Karin Y Chumbimuni-Torres1, Percy Calvo-Marzal, Joseph Wang.
Abstract
Recent advances in ion-selective electrodes have pushed the detection limits of direct potentiometry to the nanomolar concentration range. Here we present a direct comparison of the sensitivity and selectivity of potentiometric and stripping-voltammetric measurements of cadmium and lead. While both techniques offer a similar sensitivity, the potentiometric method offers higher selectivity in the presence of excess of metal ions (e.g., thallium, tin) that commonly interfere in the stripping-voltammetric operation. Because of the complementary nature of the potentiometric and stripping-voltammetric methods, it is recommended that these techniques will be selected based on the specific analytical problem or used in parallel to provide additional analytical information.Entities: