Literature DB >> 20376884

The initial extent of malapposition in ST-elevation myocardial infarction treated with drug-eluting stent: the usefulness of optical coherence tomography.

Ung Kim1, Jung-Sun Kim, Jin-Sun Kim, Jung-Myung Lee, Jung-Woo Son, Jaedeok Kim, Young-Guk Ko, Donghoon Choi, Yangsoo Jang.   

Abstract

PURPOSE: The aim of this study is to identify the extent of initial malapposition using optical coherence tomography (OCT) in ST-elevation myocardial infarctions (STEMI) treated with different types of drug-eluting stents (DES).
MATERIALS AND METHODS: Twenty four STEMI patients that underwent primary percutaneous coronary intervention (PCI) were enrolled. The OCT and intravascular ultrasound (IVUS) were performed within 72 hours after the primary PCI. Distances between the endo-luminal surface of the strut reflection and the vessel wall and the extent of malapposition were measured and analyzed.
RESULTS: Sirolimus-eluting stents (SES), paclitaxel-eluting stents (PES) and zotarolimus-eluting stents (ZES) were deployed in 7 patients (29%), 7 patients (29%) and 10 patients (42%). In total, 4951 struts in 620 mm single-stent segments were analyzed (1463 struts in SES, 1522 in PES, and 1966 in ZES). In strut analysis by OCT, the incidence of malapposition was 17 % (860/4951) and in stent analysis by IVUS, malapposition rate was 21% (5/24). The malapposition rate of strut level using OCT in 5 patients who had malapposition in IVUS was significantly higher than the 19 of those who had not (32 +/- 5% vs. 12 +/- 6%, p = 0.001). In addition, the frequency of malapposition was also significantly different (28% in SES, 11% in PES, 10% in ZES, p = 0.001). The use of SES was an independent predictor of malapposed struts.
CONCLUSION: The incidence of malapposition using OCT was quite prevalent in STEMI after primary PCI with DES implantation and SES has especially higher rates of malapposition compared to other DESs.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20376884      PMCID: PMC2852787          DOI: 10.3349/ymj.2010.51.3.332

Source DB:  PubMed          Journal:  Yonsei Med J        ISSN: 0513-5796            Impact factor:   2.759


INTRODUCTION

Intravascular optical coherence tomography (OCT) is a useful tool for the detection of a thin layer of neointimal hyperplasia (NIH) and malapposition with high-resolution (≥ 10 µm).1-5 ST-elevation myocardial infarction (STEMI) is associated with a higher risk of stent thrombosis and malapposition may be a surrogate marker.6-8 Currently, available drug eluting stents (DES) have different properties of stent expansion and apposition. However, the data are lacking in initial findings such as malapposition using OCT in STEMI treated with DES. The aim of this study is to identify the extent of initial malapposition using OCT in STEMI treated with DES and to compare the results between different stent types.

MATERIALS AND METHODS

Study population and protocol

Twenty four patients (20 males, 83%) who were diagnosed with STEMI and underwent primary percutaneous coronary intervention (PCI) were enrolled from Aug. 2007 to Mar. 2008 in the Cardiovascular Hospital of Yonsei University Medical Center. The study protocols were approved by the institutional ethics committee of Yonsei University College of Medicine, and written consent was obtained from all patients before the procedure. Emergency coronary angiography and primary PCI were performed by the femoral approach according to standard techniques in each of the patients. Using a guiding catheter for magnification and calibration, quantitative coronary angiography was performed immediately after the procedure. OCT was performed in infarct-related arteries treated with DES at the second stage within 72 hours of the primary PCI. STEMI was defined as continuous chest pain lasting > 30 min, with arrival at our hospital within 12 hours from the onset of the symptom, with ST-segment elevation ≥ 0.1 mV in two or more contiguous leads on 12-lead electrocardiogram and elevation of serum cardiac enzymes such as troponin and/or creatine kinase-MB (CK-MB). The luminal diameter proximal to the culprit lesion was required to be between 2.5 mm and 4.0 mm due to the maximal diameter of the currently available occlusion balloon. Exclusion criteria were poor left ventricular function (< 30% of ejection fraction), hemodynamic instability, significant left main coronary artery disease, renal insufficiency with baseline serum creatinine > 1.5 mg/dL, a single remaining vessel, highly tortuous vessels, and ostial or very proximal lesions (< 15 mm from ostium).

