Literature DB >> 11316036

Evaluation of the reliability of the Nidek MK-2000 microkeratome for laser in situ keratomileusis.

P Naripthaphan1, A Vongthongsri.   

Abstract

PURPOSE: To evaluate the performance of the Nidek MK-2000 microkeratome in corneal flap creation for laser in situ keratomileusis (LASIK) in Asian eyes and the simplicity of its use for the beginner.
METHODS: A prospective study of 200 Asian eyes was performed. The corneal flaps were created using the Nidek MK-2000 microkeratome 130-microm and 160-microm thickness plates with 8.5- and 9.5-mm suction rings. Ease of obtaining suction was evaluated and measurements were done for flap thickness, flap diameter, hinge size, and hinge-to-center-of-pupil distance.
RESULTS: Subjectively, even in small eyes, the procedure of obtaining the required suction and creating a good flap was easy and reliable with the Nidek MK-2000 microkeratome. Average flap thickness was 120.52 +/- 16.49 microm (range, 84 to 162 microm) for the 130-microm thickness plate with the 8.5-mm suction ring. Mean flap diameter was 9.16 +/- 0.43 mm, mean hinge size was 4.96 +/- 0.36 mm, and mean hinge-to-center-of-pupil distance was 4.16 +/- 0.28 mm. With the 9.5-mm suction ring, mean flap thickness was 122.06 +/- 18.54 microm (range, 84 to 149 microm), mean flap diameter was 9.85 +/- 0.35 mm, mean hinge size was 5.28 +/- 0.36 mm, and mean hinge-to-center-of-pupil distance was 4.49 +/- 0.27 mm. For the 160-microm plate with a 8.5-mm suction ring, the flap thickness was 172.71 +/- 27.49 microm (range, 121 to 203 microm), flap diameter was 9.13 + 0.37 mm, mean hinge size was 5.02 +/- 0.29 mm and mean hinge-to-center-of-pupil distance was 4.21 +/- 0.29 mm. There were no complications of corneal flap creation.
CONCLUSION: The Nidek MK-2000 microkeratome was effective, safe, and reliable, even in small eyes.

Entities:  

Mesh:

Year:  2001        PMID: 11316036     DOI: 10.3928/1081-597X-20010302-25

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  3 in total

Review 1.  [Femtosecond laser for refractive corneal surgery: foundations, mode of action and clinical applications].

Authors:  M Mrochen; A Donges; G Korn
Journal:  Ophthalmologe       Date:  2006-12       Impact factor: 1.059

2.  Corneal flap assessment with Rondo microkeratome in laser in situ keratomileusis.

Authors:  Eleftherios I Paschalis; Antonis P Aristeidou; Nikitas C Foudoulakis; Lambros A Razis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-06-25       Impact factor: 3.117

3.  Histopathological study of corneal flap striae following laser in situ keratomileusis in rabbits.

Authors:  Li Liu; Fang-Zhou Song; Lian-Yun Bao
Journal:  Exp Ther Med       Date:  2015-01-08       Impact factor: 2.447

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.