Feasibility of in vivo pressure measurement using a pressure-tip catheter via transventricular puncture.

Robert G Gray1, Santos E Cabreriza, T Alexander Quinn, Alan D Weinberg, Henry M Spotnitz.   

Abstract

Pressure-tip catheters (PTCs) are used to evaluate ventricular mechanics during surgical repair of congenital heart disease in children. Studies in infants require miniaturized sensors. We compared the safety and accuracy of a 2-Fr ultraminiature PTC with a 5-Fr PTC. In 10 piglets (weight 19-22 kg), a 5-Fr PTC was inserted through a 3-mm apical puncture with a #11 blade. A 20-gauge angiocatheter was inserted using a separate site. A 2-Fr PTC was threaded through the angiocatheter lumen. The angiocatheter was withdrawn, leaving the 2-Fr PTC within the left ventricle (LV). Left ventricular pressure (LVP) changes were measured during three inferior vena caval occlusions. Reliability coefficients demonstrated correlation between the 2-Fr PTC and 5-Fr PTC for LV end-diastolic pressure (0.90-0.95), peak LVP (0.92-0.99), and the maximal (0.87-0.93) and minimal (0.89-0.94) first derivatives of LVP. Bland-Altman analysis demonstrated agreement for all variables. Blood loss was trivial with pressure manipulation and catheter placement and removal. Pressure measurements using the 2-Fr PTC were accurate and comparable with those from the 5-Fr PTC. Transventricular placement of a 2-Fr PTC is feasible and should allow evaluation of ventricular mechanics during surgical repair of congenital heart disease.

Entities:  

Mesh:

Year:  2010        PMID: 20400891      PMCID: PMC2874157          DOI: 10.1097/MAT.0b013e3181d823a1

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  24 in total

1.  Development of a new catheter-tip pressure transducer.

Authors:  Y Kinefuchi; H Fukuyama; T Suzuki; M Kanazawa; M Takiguchi
Journal:  Tokai J Exp Clin Med       Date:  1999-10

2.  Contraction wave in axial direction in free wall of guinea pig left ventricle.

Authors:  Borut Kirn; Vito Starc
Journal:  Am J Physiol Heart Circ Physiol       Date:  2004-04-01       Impact factor: 4.733

3.  Transmitral pressure-flow velocity relation. Importance of regional pressure gradients in the left ventricle during diastole.

Authors:  M Courtois; S J Kovács; P A Ludbrook
Journal:  Circulation       Date:  1988-09       Impact factor: 29.690

4.  A stable ultraminiature catheter-tip pressure transducer.

Authors:  H D Millar; L E Baker
Journal:  Med Biol Eng       Date:  1973-01

5.  In vivo comparison of catheter manometer systems with the catheter-tip micromanometer.

Authors:  K L Gould; S Trenholme; J W Kennedy
Journal:  J Appl Physiol       Date:  1973-02       Impact factor: 3.531

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Measuring systolic arterial blood pressure. Possible errors from extension tubes or disposable transducer domes.

Authors:  C F Rothe; K C Kim
Journal:  Crit Care Med       Date:  1980-11       Impact factor: 7.598

8.  Combined fluid-filled and micromanometer-tip catheter system for high-fidelity pressure recordings in infants.

Authors:  S D Colan
Journal:  Cathet Cardiovasc Diagn       Date:  1984

9.  In vivo adenoviral transfer of sorcin reverses cardiac contractile abnormalities of diabetic cardiomyopathy.

Authors:  Jorge Suarez; Darrell D Belke; Bernd Gloss; Thomas Dieterle; Patrick M McDonough; Yun-Kyung Kim; Laurence L Brunton; Wolfgang H Dillmann
Journal:  Am J Physiol Heart Circ Physiol       Date:  2003-09-04       Impact factor: 4.733

10.  Ventricular mechanics in the bidirectional glenn procedure and total cavopulmonary connection.

Authors:  Yoshihisa Tanoue; Akira Sese; Yutaka Imoto; Kunitaka Joh
Journal:  Ann Thorac Surg       Date:  2003-08       Impact factor: 4.330

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.