Adjunctive medication

All patients received an intravenous bolus injection of 5,000 U of heparin. In primary PCI in STEMI, 200 mg of aspirin was given before the procedure and 600 mg of clopidogrel was given during the surgery. After the PCI, both drugs were continuously administered for at least 12 months. The use of glycoprotein IIb/IIIa inhibitors was left to the discretion of the operator.

Intravascular ultrasound examination

An intravascular ultrasound (IVUS) was performed during primary PCI if needed. However, it was done concurrently with OCT within 72 hours and the data was analyzed. IVUS assessment was performed using a commercially available IVUS system (Boston Scientific/Scimed, Natick, MA, USA) with automatic motorized 0.5 mm/s pullback. All IVUS procedures were performed after intracoronary administration of nitroglycerin 200 µg. Studies were recorded on a compact disk for offline analysis using Index (Echoplaque 2; INDEC System, Inc., Mountain View, CA, USA). Proximal reference cross-sectional area (CSA, mm2) and distal reference CSA (mm2) were measured at the most visually normal area (largest lumen with the least plaque burden) within 10 mm of the proximal or distal ends of the lesion. Malapposition was defined as the separation of at least 1 stent strut from the intimal surface of the arterial wall that was not overlapping a side branch and had evidence of blood flow (speckling) behind the strut.9

OCT examination

The OCT system used in this study consisted of a computer, a monitor display, an interface unit (Model M2 Cardiology Imaging System, LightLab Imaging, Inc., Westford, MA, USA), and a 0.014-inch wire-type imaging catheter (Image-Wire, LightLab Imaging, Inc., Westford, MA, USA). A motorized pull-back system at 1.0 mm/s was used, and OCT images were acquired at 15 frames/s. The axial resolution capacity of this system was 15 µm. All patients were given 5,000 IU of heparin intravenously before their respective procedures. A 6Fr or 7Fr guiding catheter was introduced into the coronary artery by a femoral approach, and nitroglycerin (200 µg) was administered through the guiding catheter. To remove the blood from the field of view and allow for clear images, an occlusion balloon catheter (Helios, Avantec Vascular Corp., Sunnyvale, CA, USA) and flushing with warm lac was used in an over-the-wire type method with the flush lumen and the lumen for crossing the imaging catheter. There were two metal markers at the tip and at the center of the occlusion balloon. The occlusion catheter was positioned proximal to the stent and the image catheter was positioned distal to the stent. During image acquisition, the occlusion balloon was inflated at 0.4 to 0.6 atm and Ringer lactate was infused at 0.5 to 1.0 mL/s. The image wire was pulled from distal to proximal, and continuous images were stored digitally for subsequent analysis.

OCT analysis

Cross-sectional OCT images were analyzed at 1-mm intervals (every 15 frames). Stent struts appeared as highly reflective surfaces that cast shadows on the vessel wall behind. The distances between the endo-luminal surface of the strut reflection and the vessel wall were measured by prolonging and joining the contours of the wall on either side of the strut shadow with a measurement line as perpendicular as possible to the strut and vessel wall (Fig. 1). When the strut was not fully attached to the vessel wall by visual estimation, the position of the stent strut to the vessel wall was measured by magnifying the individual stent strut to maximize accuracy. Malapposed stent struts were defined as struts with detachment from the vessel wall ≥ 160 µm for sirolimus-eluting stents (SES; Cypher select, Cordis, J&J, East Bridgewater, NJ, USA), ≥ 130 µm for paclitaxeleluting stents (PES; Taxus Liberte, Boston Scientific, Natick, MA, USA), and ≥ 110 µm for zotaroliums-eluting stents (ZES; Endeavor, Medtronic, Minneapolis, MN, USA).10 Stent overlapping segments and bifurcation lesions with major side branches were excluded from this analysis. Inter-observer and intra-observer variabilities in measured distances were assessed by the evaluation of 20 random cross-sectional images by two independent readers and by the same reader at two separate time points, respectively. The variations between measurements were calculated using the linear mixed model (one-way mixed and two-way mixed models).
Fig. 1

Measurements of strut apposition. The distances between the endo-luminal surface of the strut reflection and the vessel wall were measured. (A and B) shows OCT image of SES and strut measurements and represents malapposition. (C and D) shows OCT image of PES and strut measurement shows normal stent apposition. (E and F) represents OCT image of ZES and its measurement shows normal stent apposition. Definition of malapposition: ≥ 160 µm for SES, ≥ 130 µm for PES, ≥ 110 µm for ZES.7 OCT, optical coherence tomography; SES, sirolimus-eluting stent; PES, paclitaxel-eluting stent; ZES, zotarolimus-eluting stent.

