-158 μm); for left eye (OS): 106 ± 17 μm (range 70-147 μm). Flaps in OD were significantly thicker than in OS (p < 0.001). FT was significantly correlated to the keratometric reading (K): r = 0.121; p = 0.02. No correlation was found between FT and CCT or between FT and the attempted refractive correction (SE) (p > 0.14). Mean FT was significantly lower than the manufacturer's 130 μm specification (Mean FT = 113 ± 19 μm; p < 0.001). FT between surgeons A, B and C was significantly different (analysis of variance between surgeons; p < 0.001). Scheimpflug and ultrasound CCT measurements were significantly correlated (r = 0.921; p < 0.001) with ultrasound measuring an average 4.5 μm higher than Scheimpflug (CCT (Oculyzer) = 553.96 ± 27 μm; CCT (Ultrasound) = 558.45 ± 28 μm). Mean flap diameter was 9.2 ± 0.2 mm.
CONCLUSIONS: FT with Rondo microkeratome was significantly influenced by the mean preoperative K reading. First treated eye was significantly thicker than the fellow left eye, while both were significantly lower than the recommended 130 μm thickness. Gaining basic experience of Rondo microkeratome required an average of 90 flaps/surgeon.

Entities:  

Mesh:

Year:  2010        PMID: 20577755     DOI: 10.1007/s00417-010-1433-7

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  38 in total

1.  Corneal flap dimensions in laser in situ keratomileusis using the Innovatome automatic microkeratome.

Authors:  Y I Choi; S J Park; B J Song
Journal:  Korean J Ophthalmol       Date:  2000-06

2.  Online optical coherence pachymetry as a safety measure for laser in situ keratomileusis treatment in 1859 cases.

Authors:  Irmingard M Neuhann; Barbara A M Lege; Markus Bauer; Joerg M Hassel; Anton Hilger; Thomas F Neuhann
Journal:  J Cataract Refract Surg       Date:  2008-08       Impact factor: 3.351

3.  LASIK flap creation with the Ziemer femtosecond laser in 787 consecutive eyes.

Authors:  Juhani Pietilä; Anne Huhtala; Markus Jääskeläinen; Jarmo Jylli; Petri Mäkinen; Hannu Uusitalo
Journal:  J Refract Surg       Date:  2010-01-11       Impact factor: 3.573

4.  Experience during the learning curve of laser in situ keratomileusis.

Authors:  H V Gimbel; S Basti; G B Kaye; M Ferensowicz
Journal:  J Cataract Refract Surg       Date:  1996-06       Impact factor: 3.351

5.  Laser in situ keratomileusis to correct high myopia.

Authors:  J J Pérez-Santonja; J Bellot; P Claramonte; M M Ismail; J L Alió
Journal:  J Cataract Refract Surg       Date:  1997-04       Impact factor: 3.351

6.  Corneal ectasia induced by laser in situ keratomileusis.

Authors:  I G Pallikaris; G D Kymionis; N I Astyrakakis
Journal:  J Cataract Refract Surg       Date:  2001-11       Impact factor: 3.351

7.  Experience with the Nidek MK-2000 microkeratome in 1,220 cases.

Authors:  K A Sarkisian; A A Petrov
Journal:  J Refract Surg       Date:  2001 Mar-Apr       Impact factor: 3.573

8.  Evaluation of the reliability of the Nidek MK-2000 microkeratome for laser in situ keratomileusis.

Authors:  P Naripthaphan; A Vongthongsri
Journal:  J Refract Surg       Date:  2001 Mar-Apr       Impact factor: 3.573

9.  Nidek MK 2000 microkeratome clinical evaluation.

Authors:  Maria Clara Arbelaez
Journal:  J Refract Surg       Date:  2002 May-Jun       Impact factor: 3.573

10.  Iatrogenic keratectasia after laser in situ keratomileusis.

Authors:  T Seiler; K Koufala; G Richter
Journal:  J Refract Surg       Date:  1998 May-Jun       Impact factor: 3.573

View more
  2 in total

1.  Digital analysis of flap parameter accuracy and objective assessment of opaque bubble layer in femtosecond laser-assisted LASIK: a novel technique.

Authors:  A John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2013-02-15

Review 2.  The evolution of corneal and refractive surgery with the femtosecond laser.

Authors:  Antonis Aristeidou; Elise V Taniguchi; Michael Tsatsos; Rodrigo Muller; Colm McAlinden; Roberto Pineda; Eleftherios I Paschalis
Journal:  Eye Vis (Lond)       Date:  2015-07-14
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.