Literature DB >> 20537831

HydroSoft coil versus HydroCoil for endovascular aneurysm occlusion study: a single center experience.

Xin-Bin Guo1, Yi-Mu Fan, Jian-Ning Zhang.   

Abstract

BACKGROUND AND
PURPOSE: The HydroCoil Embolic System (HES) was developed to reduce recurrences of aneurysms relative to platinum coils. But the HydroCoil Embolic System was characterized with many limitations. The manufacturer had recognized the challenge and recently a new design of hydrogel-coated coil-HydroSoft has become available in the market as the new generation HydroCoil. We reported our initial experience using HydroSoft coil versus HydroCoil in our center.
METHODS: 75 aneurysms embolized primarily using HydroSoft Coils from July 2008 to May 2009 were compared with 66 volume- and shape-matched aneurysms treated with HydroCoils from March 2006 to August 2008. Outcome measures included length and number of coils used, contrast volume, and length of hospital stay. During embolization, a stable framework was first established with bare coils, and hydrogel-coated coils were used subsequently to increase the packing density. Follow-up angiographic results 6 months after treatment were evaluated among some of the patients.
RESULTS: Successful coil embolization was achieved in all patients. There were no differences in average total coil length used per aneurysm. There were no differences in length of hospital stay and packing density. HydroSoft coils were more suitable using as the finishing or final coil. HydroSoft coil decreased the procedure-related retreated rates, and aneurysm packing was finished with soft, flexible HydroSoft coil and decreased the neck remnant rates. Follow-up angiography in HydroSoft-treated patients at 6 months revealed aneurysm stability without significant residual neck.
CONCLUSIONS: HydroSoft coil allowed us to deploy coated coils with good packing density. A slight expansion of these coils at the neck can be expected to reduce neck remnant and potentially inhibit recurrence.
Copyright © 2010 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20537831     DOI: 10.1016/j.ejrad.2010.04.031

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  11 in total

1.  GREAT-a randomized aneurysm trial. Design of a randomized controlled multicenter study comparing HydroSoft/HydroFrame and bare platinum coils for endovascular aneurysm treatment.

Authors:  Christian Taschner; René Chapot; Vincent Costalat; Patrick Courthéoux; Xavier Barreau; Jerome Berge; Laurent Pierot; Kryzsztof Kadziolka; Betty Jean; Raphael Blanc; Alessandra Biondi; Hervé Brunel; Sophie Gallas; Ansgar Berlis; Denis Herbreteau; Joachim Berkefeld; Christoph Groden; Horst Urbach; Samer El Shikh; Erika Graf; Alain Bonafé
Journal:  Neuroradiology       Date:  2015-02-28       Impact factor: 2.804

2.  Aneurysm Recurrence Volumetry Is More Sensitive than Visual Evaluation of Aneurysm Recurrences.

Authors:  M H Schönfeld; V Schlotfeldt; N D Forkert; E Goebell; M Groth; E Vettorazzi; Y D Cho; M H Han; H-S Kang; J Fiehler
Journal:  Clin Neuroradiol       Date:  2014-08-27       Impact factor: 3.649

3.  Embolization of Ruptured Aneurysm Arising From Basilar Artery Fenestration Using Hydrocoils.

Authors:  Karuna Tamrakar; Duan Chuan Zhi
Journal:  Cureus       Date:  2015-09-18

4.  Embolization of intracranial aneurysms with HydroSoft coils: results of the Korean multicenter study.

Authors:  J H Park; H-S Kang; M H Han; P Jeon; D-S Yoo; T H Lee
Journal:  AJNR Am J Neuroradiol       Date:  2011-09-15       Impact factor: 3.825

5.  Endovascular treatment of cerebral aneurysms using the hydrocoil embolic system.

Authors:  F Briganti; G Leone; K Panagiotopoulos; M Marseglia; G Mariniello; M Napoli; F Caranci
Journal:  Neuroradiol J       Date:  2013-08-27

Review 6.  Design and biocompatibility of endovascular aneurysm filling devices.

Authors:  Jennifer N Rodriguez; Wonjun Hwang; John Horn; Todd L Landsman; Anthony Boyle; Mark A Wierzbicki; Sayyeda M Hasan; Douglas Follmer; Jesse Bryant; Ward Small; Duncan J Maitland
Journal:  J Biomed Mater Res A       Date:  2014-08-04       Impact factor: 4.396

7.  New generation Hydrogel Endovascular Aneurysm Treatment Trial (HEAT): a study protocol for a multicenter randomized controlled trial.

Authors:  Karl R Abi-Aad; Rami James N Aoun; Rudy J Rahme; Jennifer D Ward; Jason Kniss; Mary Jeanne Kwasny; Mithun G Sattur; Matthew E Welz; Bernard R Bendok
Journal:  Neuroradiology       Date:  2018-08-17       Impact factor: 2.804

Review 8.  Shape Memory Polymer-Based Endovascular Devices: Design Criteria and Future Perspective.

Authors:  Sergio A Pineda-Castillo; Aryn M Stiles; Bradley N Bohnstedt; Hyowon Lee; Yingtao Liu; Chung-Hao Lee
Journal:  Polymers (Basel)       Date:  2022-06-21       Impact factor: 4.967

9.  GREAT-a randomized controlled trial comparing HydroSoft/HydroFrame and bare platinum coils for endovascular aneurysm treatment: procedural safety and core-lab-assessedangiographic results.

Authors:  Christian A Taschner; René Chapot; Vincent Costalat; Paolo Machi; Patrick Courthéoux; Xavier Barreau; Jérôme Berge; Laurent Pierot; Kryzsztof Kadziolka; Betty Jean; Raphaël Blanc; Alessandra Biondi; Hervé Brunel; Sophie Gallas; Ansgar Berlis; Denis Herbreteau; Joachim Berkefeld; Horst Urbach; Samer El Shikh; Jens Fiehler; Hubert Desal; Erika Graf; Alain Bonafé
Journal:  Neuroradiology       Date:  2016-04-30       Impact factor: 2.804

10.  Differences in the angiographic evaluation of coiled cerebral aneurysms between a core laboratory reader and operators: results of the Cerecyte Coil Trial.

Authors:  I Rezek; R K Lingineni; M Sneade; A J Molyneux; A J Fox; D F Kallmes
Journal:  AJNR Am J Neuroradiol       Date:  2013-07-18       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.