Statistical analysis

Results are expressed as a mean ± SD or number (%). Comparisons of categorical variables were made using the Chi-square test and the student's t-test while one way ANOVA were used to compare continuous variables. The intra-observer and inter-observer agreements (reproducibility) were assessed by determining the mean ± SD of differences between observations and between observers, respectively. If the distributions were skewed, a non-parametric test was used. Because the observation of struts in the same stent is not independent of each other and the hierarchical nature of the data, mixed effect multiple logistic regression with malapposition as the outcome variable was applied to address random and fixed effects at the strut, lesion and patient levels. At the lesion level, stent type (SES versus non-SES), stent diameter, stent length, reference vessel diameter, minimal luminal diameter and maximum inflation pressure were considered. At the patient level, age, sex and diabetes mellitus were considered.11 All analyses were performed using the Statistical Analysis System software (SAS 9.1.3., SAS Institute, Cary, NC, USA; and R 2.8.1., R Foundation, Vienna, Austria) and p < 0.05 was considered statistically significant.

RESULTS

Baseline characteristics and angiographic findings

Baseline characteristics of patients are presented in Table 1.
Table 1

Baseline Characteristics of Patients

CABG, coronary artery bypass graft.

Values are presented as means ± SD or percentages.

The patients' mean age was 59 ± 10 years and they were predominantly male. Seventy percent of patients (16, 66%) had multi-vessel disease, with culprit lesions predominately in the left descending artery (14, 58%). SES was implanted in 7 patients (29%), PES in 7 patients (29%), and ZES in 10 patients (42%). Angiographic results are shown in Table 2. A pre-stenting balloon was applied in all patients. High pressures (15.0 ± 2.0 atm) were applied in all implanted stents but it was not statistically significant (Table 3). The quantitative angiographic analysis in each group is shown in Table 3.
Table 2

Angiographic Results of Patients

RCA, right coronary artery; LAD, left anterior descending artery; LCX, left circumflex artery.

Values are presented as mean ± SD or percentages.

Table 3

Procedural Results, Quantitative Coronary Angiographic Analysis and IVUS Measurements Among Stents

IVUS, intravascular ultrasound; SES, sirolimus-eluting stent; PES, paclitaxel-eluting stent; ZES, zotarolimus-eluting stent; EEM, external elastic membrane; CSA, cross sectional area.

Values are presented as means ± SD or percentages.

IVUS measurements

Post-intervention IVUS measurements are represented in Table 3. Among the groups, IVUS parameters were not significantly different except for minimal stent CSA. The initial rate of malapposition measured by IVUS was 21%, and had a higher trend toward to SES.

OCT measurements

The total number of measured struts was 4951, including 1463 (30%) in SES, 1522 (31%) in PES, and 1966 (39%) in ZES. The average distances between the lumen and stent were different, and the SES was significantly longer than the other stents (188 ± 46 in SES, 144 ± 18 in PES, 127 ± 34 in ZES, p = 0.00) (Fig. 2). The numbers of malapposition struts were 28% in SES, but only 11% in PES and 10% in ZES; the malapposition rates of SES were significantly higher than those of the other two stents (Fig. 3). The rate of malapposition of 5 patients who had malapposition in IVUS showed 32% in strut level by OCT but, 12% in 19 patients who had not and this is statistically significant (32 ± 5% vs. 12 ± 6%, p = 0.00) (Table 4).
Fig. 2

Average distances between lumen and stent. *Represents statistical significance (p = 0.00). Definition of incomplete stent apposition: ≥ 160 µm for SES, ≥ 130 µm for PES, ≥ 110 µm for ZES. SES, sirolimus-eluting stent; PES, paclitaxel-eluting stent; ZES, zotarolimus-eluting stent.7

Fig. 3

Comparison of malapposition strut numbers according to stent. *Represents statistical significance (p = 0.00), †indicates non-significant results (p = 0.81). Definition of malapposition: ≥ 160 µm for SES, ≥ 130 µm for PES, ≥ 110 µm for ZES.7 SES, sirolimus-eluting stent; PES, paclitaxel-eluting stent; ZES, zotarolimus-eluting stent.

Table 4

Malapposition Rates in Optical Coherence Tomography (OCT) according to the Results of Intravascular Ultrasound (IVUS)

OCT, optical coherence tomography; IVUS, intravascular ultrasonography.

Values are presented as means ± SD or percentages.

If we consider the definition of stent malapposition in OCT as having malapposed struts over 10% in all struts, the malapposition rate was 67% (16/24) or if over 5%, 83% (20/24) or if any malapposition, 100% (24/24). Regarding sensitivity of OCT as 100%, that of IVUS could be calculated as 33%. The intra-observer correlation coefficient between variability of observer1 was 0.981 (95% CI: 0.975-0.985) using the one-way mixed model, where patient effects are random. The inter-observer correlation coefficient between the variability of two observers using the two-way mixed model, where patient effects are random and observers effects are fixed, were 0.982 (95% CI: 0.977-0.986) and 0.983 (95% CI: 0.978-0.987), respectively. Multivariate predictors of malapposed stent struts on mixed effect logistic regression analysis are shown in Table 5. The use of SES was an independent predictor of malapposed struts together with large reference vessel diameter, small stent size and high maximal pressure.
Table 5

Multivariate Predictors of Malapposed Struts

ZES, zotarolimus-eluting stent; MLD, minimal luminal diameter; PCI, percutaneous coronary intervention; SES, sirolimus-eluting stent.

Independent predictors of malapposed strut was identified using mixed effect multivariate logistic analysis after controlling age, gender, diabetes, stent type (SES versus non-SES), stent size, stent length, reference vessel diameter, minimal luminal diameter and maximum inflation pressure.

DISCUSSION

This may be the first report assessing initial malapposition in STEMI patients undergoing primary PCI by OCT. The major findings of this paper indicate that the considerable rates of malapposition using OCT were detected during primary PCI in STEMI patients treated with DES and total malapposition rates were 17% but malapposition rates in SES were significantly higher than those of other DESs (28% in SES vs. 11% in PES vs. 10% in ZES, p = 0.00). SES was also an independent predictor of a malapposed strut after controlling the patients and lesion factors. The frequency of initial stent malapposition by IVUS was previously reported and the rates from the results had such a wide range that one study showed the rates were as low as 8.7% in SES and 2.4% in PES, but another studies showed 16% in SES, 19% in SES and 12.6% in ZES.9,12,13 However, previous reports have a heterogeneous study population from stable angina to acute myocardial infarction. We performed IVUS before OCT in all patients performing primary PCI diagnosed as STEMI, therefore, the initial rates of malapposition on IVUS seemed to be relatively higher than those of the above mentioned report (21% vs. 11 to 19%). Also these variable rates of malapposition might be affected due to unique characteristics of each stent like strut thickness. This malapposition rate on IVUS was comparable with OCT findings although there was somewhat different method of measurement (21% in IVUS vs. 16% in OCT). But, IVUS cannot measure stent struts more accurately than OCT, because the IVUS resolution to evaluate each strut of a stent compared to OCT is limited. OCT can accurately measure the intima-media thickness and identify fine abnormal structures such as detailed components of coronary atherosclerotic plaques, hematoma, dissection, intra-stent tissue prolapse, stent fracture, and stent malapposition.14-17 The present study implies specific subsets of patients with STEMI after primary PCI with OCT. Although SES is effective in preventing restenosis, currently their safety has been questioned and many defects such as stent fractures and malapposition have been challenging. Therefore, initial malapposition might be possible to be relate to late clinical outcomes of SES. Although we do not know the clinical implication of initial malapposition, it is certain that this finding is too significant to pass over and could not exclude the association with late clinical adverse events such as stent thrombosis through the inhibition of healthy endothelial cell growth and neointimal coverage of stent strut. Given that SES has a relatively stronger and thicker stent strut body than other stents, malapposition may occur more often; nevertheless, high pressure is applied. This finding suggests that malapposition might have a role to retard the neointimal growth addition to the effects of the immunosuppressant drug and polymer because SES have been know to reduced the significantly neointima than other DESs. These concepts were also evidenced by Tanigawa, et al.11 They showed that despite angiographic optimization with high pressures and adequately sized balloons, malapposed stent struts are frequently found in implantation of Cypher Select stents which have a thicker stent strut and closed cell design. Therefore, the present study suggests that a close clinical follow-up and strict dual anti-platelet therapy may be necessary to prevent the cardiac events after DES implantation in STEMI because vulnerability could remain even in successful treatment with DES. One limitations of this study is that only the initial period after the stent was evaluated. Therefore, the follow-up OCT and clinical finding are needed to investigate and although this was not performed in the current study it is now in progress. Another limitation is the number of patients, which was fairly small and there is the possibility of selection bias, although the measured strut number was quite large. However, we believe that this study is valuable because it is the first report about the initial extent of malapposition of DES using OCT during primary PCI. Another limitation is that this malapposition observed within 72 hours after stenting may be associated with procedural issues. However, we performed a primary PCI by one fully experienced operator and the procedures including ballooning or high pressure before and after stenting were also standardized in all patients to reduce errors. In conclusion, considerable rates of malapposition were detected in STEMI after primary PCI with DES implantation and OCT can be a useful tool for finding an initial malapposition. Furthermore, follow-up clinical, angiographic, and OCT findings of stent malapposition are needed to evaluate the clinical implications and fate of initial stent malapposition in STEMI after primary PCI.
  17 in total

1.  Stent thrombosis in the modern era: a pooled analysis of multicenter coronary stent clinical trials.

Authors:  D E Cutlip; D S Baim; K K Ho; J J Popma; A J Lansky; D J Cohen; J P Carrozza; M S Chauhan; O Rodriguez; R E Kuntz
Journal:  Circulation       Date:  2001-04-17       Impact factor: 29.690

2.  Visualization of coronary atherosclerotic plaques in patients using optical coherence tomography: comparison with intravascular ultrasound.

Authors:  Ik-Kyung Jang; Brett E Bouma; Dong-Heon Kang; Seung-Jung Park; Seong-Wook Park; Ki-Bae Seung; Kyu-Bo Choi; Milen Shishkov; Kelly Schlendorf; Eugene Pomerantsev; Stuart L Houser; H Thomas Aretz; Guillermo J Tearney
Journal:  J Am Coll Cardiol       Date:  2002-02-20       Impact factor: 24.094

3.  Sirolimus-eluting stent fracture with thrombus, visualization by optical coherence tomography.

Authors:  Junya Shite; Daisuke Matsumoto; Mitsuhiro Yokoyama
Journal:  Eur Heart J       Date:  2005-10-24       Impact factor: 29.983

4.  Late incomplete stent apposition after sirolimus-eluting stent implantation: a serial intravascular ultrasound analysis.

Authors:  Junya Ako; Yoshihiro Morino; Yasuhiro Honda; Ali Hassan; Shinjo Sonoda; Paul G Yock; Martin B Leon; Jeffrey W Moses; Heidi N Bonneau; Peter J Fitzgerald
Journal:  J Am Coll Cardiol       Date:  2005-09-20       Impact factor: 24.094

5.  Assessment of coronary intima--media thickness by optical coherence tomography: comparison with intravascular ultrasound.

Authors:  Teruyoshi Kume; Takashi Akasaka; Takahiro Kawamoto; Nozomi Watanabe; Eiji Toyota; Yoji Neishi; Renan Sukmawan; Yoshito Sadahira; Kiyoshi Yoshida
Journal:  Circ J       Date:  2005-08       Impact factor: 2.993

6.  Late stent malapposition after drug-eluting stent implantation: an intravascular ultrasound analysis with long-term follow-up.

Authors:  Myeong-Ki Hong; Gary S Mintz; Cheol Whan Lee; Duk-Woo Park; Kyoung-Min Park; Bong-Ki Lee; Young-Hak Kim; Jong-Min Song; Ki-Hoon Han; Duk-Hyun Kang; Sang-Sig Cheong; Jae-Kwan Song; Jae-Joong Kim; Seong-Wook Park; Seung-Jung Park
Journal:  Circulation       Date:  2006-01-24       Impact factor: 29.690

7.  Intravascular optical coherence tomography: optimisation of image acquisition and quantitative assessment of stent strut apposition.

Authors:  Jun Tanigawa; Peter Barlis; Carlo Di Mario
Journal:  EuroIntervention       Date:  2007-05       Impact factor: 6.534

8.  Optical coherence tomography as a tool for percutaneous coronary interventions.

Authors:  Larry J Diaz-Sandoval; Brett E Bouma; Guillermo J Tearney; Ik-Kyung Jang
Journal:  Catheter Cardiovasc Interv       Date:  2005-08       Impact factor: 2.692

9.  In vivo characterization of coronary atherosclerotic plaque by use of optical coherence tomography.

Authors:  Ik-Kyung Jang; Guillermo J Tearney; Briain MacNeill; Masamichi Takano; Fabian Moselewski; Nicusor Iftima; Milen Shishkov; Stuart Houser; H Thomas Aretz; Elkan F Halpern; Brett E Bouma
Journal:  Circulation       Date:  2005-03-21       Impact factor: 29.690

