Comparison of macular thickness measurement and segmentation error rate between stratus and fourier-domain optical coherence tomography.

Sara Modjtahedi1, Christine Chiou, Bobeck Modjtahedi, David G Telander, Lawrence S Morse, Susanna S Park.   

Abstract

BACKGROUND AND
OBJECTIVE: To compare macular thickness measurements and segmentation error rates between Stratus optical coherence tomography (OCT) (Carl Zeiss Meditec, Inc., Dublin, CA), and Fourier-domain OCT (RTVue, Optovue, Inc., Fremont, CA). PATIENTS AND METHODS: A retrospective study was performed of 93 normal and pathologic eyes from 79 subjects imaged with both OCT instruments on the same day. Both the macular thickness measurement for each Early Treatment Diabetic Retinopathy Study (ETDRS) zone and the incidence of segmentation error in the central macula between the two instruments were compared.
RESULTS: Macular thickness measurements for all nine ETDRS zones were higher with RTVue compared with Stratus OCT (P < .01). Linear regression analysis showed the highest correlation in the central macula (R(2) = 0.88), with progressively lower correlation peripherally. The overall segmentation error rate was 29% with Stratus OCT versus 32% with RTVue (P > .05).
CONCLUSION: Macular thickness measurement was greater with RTVue than with Stratus OCT in all ETDRS areas, with the best correlation seen in the central macula. No difference in segmentation error rate was noted between the two instruments. Copyright 2010, SLACK Incorporated.

Entities:  

Mesh:

Year:  2010        PMID: 20507013     DOI: 10.3928/15428877-20100430-02

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging        ISSN: 1542-8877


  3 in total

1.  Foveal and parafoveal choroidal thickness pattern measuring by swept source optical coherence tomography.

Authors:  Pear Pongsachareonnont; Thanapong Somkijrungroj; Buravej Assavapongpaiboon; Theerada Chitamara; Maytavee Chuntarapas; Disorn Suwajanakorn
Journal:  Eye (Lond)       Date:  2019-04-08       Impact factor: 3.775

2.  Reproducibility of Optovue RTVue Optical Coherence Tomography Retinal Thickness Measurements and Conversion to Equivalent Zeiss Stratus Metrics in Diabetic Macular Edema.

Authors:  Susan B Bressler; Allison R Edwards; Christopher M Andreoli; Paul Andrew Edwards; Adam R Glassman; Glenn J Jaffe; Michele Melia; Jennifer K Sun
Journal:  Transl Vis Sci Technol       Date:  2015-01-22       Impact factor: 3.283

3.  Comparison of macular thickness in diabetic macular edema using spectral-domain optical coherence tomography and time-domain optical coherence tomography.

Authors:  Masashi Kakinoki; Taichirou Miyake; Osamu Sawada; Tomoko Sawada; Hajime Kawamura; Masahito Ohji
Journal:  J Ophthalmol       Date:  2012-01-18       Impact factor: 1.909

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.