Literature DB >> 20307392

Fabrication of nanoindented electrodes for glucose detection.

Gymama Slaughter1.   

Abstract

BACKGROUND: The objective of this article was to design, fabricate, and evaluate a novel type of glucose biosensors based on the use of atomic force microscopy to create nanoindented electrodes (NIDEs) for the selective detection of glucose.
METHODS: Atomic force microscopy nanoindentation techniques were extended to covalently immobilized glucose oxidase on NIDEs via composite hydrogel membranes composed of interpenetrating networks of inherently conductive poly(3,4-ethylenedioxythiophene) tetramethacrylate grown within ultraviolet cross-linked hydroxyethylmethacrylate-based hydrogels to produce an in vitro amperometric NIDE biosensor for the long-term monitoring of glucose.
RESULTS: The calibration curve for glucose was linear from 0.25 to 20 mM. Results showed that the NIDE glucose biosensor has a much higher detection sensitivity of 0.32 microA/mM and rapid response times (<5 seconds). There was no interference from the competing interferent (fructose) present; the only interference was from species that react with H(2)O(2) (ascorbic acid). The linear equation was B(response) (microA) = 0.323 [glucose] (mM) + 0.634 (microA); n = 24, r(2) = 0.994.
CONCLUSION: Results showed that the resultant NIDE glucose biosensor increases the dynamic range, device sensitivity, and response time and has excellent detecting performance for glucose. (c) 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20307392      PMCID: PMC2864167          DOI: 10.1177/193229681000400212

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  15 in total

Review 1.  Biomolecule immobilization on electrode surfaces by entrapment or attachment to electrochemically polymerized films. A review.

Authors:  S Cosnier
Journal:  Biosens Bioelectron       Date:  1999-05-31       Impact factor: 10.618

2.  Sol-gel based glucose biosensors employing optical oxygen transducers, and a method for compensating for variable oxygen background.

Authors:  O S Wolfbeis; I Oehme; N Papkovskaya; I Klimant
Journal:  Biosens Bioelectron       Date:  2000-03       Impact factor: 10.618

3.  Electrode systems for continuous monitoring in cardiovascular surgery.

Authors:  L C CLARK; C LYONS
Journal:  Ann N Y Acad Sci       Date:  1962-10-31       Impact factor: 5.691

4.  A wire-based dual-analyte sensor for glucose and lactate: in vitro and in vivo evaluation.

Authors:  W Kenneth Ward; Jody L House; Jonathan Birck; Ellen M Anderson; Lawrence B Jansen
Journal:  Diabetes Technol Ther       Date:  2004-06       Impact factor: 6.118

5.  Lactate/glucose dynamics after rat fluid percussion brain injury.

Authors:  T Chen; Y Z Qian; X Di; A Rice; J P Zhu; R Bullock
Journal:  J Neurotrauma       Date:  2000-02       Impact factor: 5.269

6.  Polypyrrole-hydrogel composites for the construction of clinically important biosensors.

Authors:  Sean Brahim; Dyer Narinesingh; Anthony Guiseppi-Elie
Journal:  Biosens Bioelectron       Date:  2002-01       Impact factor: 10.618

7.  The enzyme electrode.

Authors:  S J Updike; G P Hicks
Journal:  Nature       Date:  1967-06-03       Impact factor: 49.962

8.  Highly selective microbiosensors for in vivo measurement of glucose, lactate and glutamate.

Authors:  O M Schuvailo; O O Soldatkin; A Lefebvre; R Cespuglio; A P Soldatkin
Journal:  Anal Chim Acta       Date:  2006-03-16       Impact factor: 6.558

9.  Photo-crosslinked copolymers of 2-hydroxyethyl methacrylate, poly(ethylene glycol) tetra-acrylate and ethylene dimethacrylate for improving biocompatibility of biosensors.

Authors:  C P Quinn; C P Pathak; A Heller; J A Hubbell
Journal:  Biomaterials       Date:  1995-03       Impact factor: 12.479

10.  A comparative physical study of two different hydrophilic synthetic latex matrices for the construction of a glucose biosensor.

Authors:  S Cosnier; S Szunerits; R S Marks; A Novoa; L Puech; E Perez; I Rico-Lattes
Journal:  Talanta       Date:  2001-12-13       Impact factor: 6.057

View more
  3 in total

Review 1.  Application of Semipermeable Membranes in Glucose Biosensing.

Authors:  Tanmay Kulkarni; Gymama Slaughter
Journal:  Membranes (Basel)       Date:  2016-12-14

2.  Cu and Ni Co-sputtered heteroatomic thin film for enhanced nonenzymatic glucose detection.

Authors:  Brianna Barbee; Baleeswaraiah Muchharla; Adetayo Adedeji; Abdennaceur Karoui; Kishor Kumar Sadasivuni; Mizaj Shabil Sha; Aboubakr M Abdullah; Gymama Slaughter; Bijandra Kumar
Journal:  Sci Rep       Date:  2022-05-07       Impact factor: 4.996

3.  Flexible battery-less wireless glucose monitoring system.

Authors:  Saikat Banerjee; Gymama Slaughter
Journal:  Sci Rep       Date:  2022-07-19       Impact factor: 4.996

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.