Speckle tracking in a phantom and feature-based tracking in liver in the presence of respiratory motion using 4D ultrasound.

Emma J Harris1, Naomi R Miller, Jeffrey C Bamber, J Richard N Symonds-Tayler, Philip M Evans.   

Abstract

We have evaluated a 4D ultrasound-based motion tracking system developed for tracking of abdominal organs during therapy. Tracking accuracy and precision were determined using a tissue-mimicking phantom, by comparing tracked motion with known 3D sinusoidal motion. The feasibility of tracking 3D liver motion in vivo was evaluated by acquiring 4D ultrasound data from four healthy volunteers. For two of these volunteers, data were also acquired whilst simultaneously measuring breath flow using a spirometer. Hepatic blood vessels, tracked off-line using manual tracking, were used as a reference to assess, in vivo, two types of automated tracking algorithm: incremental (from one volume to the next) and non-incremental (from the first volume to each subsequent volume). For phantom-based experiments, accuracy and precision (RMS error and SD) were found to be 0.78 mm and 0.54 mm, respectively. For in vivo measurements, mean absolute distance and standard deviation of the difference between automatically and manually tracked displacements were less than 1.7 mm and 1 mm respectively in all directions (left-right, anterior-posterior and superior-inferior). In vivo non-incremental tracking gave the best agreement. In both phantom and in vivo experiments, tracking performance was poorest for the elevational component of 3D motion. Good agreement between automatically and manually tracked displacements indicates that 4D ultrasound-based motion tracking has potential for image guidance applications in therapy.

Entities:  

Mesh:

Year:  2010        PMID: 20505224     DOI: 10.1088/0031-9155/55/12/007

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  21 in total

1.  Monte Carlo modeling of ultrasound probes for image guided radiotherapy.

Authors:  Magdalena Bazalova-Carter; Jeffrey Schlosser; Josephine Chen; Dimitre Hristov
Journal:  Med Phys       Date:  2015-10       Impact factor: 4.071

2.  In vivo reproducibility of robotic probe placement for a novel ultrasound-guided radiation therapy system.

Authors:  Muyinatu A Lediju Bell; H Tutkun Sen; Iulian Iordachita; Peter Kazanzides; John Wong
Journal:  J Med Imaging (Bellingham)       Date:  2014-07-23

3.  System Integration and In Vivo Testing of a Robot for Ultrasound Guidance and Monitoring During Radiotherapy.

Authors:  Hasan Tutkun Sen; Muyinatu A Lediju Bell; Yin Zhang; Kai Ding; Emad Boctor; John Wong; Iulian Iordachita; Peter Kazanzides
Journal:  IEEE Trans Biomed Eng       Date:  2016-10-03       Impact factor: 4.538

4.  A 4D ultrasound real-time tracking system for external beam radiotherapy of upper abdominal lesions under breath-hold.

Authors:  Dwi Seno Kuncoro Sihono; Lena Vogel; Christel Weiß; Johannes Thölking; Frederik Wenz; Frank Lohr; Judit Boda-Heggemann; Hansjörg Wertz
Journal:  Strahlenther Onkol       Date:  2016-12-05       Impact factor: 3.621

5.  Feasibility of Swept Synthetic Aperture Ultrasound Imaging.

Authors:  Nick Bottenus; Will Long; Haichong K Zhang; Marko Jakovljevic; David P Bradway; Emad M Boctor; Gregg E Trahey
Journal:  IEEE Trans Med Imaging       Date:  2016-02-03       Impact factor: 10.048

6.  System Integration and Preliminary In-Vivo Experiments of a Robot for Ultrasound Guidance and Monitoring during Radiotherapy.

Authors:  H Tutkun Şen; Muyinatu A Lediju Bell; Yin Zhang; Kai Ding; John Wong; Iulian Iordachita; Peter Kazanzides
Journal:  Proc Int Conf Adv Robot       Date:  2015-07

7.  A block matching based approach with multiple simultaneous templates for the real-time 2D ultrasound tracking of liver vessels.

Authors:  Andrew J Shepard; Bo Wang; Thomas K F Foo; Bryan P Bednarz
Journal:  Med Phys       Date:  2017-10-13       Impact factor: 4.071

Review 8.  Particle therapy of moving targets-the strategies for tumour motion monitoring and moving targets irradiation.

Authors:  Tomasz Kubiak
Journal:  Br J Radiol       Date:  2016-07-19       Impact factor: 3.039

9.  3D dosimetric validation of ultrasound-guided radiotherapy with a dynamically deformable abdominal phantom.

Authors:  Charles K Matrosic; Wesley Culberson; Andrew Shepard; Sydney Jupitz; Bryan Bednarz
Journal:  Phys Med       Date:  2021-04-23       Impact factor: 2.685

10.  Optofluidic phantom mimicking optical properties of porcine livers.

Authors:  Ruiqi Long; Travis King; Tony Akl; M Nance Ericson; Mark Wilson; Gerard L Coté; Michael J McShane
Journal:  Biomed Opt Express       Date:  2011-06-09       Impact factor: 3.732

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.