Literature DB >> 20230462

Actual pacemaker longevity: the benefit of stimulation by automatic capture verification.

Mauro Biffi1, Matteo Bertini, Davide Saporito, Matteo Ziacchi, Cristian Martignani, Igor Diemberger, Giuseppe Boriani.   

Abstract

BACKGROUND: We evaluated the impact of an algorithm for automatic right ventricular (RV) stimulation compared to fixed-output pacing (FOP) stimulation on actual pacemaker longevity over a 9-year follow-up.
METHODS: Prospective observation of 300 patients implanted with VDDR/DDDR pacemakers in 1999-2000 up to October 31, 2008. Sixty-one patients were paced by Autocapture pacing (ACP), 239 were paced by FOP; they were seen twice yearly at the pacemaker clinic. Factors known to affect pacemaker longevity were collected: median heart rate, %A&V paced activity, pacing output, and impedance. Patients dead before pacemaker replacement, lost to follow-up, or who developed permanent atrial fibrillation were excluded from analysis.
RESULTS: One hundred twenty-six of three hundred patients completed the study. Adverse clinical events due to an increased RV threshold occurred in two FOP patients compared to none among ACP. Pacemaker replacement occurred in 1/34 ACP patients versus 60/92 FOP patients (P < 0.001). ACP was the single independent predictor of pacemaker longevity at multivariable analysis (hazard ratio = 0.03, P < 0.001) either in the overall population or in the specific patients subgroups (sick sinus syndrome, atrioventricular block, and neurally mediated syncope).
CONCLUSIONS: Automatic verification of stimulation is reliable at long term, and warrants superior safety in the event of pacing threshold changes. It allows a significant longevity increase compared to FOP stimulation that may heavily impact the patients' quality of life and the cost of pacing therapy. Moreover, it is a fundamental technology in a strategy of remote patient and device monitoring, and may enable automatic device follow-up operated by trained, nonmedical personnel.

Entities:  

Mesh:

Year:  2010        PMID: 20230462     DOI: 10.1111/j.1540-8159.2010.02724.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  7 in total

1.  Clinical observation of atrial threshold monitoring algorithm: a single center experience.

Authors:  Jianqing She; Jing Zhou; Zhan Hu; Yulong Xia
Journal:  Int J Clin Exp Med       Date:  2015-04-15

Review 2.  Next-generation pacemakers: from small devices to biological pacemakers.

Authors:  Eugenio Cingolani; Joshua I Goldhaber; Eduardo Marbán
Journal:  Nat Rev Cardiol       Date:  2017-11-16       Impact factor: 32.419

Review 3.  Management of atrial fibrillation in bradyarrhythmias.

Authors:  Giuseppe Boriani; Luigi Padeletti
Journal:  Nat Rev Cardiol       Date:  2015-03-17       Impact factor: 32.419

Review 4.  Self-powered cardiovascular electronic devices and systems.

Authors:  Qiang Zheng; Qizhu Tang; Zhong Lin Wang; Zhou Li
Journal:  Nat Rev Cardiol       Date:  2020-09-07       Impact factor: 32.419

5.  Using an elastic magnifier to increase power output and performance of heart-beat harvesters.

Authors:  Antonio C Galbier; M Amin Karami
Journal:  Smart Mater Struct       Date:  2017-08-17       Impact factor: 3.585

6.  Impact of pacemaker longevity on expected device replacement rates: Results from computer simulations based on a multicenter registry (ESSENTIAL).

Authors:  Giuseppe Boriani; Matteo Bertini; Davide Saporito; Giuseppina Belotti; Fabio Quartieri; Corrado Tomasi; Angelo Pucci; Giulio Boggian; Gian Franco Mazzocca; Davide Giorgi; Paolo Diotallevi; Biagio Sassone; Diego Grassini; Alessio Gargaro; Mauro Biffi
Journal:  Clin Cardiol       Date:  2018-09-22       Impact factor: 2.882

7.  Automatic atrial capture device control in real-life practice: A multicenter experience.

Authors:  Massimo Giammaria; Gianluca Quirino; Mariangela Alberio; Umberto Parravicini; Eliana Cipolla; Guido Rossetti; Antonio Ruocco; Gaetano Senatore; Francesco Rametta; Paolo Pistelli
Journal:  J Arrhythm       Date:  2016-07-11
  7 in total