10.  Predictors of subacute stent thrombosis: results of a systematic intravascular ultrasound study.

Authors:  Edouard Cheneau; Laurent Leborgne; Gary S Mintz; Jun-ichi Kotani; Augusto D Pichard; Lowell F Satler; Daniel Canos; Marco Castagna; Neil J Weissman; Ron Waksman
Journal:  Circulation       Date:  2003-06-23       Impact factor: 29.690

View more
  12 in total

1.  Optical coherence tomography-based evaluation of malapposed strut coverage after drug-eluting stent implantation.

Authors:  Byeong-Keuk Kim; Dong-Ho Shin; Jung-Sun Kim; Young-Guk Ko; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Int J Cardiovasc Imaging       Date:  2012-03-25       Impact factor: 2.357

2.  Major determinants for the uncovered stent struts on optical coherence tomography after drug-eluting stent implantation.

Authors:  Byeong-Keuk Kim; Jung-Sun Kim; Changmyung Oh; Young-Guk Ko; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Int J Cardiovasc Imaging       Date:  2011-05-31       Impact factor: 2.357

3.  Optical coherence tomographic comparison of neointimal coverage between sirolimus- and resolute zotarolimus-eluting stents at 9 months after stent implantation.

Authors:  Jung-Sun Kim; Jin-Sun Kim; Dong-Ho Shin; Byeong-Keuk Kim; Young-Guk Ko; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Int J Cardiovasc Imaging       Date:  2011-08-21       Impact factor: 2.357

4.  Evaluation of neointimal morphology of lesions with or without in-stent restenosis: an optical coherence tomography study.

Authors:  Sung-Joo Lee; Byeong-Keuk Kim; Jung-Sun Kim; Young-Guk Ko; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Clin Cardiol       Date:  2011-09-16       Impact factor: 2.882

5.  Optical coherence tomography derived cut-off value of uncovered stent struts to predict adverse clinical outcomes after drug-eluting stent implantation.

Authors:  Hoyoun Won; Dong-Ho Shin; Byeong-Keuk Kim; Gary S Mintz; Jung-Sun Kim; Young-Guk Ko; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Int J Cardiovasc Imaging       Date:  2013-04-25       Impact factor: 2.357

6.  Favorable neointimal coverage in everolimus-eluting stent at 9 months after stent implantation: comparison with sirolimus-eluting stent using optical coherence tomography.

Authors:  Hyun Hee Choi; Jung-Sun Kim; Duck Hyoung Yoon; Kyung-Soon Hong; Tae Hoon Kim; Byeong Keuk Kim; Young-Guk Ko; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Int J Cardiovasc Imaging       Date:  2011-03-26       Impact factor: 2.357

7.  Optical coherence tomography findings of very late stent thrombosis after drug-eluting stent implantation.

Authors:  Young-Guk Ko; Dong-Min Kim; Jin Man Cho; So Yeon Choi; Jung Han Yoon; Jung-Sun Kim; Byeong-Keuk Kim; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Int J Cardiovasc Imaging       Date:  2011-06-08       Impact factor: 2.357

8.  A newly formed and ruptured atheromatous plaque within neointima after drug-eluting stent implantation: 2-year follow-up intravascular ultrasound and optical coherence tomography studies.

Authors:  Chang-Myung Oh; Jeonggeun Moon; Hee Tae Yu; Ji-Yong Jang; Jung-Sun Kim; Young-Guk Ko; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Yonsei Med J       Date:  2011-11       Impact factor: 2.759

9.  Comparisons of the effects of stent eccentricity on the neointimal hyperplasia between sirolimus-eluting stent versus paclitaxel-eluting stent.

Authors:  Byeong-Keuk Kim; Young-Guk Ko; Seungjin Oh; Jung-Sun Kim; Woong Chol Kang; Dong Woon Jeon; Joo Young Yang; Donghoon Choi; Myeong-Ki Hong; Taehoon Ahn; Yangsoo Jang
Journal:  Yonsei Med J       Date:  2010-11       Impact factor: 2.759

10.  Relationship between angiographic late loss and 5-year clinical outcome after drug-eluting stent implantation.

Authors:  Young-June Yang; Sanghoon Shin; Byeong-Keuk Kim; Jung-Sun Kim; Dong-Ho Shin; Young-Guk Ko; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Yonsei Med J       Date:  2013-01-01       Impact factor: 2.759

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.