Literature DB >> 20444425

Analytical performance of a sensitive assay for cardiac troponin I with loci technology.

M M Arrebola1, J A Lillo, M J Diez De Los Ríos, M Rodríguez, A Dayaldasani, R Yahyaoui, V Pérez.   

Abstract

OBJECTIVES: To confirm the analytical performance of the Dimension Vista LOCI troponin I assay (cTnI). DESIGN AND
METHOD: Limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ) with a 10% coefficient of variation (CV), linearity, precision, method comparison, and 99th percentile upper reference limits (URL) were analyzed. Endogenous analytes and rheumatoid factor (RF) were tested for assay interference.
RESULTS: The 99th percentile was 0.022 microg/L (CV=14%) and the LoQ was 0.036 microg/L. The ratio of 10% CV concentration to 99th percentile was 1.63. Linearity extended from 0 to 44.36 microg/L. The method comparison equation was Dimension(R) Vista=0.94 (Dimension RxL)+0.00 microg/L with bias at low levels. No interference was detected.
CONCLUSIONS: This study shows acceptable performance characteristics of the LOCI cTnI assay on Dimension Vista to diagnosis and risk stratification of patients with acute coronary syndrome symptoms. 2010. Published by Elsevier Inc.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20444425     DOI: 10.1016/j.clinbiochem.2010.04.073

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  7 in total

1.  External evaluation of the Dimension Vista 1500® intelligent lab system.

Authors:  Arnaud Bruneel; Monique Dehoux; Anne Barnier; Anne Boutten
Journal:  J Clin Lab Anal       Date:  2012-09       Impact factor: 2.352

2.  High-density lipoprotein cholesterol values independently and inversely predict cardiac troponin T and I concentration.

Authors:  Giuseppe Lippi; Claudia Lo Cascio; Giorgio Brocco; Elisa Danese; Martina Montagnana; Antonella Bassi; Beatrice Caruso; Chiara Bovo; Gian Luca Salvagno
Journal:  Ann Transl Med       Date:  2016-05

3.  Performance characteristics of loci method for measuring cardiac troponin I on the dimension EXL.

Authors:  José L Martín Calderón; Julia M Varona Pérez; Fernando Bustos; Juan C Sánchez Gómez
Journal:  Pract Lab Med       Date:  2015-04-08

4.  Type 2 Myocardial Infarction: A Geriatric Population-based Model of Pathogenesis.

Authors:  Alain Putot; Melanie Jeanmichel; Frederic Chague; Patrick Manckoundia; Yves Cottin; Marianne Zeller
Journal:  Aging Dis       Date:  2020-02-01       Impact factor: 6.745

5.  Troponin Elevation in Older Patients with Acute Pneumonia: Frequency and Prognostic Value.

Authors:  Alain Putot; Emmanuel Bouhey; Jennifer Tetu; Jérémy Barben; Eléonore Timsit; Sophie Putot; Patrick Ray; Patrick Manckoundia
Journal:  J Clin Med       Date:  2020-11-10       Impact factor: 4.241

6.  A Case of Elevated Troponin I Level After Packed Red Blood Cell Transfusion With Normal Coronary Angiography.

Authors:  Sherif Roman; Manpreet Sran; Amgad N Makaryus
Journal:  Cureus       Date:  2022-06-22

7.  The importance of having a flexible scope ISO 15189 accreditation and quality specifications based on biological variation--the case of validation of the biochemistry analyzer Dimension Vista.

Authors:  Pilar Fernandez-Calle; Sandra Pelaz; Paloma Oliver; Maria Jose Alcaide; Ruben Gomez-Rioja; Antonio Buno; Jose Manuel Iturzaeta
Journal:  Biochem Med (Zagreb)       Date:  2013       Impact factor: 2.313

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.