Clinical experience with a novel thermosensitive temporary coronary artery occluder (LeGoo).

Olivier Bouchot1, Robert L Berger, Jean Pierre Berne, François Brunotte, Roger Brenot.   

Abstract

BACKGROUND: A bloodless field is important for visualizing relevant anatomical structures and facilitating construction of high-quality coronary anastomoses during off-pump coronary artery bypass graft surgery (OPCABG). Currently employed devices can damage target arteries as well as myocardium and are not uniformly effective, especially with diseased arteries.
METHODS: LeGoo (Pluromed, Woburn, MA) is a thermosensitive polymer that is a viscous liquid at room temperatures and changes into a firm occlusive plug instantaneously when exposed to body temperatures. The polymer is nontoxic and neither metabolized nor absorbed. It dissolves spontaneously or can be dissolved at will by cooling. LeGoo was instilled through the arteriotomy of 99 coronaries in 50 patients during construction of OPCABG anastomoses.
RESULTS: LeGoo formed an occlusive endoluminal plug instantaneously and blocked blood flow into the surgical field. In 89 vessels with available information, the quality of bloodless field was satisfactory, without need for substantial use of blood removal systems or application of another device in 81 vessels (91%). Resort to other means was prompted by serious arrhythmia (1 case) epicardial bleeding (1 vessel), bleeding from collateral branches (3 vessels), and shortcomings in design of the injecting catheter (3 vessels) that has been corrected. Intraoperative and perioperative profiles of these patients were similar to those reported from other OPCABG series. Abnormal bleeding, coagulopathy, metabolic derangements, and evidence of LeGoo embolization were not observed.
CONCLUSIONS: In this first large-scale human experience, LeGoo had been an atraumatic, safe, and effective temporary coronary artery occluder. It produced a bloodless field during OPCABG in 91% of target vessels. Improvements on these results with further experience is anticipated. 2010 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20494048     DOI: 10.1016/j.athoracsur.2010.03.015

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  9 in total

1.  Feasibility of bloodless liver resection using Lumagel, a reverse thermoplastic polymer, to produce temporary, targeted hepatic blood flow interruption.

Authors:  James J Pomposelli; Mohamed Akoad; Sebastian Flacke; James J Benn; Mauricio Solano; Aarti Kalra; Peter N Madras
Journal:  HPB (Oxford)       Date:  2011-11-21       Impact factor: 3.647

2.  LeGoo polymer injection into hydatid cyst would eliminate intraoperative spillage of scolices.

Authors:  Fadhil Al-Amran; Hayder M Abdulnabi
Journal:  Gen Thorac Cardiovasc Surg       Date:  2013-02-16

3.  Temporary targeted hemostasis to facilitate bloodless partial nephrectomy.

Authors:  Niall J Harty; Alireza Moinzadeh; Sebastian Flacke; Jeffrey Pettit; James A Benn; John A Libertino; Peter N Madras
Journal:  Urology       Date:  2011-12       Impact factor: 2.649

4.  Temporary targeted renal blood flow interruption using a reverse thermosensitive polymer to facilitate bloodless partial nephrectomy: a swine survival study.

Authors:  Niall J Harty; Daniel H Laskey; Alireza Moinzadeh; Sebastian Flacke; James A Benn; Rosanna Villani; Aarti Kalra; John A Libertino; Peter N Madras
Journal:  BJU Int       Date:  2012-03-15       Impact factor: 5.588

Review 5.  (Macro)molecular self-assembly for hydrogel drug delivery.

Authors:  Matthew J Webber; E Thomas Pashuck
Journal:  Adv Drug Deliv Rev       Date:  2021-01-12       Impact factor: 15.470

6.  Incidence and Predictors of New-Onset Silent Atrial Fibrillation after Coronary Artery Bypass Graft Surgery.

Authors:  Charles Guenancia; Charline Pujos; Frederique Debomy; Ghislain Malapert; Gabriel Laurent; Olivier Bouchot
Journal:  Biomed Res Int       Date:  2015-07-28       Impact factor: 3.411

7.  Temporary vascular occlusion by rapid reverse phase polymer: a preliminary in vitro study of retrograde injection.

Authors:  Einar Dregelid
Journal:  Int J Biomater       Date:  2012-07-25

8.  Effects of temporary vascular occluder poloxamer 407 gel on the endothelium.

Authors:  Arif Gucu; Ilkin Cavusoglu; Onder Bozkurt; Cuneyt Eris; Faruk Toktas; Tugrul Goncu; Ahmet Ozyazicioglu
Journal:  J Cardiothorac Surg       Date:  2013-01-22       Impact factor: 1.637

9.  Achieving hemostasis in dermatology-Part II: Topical hemostatic agents.

Authors:  Jaimie B Glick; Ravneet R Kaur; Daniel Siegel
Journal:  Indian Dermatol Online J       Date:  2013-07