Finite element modeling of subcutaneous implantable defibrillator electrodes in an adult torso.

Matthew Jolley1, Jeroen Stinstra, Jess Tate, Steve Pieper, Rob Macleod, Larry Chu, Paul Wang, John K Triedman.   

Abstract

BACKGROUND: Total subcutaneous implantable subcutaneous defibrillators are in development, but optimal electrode configurations are not known.
OBJECTIVE: We used image-based finite element models (FEM) to predict the myocardial electric field generated during defibrillation shocks (pseudo-DFT) in a wide variety of reported and innovative subcutaneous electrode positions to determine factors affecting optimal lead positions for subcutaneous implantable cardioverter-defibrillators (S-ICD).
METHODS: An image-based FEM of an adult man was used to predict pseudo-DFTs across a wide range of technically feasible S-ICD electrode placements. Generator location, lead location, length, geometry and orientation, and spatial relation of electrodes to ventricular mass were systematically varied. Best electrode configurations were determined, and spatial factors contributing to low pseudo-DFTs were identified using regression and general linear models.
RESULTS: A total of 122 single-electrode/array configurations and 28 dual-electrode configurations were simulated. Pseudo-DFTs for single-electrode orientations ranged from 0.60 to 16.0 (mean 2.65 +/- 2.48) times that predicted for the base case, an anterior-posterior configuration recently tested clinically. A total of 32 of 150 tested configurations (21%) had pseudo-DFT ratios </=1, indicating the possibility of multiple novel, efficient, and clinically relevant orientations. Favorable alignment of lead-generator vector with ventricular myocardium and increased lead length were the most important factors correlated with pseudo-DFT, accounting for 70% of the predicted variation (R(2) = 0.70, each factor P < .05) in a combined general linear model in which parameter estimates were calculated for each factor.
CONCLUSION: Further exploration of novel and efficient electrode configurations may be of value in the development of the S-ICD technologies and implant procedure. FEM modeling suggests that the choice of configurations that maximize shock vector alignment with the center of myocardial mass and use of longer leads is more likely to result in lower DFT. Copyright 2010 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20230927      PMCID: PMC3103844          DOI: 10.1016/j.hrthm.2010.01.030

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  24 in total

Review 1.  Optical imaging of the heart.

Authors:  Igor R Efimov; Vladimir P Nikolski; Guy Salama
Journal:  Circ Res       Date:  2004-07-09       Impact factor: 17.367

2.  Long-term comparison of the implantable cardioverter defibrillator versus amiodarone: eleven-year follow-up of a subset of patients in the Canadian Implantable Defibrillator Study (CIDS).

Authors:  Fayez Bokhari; David Newman; Mary Greene; Victoria Korley; Iqwal Mangat; Paul Dorian
Journal:  Circulation       Date:  2004-07-06       Impact factor: 29.690

3.  The specific resistance of biological material--a compendium of data for the biomedical engineer and physiologist.

Authors:  L A Geddes; L E Baker
Journal:  Med Biol Eng       Date:  1967-05

4.  Minimally invasive cardioverter defibrillator implantation for children: an animal model and pediatric case report.

Authors:  C I Berul; J K Triedman; J Forbess; L M Bevilacqua; M E Alexander; D Dahlby; J O Gilkerson; E P Walsh
Journal:  Pacing Clin Electrophysiol       Date:  2001-12       Impact factor: 1.976

5.  A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators.

Authors:  A E Buxton; K L Lee; J D Fisher; M E Josephson; E N Prystowsky; G Hafley
Journal:  N Engl J Med       Date:  1999-12-16       Impact factor: 91.245

6.  A computer modeling tool for comparing novel ICD electrode orientations in children and adults.

Authors:  Matthew Jolley; Jeroen Stinstra; Steve Pieper; Rob Macleod; Dana H Brooks; Frank Cecchin; John K Triedman
Journal:  Heart Rhythm       Date:  2008-01-17       Impact factor: 6.343

7.  Incidence of ICD lead related complications during long-term follow-up: comparison of epicardial and endocardial electrode systems.

