Literature DB >> 20206615

A rapid and fully-automated method for the quantitation of tricyclic antidepressants in serum using turbulent-flow liquid chromatography-tandem mass spectrometry.

Autumn R Breaud1, Robert Harlan, Joseph M Di Bussolo, Gwendolyn A McMillin, William Clarke.   

Abstract

BACKGROUND: We describe a fully-automated turbulent-flow liquid chromatography-tandem mass spectrometry method for the detection of tricyclic antidepressant drugs (amitriptyline, desipramine, imipramine, and nortriptyline) in serum.
METHODS: Human serum and an internal standard were injected directly onto a Cyclone-P online solid-phase extraction (SPE) column (0.5 x 50 mm). Following removal of serum proteins and other components the analytes were transferred to a Hypersil Gold C-18 (50 x 3 mm) analytical column. Elution occurred with a gradient of water and acetonitrile each with 0.1% formic acid. Analytes were ionized and detected over a 3.5 min analysis time by electrospray-ionization mass spectrometry with selected reaction monitoring (SRM). Matrix effects were well-characterized and carryover, precision, linearity, recovery and limits of detection and quantitation were evaluated.
RESULTS: The simple and complex precision CVs for all compounds were < or = 16%. The limits of detection and quantitation for all drugs were < or = 3 ng/ml and <20 ng/ml, respectively. Recoveries were between 97 and 114%. Slopes for method comparison plots were all >0.96. Proficiency testing materials had values within 2 SDI of peer group means for all drugs.
CONCLUSION: Based on validation data, this is a specific, sensitive fully-automated method for rapid quantitation of tricyclic antidepressants in serum. Copyright 2010 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20206615     DOI: 10.1016/j.cca.2010.02.067

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  2 in total

1.  A Simple, Rapid and Reliable Method to Determine Imipramine and Desipramine in Mouse Serum Using Ultra-High-Performance Liquid Chromatography-Quadrupole-Time-of-Flight Mass Spectrometry.

Authors:  Jing Zhao; Yujin Shin; Kwang-Hoon Chun; Hye-Ran Yoon; Jeongmi Lee
Journal:  J Chromatogr Sci       Date:  2015-12-19       Impact factor: 1.618

2.  Alternative calibration strategies for the clinical laboratory: application to nortriptyline therapeutic drug monitoring.

Authors:  Matthew T Olson; Autumn Breaud; Robert Harlan; Nkechinyere Emezienna; Sabitha Schools; Alfred L Yergey; William Clarke
Journal:  Clin Chem       Date:  2013-02-20       Impact factor: 8.327

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.