Estimation of environmental force for the haptic interface of robotic surgery.

Hyoung Il Son1, Tapomayukh Bhattacharjee, Doo Yong Lee.   

Abstract

BACKGROUND: The success of a telerobotic surgery system with haptic feedback requires accurate force-tracking and position-tracking capacity of the slave robot. The two-channel force-position control architecture is widely used in teleoperation systems with haptic feedback for its better force-tracking characteristics and superior position-tracking capacity for the maximum stability margin. This control architecture, however, requires force sensors at the end-effector of the slave robot to measure the environment force. However, it is difficult to attach force sensors to slave robots, mainly due to their large size, insulation issues and also large currents often flowing through the end-effector for incision or cautery of tissues.
METHODS: This paper provides a method to estimate the environment force, using a function parameter matrix and a recursive least-squares method. The estimated force is used to feed back the force information to the surgeon through the control architecture without involving the force sensors.
RESULTS: The simulation and experimental results verify the efficacy of the proposed method. The force estimation error is negligible and the slave device successfully tracks the position of the master device while the stability of the teleoperation system is maintained.
CONCLUSIONS: The developed method allows practical haptic feedback for telerobotic surgery systems in the two-channel force-position control scheme without the direct employment of force sensors at the end-effector of the slave robot. (c) 2010 John Wiley & Sons, Ltd.

Mesh:

Year:  2010        PMID: 20506442     DOI: 10.1002/rcs.311

Source DB:  PubMed          Journal:  Int J Med Robot        ISSN: 1478-5951            Impact factor:   2.547


  3 in total

1.  Grasper having tactile sensing function using acoustic reflection for laparoscopic surgery.

Authors:  Hiep Hoang Ly; Yoshihiro Tanaka; Tomohiro Fukuda; Akihito Sano
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-04-28       Impact factor: 2.924

Review 2.  Prevalence of haptic feedback in robot-mediated surgery: a systematic review of literature.

Authors:  Farshid Amirabdollahian; Salvatore Livatino; Behrad Vahedi; Radhika Gudipati; Patrick Sheen; Shan Gawrie-Mohan; Nikhil Vasdev
Journal:  J Robot Surg       Date:  2017-12-01

3.  Electrically Elicited Force Response Characteristics of Forearm Extensor Muscles for Electrical Muscle Stimulation-Based Haptic Rendering.

Authors:  Jungeun Lee; Yeongjin Kim; Hoeryong Jung
Journal:  Sensors (Basel)       Date:  2020-10-04       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.