Backprojection-filtration reconstruction without invoking a spatially varying weighting factor.

Dan Xia1, Seungryong Cho, Xiaochuan Pan.   

Abstract

PURPOSE: To develop a backprojection-filtration (BPF) algorithm with improved noise properties over the existing BPF algorithm through utilizing (approximate) redundant information in circular cone-beam or fan-beam scans.
METHODS: The backprojection steps in the existing filtered-backprojection (FBP) and BPF algorithms for fan-beam and cone-beam projections invoke spatially varying weighting factors, which may not only increase the computational load in image reconstruction but also, more importantly, result in reconstruction artifacts. Redundant information in fan-beam projections has been exploited for eliminating the weighting factor in the existing FBP algorithm. However, the new FBP algorithm cannot be applied to image reconstruction in a region of interest from transversely truncated data. In this work, the authors identify approximate data redundancy in circular cone-beam projections and propose a new BPF algorithm in which the approximate data redundancy is exploited for eliminating the spatially varying weighting factor in the existing BPF algorithm.
RESULTS: The authors have implemented and evaluated the proposed BPF algorithm in numerical studies of reconstructing 3D images from both the nontruncated and truncated projection data in a circular cone-beam scan. The results of numerical studies demonstrate that the proposed BPF algorithm retains the resolution property of the existing BPF algorithm, and that it can also reconstruct accurately ROI images from truncated data. More importantly, the results also indicate that the proposed BPF algorithm not only is computationally more efficient but also yields generally lower image variances than the existing BPF algorithm.
CONCLUSIONS: A BPF algorithm was proposed that not only retains the desirable properties of the existing BPF algorithm but also possesses improved computational and noise properties over the latter.

Mesh:

Year:  2010        PMID: 20384257      PMCID: PMC2837731          DOI: 10.1118/1.3285041

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  Exact helical reconstruction using native cone-beam geometries.

Authors:  Frédéric Noo; Jed Pack; Dominic Heuscher
Journal:  Phys Med Biol       Date:  2003-12-07       Impact factor: 3.609

2.  Flat-panel cone-beam computed tomography for image-guided radiation therapy.

Authors:  David A Jaffray; Jeffrey H Siewerdsen; John W Wong; Alvaro A Martinez
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-08-01       Impact factor: 7.038

3.  Physical performance evaluation of a 256-slice CT-scanner for four-dimensional imaging.

Authors:  Shinichiro Mori; Masahiro Endo; Takanori Tsunoo; Susumu Kandatsu; Shuji Tanada; Hiroshi Aradate; Yasuo Saito; Hiroaki Miyazaki; Kazumasa Satoh; Satoshi Matsushita; Masahiro Kusakabe
Journal:  Med Phys       Date:  2004-06       Impact factor: 4.071

4.  General reconstruction theory for multislice X-ray computed tomography with a gantry tilt.

Authors:  Frédéric Noo; Michel Defrise; Hiroyuki Kudo
Journal:  IEEE Trans Med Imaging       Date:  2004-09       Impact factor: 10.048

5.  Exact image reconstruction on PI-lines from minimum data in helical cone-beam CT.

Authors:  Yu Zou; Xiaochuan Pan
Journal:  Phys Med Biol       Date:  2004-03-21       Impact factor: 3.609

6.  Image reconstruction for the circle and line trajectory.

Authors:  Alexander Katsevich
Journal:  Phys Med Biol       Date:  2004-11-21       Impact factor: 3.609

7.  A unified analysis of FBP-based algorithms in helical cone-beam and circular cone- and fan-beam scans.

Authors:  Xiaochuan Pan; Dan Xia; Yu Zou; Lifeng Yu
Journal:  Phys Med Biol       Date:  2004-09-21       Impact factor: 3.609

8.  A factorization approach for cone-beam reconstruction on a circular short-scan.

Authors:  Frank Dennerlein; Frédéric Noo; Harald Schöndube; Günter Lauritsch; Joachim Hornegger
Journal:  IEEE Trans Med Imaging       Date:  2008       Impact factor: 10.048

9.  Nonstationary noise characteristics of the helical scan and its impact on image quality and artifacts.

Authors:  J Hsieh
Journal:  Med Phys       Date:  1997-09       Impact factor: 4.071

10.  Fundamental image quality limits for microcomputed tomography in small animals.

Authors:  N L Ford; M M Thornton; D W Holdsworth
Journal:  Med Phys       Date:  2003-11       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.