Improvements in critical dosimetric endpoints using the Contura multilumen balloon breast brachytherapy catheter to deliver accelerated partial breast irradiation: preliminary dosimetric findings of a phase iv trial.

Douglas W Arthur1, Frank A Vicini, Dorin A Todor, Thomas B Julian, Maureen R Lyden.   

Abstract

PURPOSE: Dosimetric findings in patients treated with the Contura multilumen balloon (MLB) breast brachytherapy catheter to deliver accelerated partial breast irradiation (APBI) on a multi-institutional Phase IV registry trial are presented. METHODS AND MATERIALS: Computed tomography-based three-dimensional planning with dose optimization was performed. For the trial, new ideal dosimetric goals included (1) ≥95% of the prescribed dose (PD) covering ≥90% of the target volume, (2) a maximum skin dose ≤125% of the PD, (3) maximum rib dose ≤145% of the PD, and (4) the V150 ≤50 cc and V200 ≤10 cc. The ability to concurrently achieve these dosimetric goals using the Contura MLB was analyzed.
RESULTS: 144 cases were available for review. Using the MLB, all dosimetric criteria were met in 76% of cases. Evaluating dosimetric criteria individually, 92% and 89% of cases met skin and rib dose criteria, respectively. In 93% of cases, ideal target volume coverage goals were met, and in 99%, dose homogeneity criteria (V150 and V200) were satisfied. When skin thickness was ≥5 mm to <7 mm, the median skin dose was limited to 120.1% of the PD, and when skin thickness was <5 mm, the median skin dose was 124.2%. When rib distance was <5 mm, median rib dose was reduced to 136.5% of the PD. When skin thickness was <7 mm and distance to rib was <5 mm, median skin and rib doses were jointly limited to 120.6% and 142.1% of the PD, respectively.
CONCLUSION: The Contura MLB catheter provided the means of achieving the imposed higher standard of dosimetric goals in the majority of clinical scenarios encountered.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20378268     DOI: 10.1016/j.ijrobp.2009.10.025

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  6 in total

1.  A PHASE II TRIAL OF BALLOON-CATHETER PARTIAL BREAST BRACHYTHERAPY OPTIMIZATION IN THE TREATMENT OF STAGE 0, I AND IIA BREAST CARCINOMA.

Authors:  Sameer K Nath; Zhe J Chen; Bryan P Rowe; Rachel C Blitzblau; Sanjay Aneja; Baiba J Grube; Nina R Horowitz; Joanne B Weidhaas
Journal:  J Radiat Oncol       Date:  2014-12

2.  Dosimetric Improvements in Balloon Based Brachytherapy Using the Contura® Multi-Lumen Balloon (MLB) Catheter to Deliver Accelerated Partial Breast Irradiation.

Authors:  Frank A Vicini; Arthur Douglas; Dorin Todor; Thomas Julian; Maureen Lyden
Journal:  J Contemp Brachytherapy       Date:  2010-04-01

3.  Dose perturbation study in a multichannel breast brachytherapy device.

Authors:  Eric D Slessinger; Rodney Fletcher; Indra J Das
Journal:  J Contemp Brachytherapy       Date:  2011-12-30

4.  Acute toxicity and early cosmetic outcome in patients treated with multilumen balloon brachytherapy with skin spacing ≤ 7.0 millimeters.

Authors:  Kiwhoon Lee; Amy Quillo; Dustin Dillon; Keith Sowards; Allison Hatmaker; Anthony Dragun
Journal:  J Contemp Brachytherapy       Date:  2012-03-30

5.  Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation?

Authors:  Janusz Skowronek; Grzegorz Bielęda; Piotr Laski; Witold Kycler
Journal:  J Contemp Brachytherapy       Date:  2013-09-20

6.  Dosimetric evaluation of multilumen intracavitary balloon applicator rotation in high-dose-rate brachytherapy for breast cancer.

Authors:  Yongbok Kim; Mark G Trombetta
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.