Comparison of ultrasonic Doppler flow monitor, oscillometric, and direct arterial blood pressure measurements in ill dogs.

Ann P Bosiack1, F A Mann, John R Dodam, Colette C Wagner-Mann, Keith R Branson.   

Abstract

OBJECTIVE: To compare blood pressure measurements obtained via ultrasonic Doppler flow monitor (DOP) and 2 oscillometric noninvasive blood pressure monitors (CAR and PAS) to invasive blood pressure (IBP) in hospitalized, conscious dogs with a range of blood pressures.
DESIGN: Prospective clinical study.
SETTING: University teaching hospital. ANIMALS: Eleven client-owned dogs aged between 4 months and 11.5 years (median 6 y), and weighing between 5.8 and 37.5 kg (median 30.2 kg).
INTERVENTIONS: Blood pressure measurement.
MEASUREMENTS AND MAIN RESULTS: Three consecutive measurements of systolic, diastolic, and mean arterial pressure (MAP) were recorded for each of the 3 indirect devices (only systolic for DOP), along with concurrent IBP measurements. The data were categorized into 3 groups: hypotensive (direct MAP<80 mm Hg), normotensive (80 mm Hg<or=direct MAP>or=100 mm Hg), and hypertensive (direct MAP>100 mm Hg). Each indirect method was compared with the corresponding direct arterial pressure using the Bland-Altman method. Within the hypotensive group, each indirect method overestimated the corresponding IBP. Within the normotensive group all indirect systolic measurements and the PAS diastolic measurements underestimated the corresponding IBP. The remaining indirect measurements overestimated the corresponding IBP. Within the hypertensive group, DOP and CAR systolic measurements underestimated the corresponding IBP, and the remaining indirect measurements overestimated the corresponding IBP. In hypertensive dogs oscillometric systolic measurements were more accurate than MAP. In hypotensive dogs MAP measurements were more accurate than systolic measurements. All indirect measurements were most accurate in hypertensive dogs.
CONCLUSIONS: The noninvasive blood pressure monitors in our study did not meet the validation standards set in human medicine. However, CAR diastolic and MAP measurements within the normotensive group, CAR MAP measurements within the hypertensive group, and PAS diastolic measurements in all groups were close to these standards. All indirect measurements showed greater bias during hypotension. Precision was poorer for all indirect systolic measurements than for MAP.

Entities:  

Mesh:

Year:  2010        PMID: 20487248     DOI: 10.1111/j.1476-4431.2010.00520.x

Source DB:  PubMed          Journal:  J Vet Emerg Crit Care (San Antonio)        ISSN: 1476-4431


  12 in total

1.  Evaluation of surgical outcome, complications, and mortality in dogs undergoing preoperative computed tomography angiography for diagnosis of an extrahepatic portosystemic shunt: 124 cases (2005-2014).

Authors:  Benjamin W Brunson; J Brad Case; Gary W Ellison; W Alexander Fox-Alvarez; Stanley E Kim; Matthew Winter; Fernando L Garcia-Pereira; Lisa L Farina
Journal:  Can Vet J       Date:  2016-01       Impact factor: 1.008

2.  Comparison of invasive and oscillometric blood pressure measurement techniques in anesthetized camelids.

Authors:  Turi K Aarnes; John A E Hubbell; Phillip Lerche; Richard M Bednarski
Journal:  Can Vet J       Date:  2012-08       Impact factor: 1.008

3.  Evaluation of the shock index in dogs presenting as emergencies.

Authors:  Adam E Porter; Elizabeth A Rozanski; Claire R Sharp; Kursten L Dixon; Lori Lyn Price; Scott P Shaw
Journal:  J Vet Emerg Crit Care (San Antonio)       Date:  2013-07-15

4.  QUANTITATIVE PLANAR AND VOLUMETRIC CARDIAC MEASUREMENTS USING 64 MDCT AND 3T MRI VS. STANDARD 2D AND M-MODE ECHOCARDIOGRAPHY: DOES ANESTHETIC PROTOCOL MATTER?

Authors:  Randi Drees; Rebecca A Johnson; Rebecca L Stepien; Alejandro Munoz Del Rio; Jimmy H Saunders; Christopher J François
Journal:  Vet Radiol Ultrasound       Date:  2015-06-16       Impact factor: 1.363

5.  Accuracy and Precision of Noninvasive Blood Pressure in Normo-, Hyper-, and Hypotensive Standing and Anesthetized Adult Horses.

Authors:  N Heliczer; O Lorello; D Casoni; C Navas de Solis
Journal:  J Vet Intern Med       Date:  2016-04-05       Impact factor: 3.333

6.  Effects of various factors on Doppler flow ultrasonic radial and coccygeal artery systolic blood pressure measurements in privately-owned, conscious dogs.

Authors:  Allison P Mooney; Dianne I Mawby; Joshua M Price; Jacqueline C Whittemore
Journal:  PeerJ       Date:  2017-03-22       Impact factor: 2.984

7.  Indirect Doppler flow systolic blood pressure measurements taken with and without headphones in privately-owned, conscious dogs.

Authors:  India R Gill; Joshua M Price; Jacqueline C Whittemore
Journal:  PeerJ       Date:  2019-08-01       Impact factor: 2.984

Review 8.  Review of oncological emergencies in small animal patients.

Authors:  Katrina L Tumielewicz; Danielle Hudak; Jennifer Kim; David W Hunley; Lisa A Murphy
Journal:  Vet Med Sci       Date:  2019-03-21

9.  ACVIM consensus statement: Guidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats.

Authors:  Mark J Acierno; Scott Brown; Amanda E Coleman; Rosanne E Jepson; Mark Papich; Rebecca L Stepien; Harriet M Syme
Journal:  J Vet Intern Med       Date:  2018-10-24       Impact factor: 3.333

10.  Hypertension, retinopathy, and acute kidney injury in dogs: A prospective study.

Authors:  Laura Pearl Cole; Rosanne Jepson; Charlotte Dawson; Karen Humm
Journal:  J Vet Intern Med       Date:  2020-07-17       Impact factor: 3.333

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.