Experimental investigation of factors affecting the absolute recovery coefficients in iodine-124 PET lesion imaging.

Walter Jentzen1.   

Abstract

The use of recovery coefficients (RCs) in (124)I PET lesion imaging is a simple method to correct the imaged activity concentration (AC) primarily for the partial-volume effect and, to a minor extent, for the prompt gamma coincidence effect. The aim of this phantom study was to experimentally investigate a number of various factors affecting the (124)I RCs. Three RC-based correction approaches were considered. These approaches differ with respect to the volume of interest (VOI) drawn, which determines the imaged AC and the RCs: a single voxel VOI containing the maximum value (maximum RC), a spherical VOI with a diameter of the scanner resolution (resolution RC) and a VOI equaling the physical object volume (isovolume RC). Measurements were performed using mainly a stand-alone PET scanner (EXACT HR(+)) and a latest-generation PET/CT scanner (BIOGRAPH mCT). The RCs were determined using a cylindrical phantom containing spheres or rotational ellipsoids and were derived from images acquired with a reference acquisition protocol. For each type of RC, the influence of the following factors on the RC was assessed: object shape, background activity spill in and iterative image reconstruction parameters. To evaluate the robustness of the RC-based correction approaches, the percentage deviation between RC-corrected and true ACs was determined from images acquired with a clinical acquisition protocol of different AC regimes. The observed results of the shape and spill-in effects were compared with simulation data derived from a convolution-based model. The study demonstrated that the shape effect was negligible and, therefore, was in agreement with theoretical expectations. In contradiction to the simulation results, the observed spill-in effect was unexpectedly small. To avoid variations in the determination of RCs due to reconstruction parameter changes, image reconstruction with a pixel length of about one-third or less of the scanner resolution and an OSEM 1 x 32 algorithm or one with somewhat higher number of effective iterations are recommended. Using the clinical acquisition protocol, the phantom study indicated that the resolution- or isovolume-based recovery-correction approaches appeared to be more appropriate to recover the ACs from patient data; however, the application of the three RC-based correction approaches to small lesions containing low ACs was, in particular, associated with large underestimations. The phantom study had several limitations, which were discussed in detail.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20360631     DOI: 10.1088/0031-9155/55/8/016

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  18 in total

1.  EANM Dosimetry Committee guidance document: good practice of clinical dosimetry reporting.

Authors:  M Lassmann; C Chiesa; G Flux; M Bardiès
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-08-27       Impact factor: 9.236

Review 2.  Clinical applications of 124I-PET/CT in patients with differentiated thyroid cancer.

Authors:  Lutz S Freudenberg; Walter Jentzen; Alexander Stahl; Andreas Bockisch; Sandra J Rosenbaum-Krumme
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-04-12       Impact factor: 9.236

3.  The influence of saliva flow stimulation on the absorbed radiation dose to the salivary glands during radioiodine therapy of thyroid cancer using 124I PET(/CT) imaging.

Authors:  Walter Jentzen; Dorothee Balschuweit; Jochen Schmitz; Lutz Freudenberg; Ernst Eising; Thomas Hilbel; Andreas Bockisch; Alexander Stahl
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-07-13       Impact factor: 9.236

4.  Diagnosis and dosimetry in differentiated thyroid carcinoma using 124I PET: comparison of PET/MRI vs PET/CT of the neck.

Authors:  James Nagarajah; Walter Jentzen; Verena Hartung; Sandra Rosenbaum-Krumme; Christian Mikat; Till Alexander Heusner; Gerald Antoch; Andreas Bockisch; Alexander Stahl
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-07-08       Impact factor: 9.236

5.  MIRD pamphlet No. 23: quantitative SPECT for patient-specific 3-dimensional dosimetry in internal radionuclide therapy.

Authors:  Yuni K Dewaraja; Eric C Frey; George Sgouros; A Bertrand Brill; Peter Roberson; Pat B Zanzonico; Michael Ljungberg
Journal:  J Nucl Med       Date:  2012-06-28       Impact factor: 10.057

Review 6.  Quantitative imaging of 124I and 86Y with PET.

Authors:  Mark Lubberink; Hans Herzog
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-04-12       Impact factor: 9.236

7.  Practical reconstruction protocol for quantitative (90)Y bremsstrahlung SPECT/CT.

Authors:  W Siman; J K Mikell; S C Kappadath
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

8.  Evaluation of a 3D point cloud tetrahedral tomographic reconstruction method.

Authors:  N F Pereira; A Sitek
Journal:  Phys Med Biol       Date:  2010-08-24       Impact factor: 3.609

9.  68Ga-DOTA-TOC uptake in neuroendocrine tumour and healthy tissue: differentiation of physiological uptake and pathological processes in PET/CT.

Authors:  A Kroiss; D Putzer; C Decristoforo; C Uprimny; B Warwitz; B Nilica; M Gabriel; D Kendler; D Waitz; G Widmann; I J Virgolini
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-01-05       Impact factor: 9.236

10.  Monte Carlo-based 3-dimensional dosimetry of salivary glands in radioiodine treatment of differentiated thyroid cancer estimated using 124I PET.

Authors:  R F Hobbs; W Jentzen; A Bockisch; G Sgouros
Journal:  Q J Nucl Med Mol Imaging       Date:  2013-03       Impact factor: 2.346

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.