The reproducibility of test contractions for calibration of electromyographic measurements.

K B Veiersted1.   

Abstract

The main purpose of this study was to evaluate the reproducibility of electromyographic (EMG) measurements and specifically to test a calibration procedure with submaximal test contractions. Bipolar surface electrodes (20 mm fixed distance) were repositioned by a tracing sheet on both trapezius muscles, halfway between acromion and processus prominens. Submaximal test contractions were performed by keeping both arms straight abducted 90 degrees and forward flexed 10 degrees for 15-s periods. The arm position could be precisely reproduced in the frontal plane, but deviated forwards by 4 degrees in the horizontal plane, where the sensitivity of the EMG response to arm position was lowest. The electrodes were repositioned within a radius of 3 mm with a probability of 90%. Large deviations in the EMG response were found within this radius and a significant depression of the EMG response was recorded over the middle part of the muscle (the innervation zone?). This change in sensitivity of the EMG response with electrode position occurred in parallel for the test and maximal contractions. The total coefficient of variation was estimated to be 23% for recurrent EMG measurements using the calibration procedure described.

Entities:  

Mesh:

Year:  1991        PMID: 2022209     DOI: 10.1007/bf00626762

Source DB:  PubMed          Journal:  Eur J Appl Physiol Occup Physiol        ISSN: 0301-5548


  17 in total

1.  The relationship between myoelectric activity and muscle work in intermittent statical contractions.

Authors:  M Hagberg; B Jonsson
Journal:  Electromyogr Clin Neurophysiol       Date:  1975-07

2.  Signal characteristics of EMG with special reference to reproducibility of measurements.

Authors:  J H Viitasalo; P V Komi
Journal:  Acta Physiol Scand       Date:  1975-04

3.  Electromyographic activity in the shoulder-neck region according to arm position and glenohumeral torque.

Authors:  S E Mathiassen; J Winkel
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1990

4.  Pattern of muscle activity during stereotyped work and its relation to muscle pain.

Authors:  K B Veiersted; R H Westgaard; P Andersen
Journal:  Int Arch Occup Environ Health       Date:  1990       Impact factor: 3.015

5.  Experimental human muscle damage: morphological changes in relation to other indices of damage.

Authors:  D A Jones; D J Newham; J M Round; S E Tolfree
Journal:  J Physiol       Date:  1986-06       Impact factor: 5.182

6.  Activity levels during isometric test contractions of neck and shoulder muscles.

Authors:  K Schüldt; K Harms-Ringdahl
Journal:  Scand J Rehabil Med       Date:  1988

7.  Effects of skin electrode position on averaged electromyographic potentials.

Authors:  E N Zuniga; X T Truong; D G Simons
Journal:  Arch Phys Med Rehabil       Date:  1970-05       Impact factor: 3.966

8.  Reproducibility of electromyographic measurements with inserted wire electrodes and surface electrodes.

Authors:  P V Komi; E R Buskirk
Journal:  Electromyography       Date:  1970 Nov-Dec

9.  The influence of temperature on the amplitude and frequency components of the EMG during brief and sustained isometric contractions.

Authors:  J S Petrofsky; A R Lind
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1980

10.  Metabolic changes following eccentric exercise in trained and untrained men.

Authors:  W J Evans; C N Meredith; J G Cannon; C A Dinarello; W R Frontera; V A Hughes; B H Jones; H G Knuttgen
Journal:  J Appl Physiol (1985)       Date:  1986-11
View more
  15 in total

1.  Physical exposure of sign language interpreters: baseline measures and reliability analysis.

Authors:  Alain Delisle; Christian Larivière; Daniel Imbeau; Marie-José Durand
Journal:  Eur J Appl Physiol       Date:  2005-04-14       Impact factor: 3.078

2.  Effect of mechanical compression due to load carrying on shoulder muscle fatigue during sustained isometric arm abduction: an electromyographic study.

Authors:  Julien Piscione; Didier Gamet
Journal:  Eur J Appl Physiol       Date:  2006-06-10       Impact factor: 3.078

3.  Reliability of bloodflux measurements from the upper trapezius muscle during muscle contractions.

Authors:  Cecilie Røe; Elin Damsgård; Stein Knardahl
Journal:  Eur J Appl Physiol       Date:  2007-11-15       Impact factor: 3.078

4.  Reproducibility of kinetics of electromyogram spectrum parameters during dynamic exercise.

Authors:  D Gamet; J Duchêne; F Goubel
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1996

5.  Normalization of electromyogram in the neck-shoulder region.

Authors:  H Nieminen; E P Takala; E Viikari-Juntura
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1993

6.  The influence of electrode position on bipolar surface electromyogram recordings of the upper trapezius muscle.

Authors:  C Jensen; O Vasseljen; R H Westgaard
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1993

7.  Trapezius muscle load as a risk indicator for occupational shoulder-neck complaints.

Authors:  C Jensen; K Nilsen; K Hansen; R H Westgaard
Journal:  Int Arch Occup Environ Health       Date:  1993       Impact factor: 3.015

8.  Reproducibility of computer measurement of maximal isometric strength and electromyography in sedentary middle-aged women.

Authors:  A Heinonen; H Sievänen; J Viitasalo; M Pasanen; P Oja; I Vuori
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1994

Review 9.  An Evidence-Based Framework for Strengthening Exercises to Prevent Hamstring Injury.

Authors:  Matthew N Bourne; Ryan G Timmins; David A Opar; Tania Pizzari; Joshua D Ruddy; Casey Sims; Morgan D Williams; Anthony J Shield
Journal:  Sports Med       Date:  2018-02       Impact factor: 11.136

10.  Effect of side dominance on myoelectric manifestations of muscle fatigue in the human upper trapezius muscle.

Authors:  Dario Farina; Laura A C Kallenberg; Roberto Merletti; Hermie J Hermens
Journal:  Eur J Appl Physiol       Date:  2003-07-26       Impact factor: 3.078

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.