Authors:  T Korte; W Jung; S Spehl; C Wolpert; R Moosdorf; M Manz; B Lüderitz
Journal:  Pacing Clin Electrophysiol       Date:  1995-11       Impact factor: 1.976

8.  A comparison of biventricular and conventional transvenous defibrillation: a computational study using patient derived models.

Authors:  Daniel Mocanu; Joachim Kettenbach; Michael O Sweeney; Ron Kikinis; Bruce H Kenknight; Solomon R Eisenberg
Journal:  Pacing Clin Electrophysiol       Date:  2004-05       Impact factor: 1.976

9.  Incidence of venous obstruction following insertion of an implantable cardioverter defibrillator. A study of systematic contrast venography on patients presenting for their first elective ICD generator replacement.

Authors:  Lars Lickfett; Alexander Bitzen; Aravind Arepally; Khurram Nasir; Christian Wolpert; Kyung Mi Jeong; Ulf Krause; Rainer Schimpf; Thorsten Lewalter; Hugh Calkins; Werner Jung; Berndt Lüderitz
Journal:  Europace       Date:  2004-01       Impact factor: 5.214

10.  Computational studies of transthoracic and transvenous defibrillation in a detailed 3-D human thorax model.

Authors:  D B Jorgenson; D R Haynor; G H Bardy; Y Kim
Journal:  IEEE Trans Biomed Eng       Date:  1995-02       Impact factor: 4.538

View more
  17 in total

1.  Validating defibrillation simulation in a human-shaped phantom.

Authors:  Jess D Tate; Thomas A Pilcher; Kedar K Aras; Brett M Burton; Rob S MacLeod
Journal:  Heart Rhythm       Date:  2019-11-23       Impact factor: 6.343

2.  Measuring defibrillator surface potentials: The validation of a predictive defibrillation computer model.

Authors:  Jess Tate; Jeroen Stinstra; Thomas Pilcher; Ahrash Poursaid; Matthew A Jolley; Elizabeth Saarel; John Triedman; Rob S MacLeod
Journal:  Comput Biol Med       Date:  2018-08-29       Impact factor: 4.589

3.  Substernal lead implantation: a novel option to manage DFT failure in S-ICD patients.

Authors:  M Guenther; S Kolschmann; M Knaut
Journal:  Clin Res Cardiol       Date:  2014-10-02       Impact factor: 5.460

4.  Prospective evaluation of defibrillation threshold and postshock rhythm in young ICD recipients.

Authors:  Andrew E Radbill; John K Triedman; Charles I Berul; Edward P Walsh; Mark E Alexander; Gregory Webster; Frank Cecchin
Journal:  Pacing Clin Electrophysiol       Date:  2012-09-14       Impact factor: 1.976

Review 5.  How computer simulations of the human heart can improve anti-arrhythmia therapy.

Authors:  Natalia A Trayanova; Kelly C Chang
Journal:  J Physiol       Date:  2016-01-18       Impact factor: 5.182

6.  Placement of implantable cardioverter-defibrillators in paediatric and congenital heart defect patients: a pipeline for model generation and simulation prediction of optimal configurations.

Authors:  Lukas J Rantner; Fijoy Vadakkumpadan; Philip J Spevak; Jane E Crosson; Natalia A Trayanova
Journal:  J Physiol       Date:  2013-06-24       Impact factor: 5.182

7.  Temporal Dilation of Animal Cardiac Recordings Registered to Human Torso Geometries.

Authors:  Karli Gillette; Jess Tate; Brianna Kindall; Wilson Good; Jeff Wilkinson; Narendra Simha; Rob MacLeod
Journal:  Comput Cardiol (2010)       Date:  2017-03-02

8.  The subcutaneous defibrillator.

Authors:  Christopher P Rowley; S Suave Lobodzinski; Michael R Gold
Journal:  Curr Treat Options Cardiovasc Med       Date:  2012-10

9.  Virtual 3D heart models to aid pacemaker implantation in children.

Authors:  Natalia A Trayanova
Journal:  Future Cardiol       Date:  2014-01

10.  Verification of a Defibrillation Simulation Using Internal Electric Fields in a Human Shaped Phantom.

Authors:  Jess Tate; Thomas Pilcher; Kedar Aras; Brett Burton; Rob MacLeod
Journal:  Comput Cardiol (2010)       Date:  2014-09-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.