Literature DB >> 20228885

Comparison Between Potentiometric and Stripping Voltammetric Detection of Trace Metals: Measurements of Cadmium and Lead in the Presence of Thalium, Indium, and Tin.

Karin Y Chumbimuni-Torres1, Percy Calvo-Marzal, Joseph Wang.   

Abstract

Recent advances in ion-selective electrodes have pushed the detection limits of direct potentiometry to the nanomolar concentration range. Here we present a direct comparison of the sensitivity and selectivity of potentiometric and stripping-voltammetric measurements of cadmium and lead. While both techniques offer a similar sensitivity, the potentiometric method offers higher selectivity in the presence of excess of metal ions (e.g., thallium, tin) that commonly interfere in the stripping-voltammetric operation. Because of the complementary nature of the potentiometric and stripping-voltammetric methods, it is recommended that these techniques will be selected based on the specific analytical problem or used in parallel to provide additional analytical information.

Entities:  

Year:  2009        PMID: 20228885      PMCID: PMC2836769          DOI: 10.1002/elan.200904613

Source DB:  PubMed          Journal:  Electroanalysis        ISSN: 1040-0397            Impact factor:   3.223


  6 in total

Review 1.  The new wave of ion-selective.

Authors:  Eric Bakker; Ernö Pretsch
Journal:  Anal Chem       Date:  2002-08-01       Impact factor: 6.986

2.  Potentiometric immunoassay with quantum dot labels.

Authors:  Reto Thürer; Tamás Vigassy; Martina Hirayama; Joseph Wang; Eric Bakker; Ernö Pretsch
Journal:  Anal Chem       Date:  2007-05-26       Impact factor: 6.986

3.  Solid contact potentiometric sensors for trace level measurements.

Authors:  Karin Y Chumbimuni-Torres; Nastassia Rubinova; Aleksandar Radu; Lauro T Kubota; Eric Bakker
Journal:  Anal Chem       Date:  2006-02-15       Impact factor: 6.986

4.  Electrochemical sample matrix elimination for trace-level potentiometric detection with polymeric membrane ion-selective electrodes.

Authors:  Karin Y Chumbimuni-Torres; Percy Calvo-Marzal; Joseph Wang; Eric Bakker
Journal:  Anal Chem       Date:  2008-06-21       Impact factor: 6.986

5.  Ion-transport and diffusion coefficients of non-plasticised methacrylic-acrylic ion-selective membranes.

Authors:  Lee Yook Heng; Klara Toth; Elizabeth A H Hall
Journal:  Talanta       Date:  2004-05-10       Impact factor: 6.057

6.  Aptamer-based potentiometric measurements of proteins using ion-selective microelectrodes.

Authors:  Apon Numnuam; Karin Y Chumbimuni-Torres; Yun Xiang; Ralph Bash; Panote Thavarungkul; Proespichaya Kanatharana; Ernö Pretsch; Joseph Wang; Eric Bakker
Journal:  Anal Chem       Date:  2008-01-10       Impact factor: 6.986

  6 in total