| Literature DB >> 20495677 |
Domingos D Afonso1, Sitki Baytak, Zikri Arslan.
Abstract
Performance of potassium ferricyanide, K(3)(Fe(CN)(6), for simultaneous generation of hydrides of Bi, Pb and Sn in dilute HCl is investigated for determination by ICP-AES. On-line addition of K(3)Fe(CN)(6) to sample solution was essential to achieve optimum signals and stability in generation of BiH(3) and SnH(4). Off-line addition caused instability for Bi(III) and Sn(IV) that resulted in substantial loss in hydride generation efficiency within 24 h. Lead hydride (PbH(4)) generation, however, was not influenced from on-line or off-line addition of [Fe(CN)(6)](3-), nor did it show any instability under the same conditions indicating that [Fe(CN)(6)](3-) affects generation of PbH(4) differently from those of BiH(3) and SnH(4). The effects of transition metals and hydride forming elements were not significant, except Cr(VI) and Cu(II) that suppressed the signals of Bi and Sn, and Pb, respectively, at and above 1.0 μg mL(-1). The detection limits (3s, n = 11) were 0.20, 0.13 and 0.10 μg L(-1) for Bi, Pb and Sn, respectively. The method was applied to the analysis of calcium-rich biominerals - fish otoliths and NIST bone ash certified reference material (SRM 1400).Entities:
Year: 2010 PMID: 20495677 PMCID: PMC2873226 DOI: 10.1039/b920280c
Source DB: PubMed Journal: J Anal At Spectrom ISSN: 0267-9477 Impact factor: 4.023