BACKGROUND: Minimally invasive surgery (MIS) suffers from the inability to directly palpate organs for tumour localization. A tactile sensing system (TSS), consisting of a probe and a visualization interface, was developed to present an active pressure map of the contact surface to locate tumours during MIS. METHODS: The TSS performance was compared to MIS graspers to locate occult 10 mm phantom tumours in ex vivo bovine liver and ex vivo porcine lung. Performance assessment included applied pressure, localization distance and accuracy. RESULTS: The TSS realized a relative 71% reduction in maximum applied pressure and a 31% increase in detection accuracy in liver tissue (when compared to MIS graspers) and demonstrated no significant differences in performance when palpating lung tissue. CONCLUSIONS: The TSS may help surgeons to identify occult tumours during surgery by restoring some of the haptic information lost during MIS. (c) 2010 John Wiley & Sons, Ltd.
BACKGROUND: Minimally invasive surgery (MIS) suffers from the inability to directly palpate organs for tumour localization. A tactile sensing system (TSS), consisting of a probe and a visualization interface, was developed to present an active pressure map of the contact surface to locate tumours during MIS. METHODS: The TSS performance was compared to MIS graspers to locate occult 10 mm phantom tumours in ex vivo bovine liver and ex vivo porcine lung. Performance assessment included applied pressure, localization distance and accuracy. RESULTS: The TSS realized a relative 71% reduction in maximum applied pressure and a 31% increase in detection accuracy in liver tissue (when compared to MIS graspers) and demonstrated no significant differences in performance when palpating lung tissue. CONCLUSIONS: The TSS may help surgeons to identify occult tumours during surgery by restoring some of the haptic information lost during MIS. (c) 2010 John Wiley & Sons, Ltd.
Authors: Johanna Miller; Manuel Braun; Johannes Bilz; Sebastian Matich; Carsten Neupert; Wolfgang Kunert; Andreas Kirschniak Journal: Surg Endosc Date: 2020-07-22 Impact factor: 4.584