Transcutaneous bilirubin--comparing the accuracy of BiliChek® and JM 103® in a regional postnatal unit.

Yvonne M Qualter1, Nicholas M Allen, John D Corcoran, Donough J O'Donovan.   

Abstract

OBJECTIVE: Transcutaneous bilirubin (TcB) has the potential to reduce serum bilirubin sampling. During a recent survey on the use of TcB in postnatal units in the Republic of Ireland, we identified that only 58% of the 19 units were using TcB and that only two devices were in use, the BiliChek® and JM 103®. We aimed to evaluate and compare these two devices in a regional postnatal unit.
METHODS: To evaluate and compare the accuracy of the BiliChek® and JM 103®, we studied simultaneous TcB and total serum bilirubin (TSB) measurements from a population of jaundiced term and near term infants. We evaluated each device with regard to correlation with TSB and potential to safely reduce serum bilirubin testing.
RESULTS: Both TcB devices strongly correlated with TSB (r = 0.88 for BiliChek® and r = 0.70 for JM 103®. The BiliChek® and JM 103® were accurate up to cut-off values of 200 μmol/L and 180 μmol/L, respectively. Using Bhutani's nomogram, 100% sensitivity was achieved using the 75th percentile for BiliChek® and the 40th percentile for JM 103®.
CONCLUSION: Both TcB devices correlated closely with moderately increased TSB levels and are suitable screening tools to identify jaundiced infants that require a serum bilirubin, with upper limit cut-off values. Both devices reduced the need for TSB levels. We found the BiliChek® slightly more accurate than the JM 103® for our study population. TcB however, is not in widespread use.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20482289     DOI: 10.3109/14767058.2010.484471

Source DB:  PubMed          Journal:  J Matern Fetal Neonatal Med        ISSN: 1476-4954


  3 in total

1.  Defining the limitations of transcutaneous bilirubin measurement in late preterm newborns.

Authors:  K L Fine; W A Carey; J A W Schuster; S C Bryant; W J Cook; B S Karon
Journal:  J Perinatol       Date:  2017-02-16       Impact factor: 2.521

2.  Measuring transcutaneous bilirubin: a comparative analysis of three devices on a multiracial population.

Authors:  Francesco Raimondi; Silvia Lama; Francesca Landolfo; Maria Sellitto; Angela Carla Borrelli; Rosalba Maffucci; Paola Milite; Letizia Capasso
Journal:  BMC Pediatr       Date:  2012-06-14       Impact factor: 2.125

3.  Quality improvement initiative using transcutaneous bilirubin nomogram to decrease serum bilirubin sampling in low-risk babies.

Authors:  Muhammad Hussain Shah; Shabina Ariff; Syed Rehan Ali; Rayaan Asad Chaudhry; Maryam Pyar Ali Lakhdir; Fatima Qaiser; Simon Demas; Ali Shabbir Hussain
Journal:  BMJ Paediatr Open       Date:  2019-04-24
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.