Reconsideration of patient-prosthesis mismatch definition from the valve indexed effective orifice area.

Yoshimasa Sakamoto1, Michio Yoshitake, Hirokuni Naganuma, Noriyasu Kawada, Katsushi Kinouchi, Kazuhiro Hashimoto.   

Abstract

BACKGROUND: The objective of this study was to reassess the validity of defining patient-prosthesis mismatch (PPM) in the aortic position on the basis of an indexed effective orifice area (iEOA) less than 0.85 cm(2)/m(2).
METHODS: From June 1996 to March 2008, 342 patients underwent aortic valve replacement with a Carpentier-Edwards Perimount valve. From the data collected, the transvalvular pressure gradient was determined by the modified Bernoulli equation, and EOA was calculated from the standard continuity equation.
RESULTS: The actuarial survival rate at 10 years after surgery was 84.0% +/- 8.2%. The prevalence of PPM was 6.1% when a projected iEOA less than 0.85 cm(2)/m(2) was defined as indicating significant PPM. There was no difference between patients with moderate PPM (85.2% +/- 9.8%) and patients without PPM (81.0% +/- 8.7%; p = 0.44). The relation between mean transvalvular pressure gradient and iEOA demonstrated a gentler slope than that reported previously. Postoperative mean transvalvular pressure gradient was 17.4 +/- 5.6 mm Hg and 14.5 +/- 5.6 mm Hg in patients with an iEOA less than 0.85 and 0.85 or greater, respectively. Most patients had a postoperative mean transvalvular pressure gradient more than 10 mm Hg regardless of PPM.
CONCLUSIONS: Our analysis suggested that an iEOA less than 2.0 cm(2)/m(2) might be the threshold for PPM, which should not be passed to achieve a low mean transvalvular pressure gradient (less than 10 mm Hg) with the Carpentier-Edwards Perimount valve. The implications of these findings include the necessity for reassessing the hemodynamic performance of each type of prosthesis when attempting to define PPM, to avoid residual significant transvalvular pressure gradient. 2010 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20494055     DOI: 10.1016/j.athoracsur.2010.03.011

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  2 in total

Review 1.  Update on aortic valve prosthesis-patient mismatch in Japan.

Authors:  Yoshimasa Sakamoto; Kazuhiro Hashimoto
Journal:  Gen Thorac Cardiovasc Surg       Date:  2013-04-13

2.  Is the threshold for postoperative prosthesis-patient mismatch the same for all prostheses?

Authors:  Homare Okamura; Atsushi Yamaguchi; Hideki Morita; Kazuhiro Naito; Koichi Yuri; Hideo Adachi
Journal:  Surg Today       Date:  2012-08-25       Impact factor: 2.549