Monitoring linear accelerator output constancy using the PTW Linacheck.

Garry M McDermott1, Andrew H Buckle.   

Abstract

The PTW-Linacheck was assessed for its ability to monitor linear accelerator radiation output constancy. The key issues that were considered were the setup for daily output measurements, e.g., requirements for build-up and backscatter material, and the reproducibility and linearity of the device with linear accelerator output. An appropriate measurement setup includes a 10 × 10 cm field at 100 cm FSD, 5 cm backscatter, and no added build-up for 4 MeV electron beams, 1 cm added build-up for 6-16 MeV electron beams and 5 cm added build-up for 6-15 MV photon beams. Using this measurement setup, the dose linearity and short-term reproducibility were acceptable; however, the Linacheck should be recalibrated on a monthly basis to ensure acceptable long-term reproducibility.
Copyright © 2011 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2010        PMID: 20346645     DOI: 10.1016/j.meddos.2009.12.001

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482