| Literature DB >> 20376884 |
Ung Kim1, Jung-Sun Kim, Jin-Sun Kim, Jung-Myung Lee, Jung-Woo Son, Jaedeok Kim, Young-Guk Ko, Donghoon Choi, Yangsoo Jang.
Abstract
PURPOSE: The aim of this study is to identify the extent of initial malapposition using optical coherence tomography (OCT) in ST-elevation myocardial infarctions (STEMI) treated with different types of drug-eluting stents (DES).Entities:
Mesh:
Substances:
Year: 2010 PMID: 20376884 PMCID: PMC2852787 DOI: 10.3349/ymj.2010.51.3.332
Source DB: PubMed Journal: Yonsei Med J ISSN: 0513-5796 Impact factor: 2.759
Fig. 1Measurements of strut apposition. The distances between the endo-luminal surface of the strut reflection and the vessel wall were measured. (A and B) shows OCT image of SES and strut measurements and represents malapposition. (C and D) shows OCT image of PES and strut measurement shows normal stent apposition. (E and F) represents OCT image of ZES and its measurement shows normal stent apposition. Definition of malapposition: ≥ 160 µm for SES, ≥ 130 µm for PES, ≥ 110 µm for ZES.7 OCT, optical coherence tomography; SES, sirolimus-eluting stent; PES, paclitaxel-eluting stent; ZES, zotarolimus-eluting stent.
Baseline Characteristics of Patients
CABG, coronary artery bypass graft.
Values are presented as means ± SD or percentages.
Angiographic Results of Patients
RCA, right coronary artery; LAD, left anterior descending artery; LCX, left circumflex artery.
Values are presented as mean ± SD or percentages.
Procedural Results, Quantitative Coronary Angiographic Analysis and IVUS Measurements Among Stents
IVUS, intravascular ultrasound; SES, sirolimus-eluting stent; PES, paclitaxel-eluting stent; ZES, zotarolimus-eluting stent; EEM, external elastic membrane; CSA, cross sectional area.
Values are presented as means ± SD or percentages.
Fig. 2Average distances between lumen and stent. *Represents statistical significance (p = 0.00). Definition of incomplete stent apposition: ≥ 160 µm for SES, ≥ 130 µm for PES, ≥ 110 µm for ZES. SES, sirolimus-eluting stent; PES, paclitaxel-eluting stent; ZES, zotarolimus-eluting stent.7
Fig. 3Comparison of malapposition strut numbers according to stent. *Represents statistical significance (p = 0.00), †indicates non-significant results (p = 0.81). Definition of malapposition: ≥ 160 µm for SES, ≥ 130 µm for PES, ≥ 110 µm for ZES.7 SES, sirolimus-eluting stent; PES, paclitaxel-eluting stent; ZES, zotarolimus-eluting stent.
Malapposition Rates in Optical Coherence Tomography (OCT) according to the Results of Intravascular Ultrasound (IVUS)
OCT, optical coherence tomography; IVUS, intravascular ultrasonography.
Values are presented as means ± SD or percentages.
Multivariate Predictors of Malapposed Struts
ZES, zotarolimus-eluting stent; MLD, minimal luminal diameter; PCI, percutaneous coronary intervention; SES, sirolimus-eluting stent.
Independent predictors of malapposed strut was identified using mixed effect multivariate logistic analysis after controlling age, gender, diabetes, stent type (SES versus non-SES), stent size, stent length, reference vessel diameter, minimal luminal diameter and maximum inflation pressure.