Alternative calibration strategies for the clinical laboratory: application to nortriptyline therapeutic drug monitoring.

Matthew T Olson1, Autumn Breaud, Robert Harlan, Nkechinyere Emezienna, Sabitha Schools, Alfred L Yergey, William Clarke.   

Abstract

BACKGROUND: The addition of a calibration curve with every run is both time-consuming and expensive for clinical mass spectrometry assays. We present alternative calibration strategies that eliminate the need for a calibration curve except as required by laboratory regulations.
METHODS: We measured serum nortriptyline concentrations prospectively in 68 patients on 16 days over a 2-month period using a method employing calibration schemes that relied on the measurement of the response ratio (RR) corrected by the response factor (RF), i.e., a measurement of the RR for an equimolar mixture of the analyte and internal standard. Measurements were taken with contemporaneous RF (cRF) measurements as well as sporadic RF (sRF) measurements. The measurements with these alternative calibration schemes were compared against the clinical results obtained by interpolation on a calibration curve, and those differences were evaluated for analytical and clinical significance.
RESULTS: The differences between the values measured by cRF and sRF calibration and interpolation on a calibration curve were not significant (P = 0.088 and P = 0.091, respectively). Both the cRF- and sRF-based calibration results demonstrated a low mean bias against the calibration curve interpolations of 3.69% (95% CI, -15.8% to 23.2%) and 3.11% (95% CI, -16.4% to 22.6%), respectively. When these results were classified as subtherapeutic, therapeutic, or supratherapeutic, there was categorical agreement in 95.6% of the cRF calibration results and 94.1% of the sRF results.
CONCLUSIONS: cRF and sRF calibration in a clinically validated liquid chromatography-tandem mass spectrometry assay yields results that are analytically and clinically commensurate to those produced by interpolation with a calibration curve.
© 2013 American Association for Clinical Chemistry.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23426427      PMCID: PMC4162085          DOI: 10.1373/clinchem.2012.194639

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  20 in total

1.  Successful and cost-efficient replacement of immunoassays by tandem mass spectrometry for the quantification of immunosuppressants in the clinical laboratory.

Authors:  Pierre-Olivier Hétu; Robert Robitaille; Bernard Vinet
Journal:  J Chromatogr B Analyt Technol Biomed Life Sci       Date:  2011-11-03       Impact factor: 3.205

2.  Liquid chromatography-tandem mass spectrometry or automated immunoassays: what are the future trends in therapeutic drug monitoring?

Authors:  Gunnar Brandhorst; Michael Oellerich; Gregory Maine; Paul Taylor; Gerhard Veen; Pierre Wallemacq
Journal:  Clin Chem       Date:  2012-01-17       Impact factor: 8.327

Review 3.  A statistical overview on univariate calibration, inverse regression, and detection limits: Application to gas chromatography/mass spectrometry technique.

Authors:  Irma Lavagnini; Franco Magno
Journal:  Mass Spectrom Rev       Date:  2007 Jan-Feb       Impact factor: 10.946

4.  Direct quantification in bioanalytical LC-MS/MS using internal calibration via analyte/stable isotope ratio.

Authors:  Lars B Nilsson; Göran Eklund
Journal:  J Pharm Biomed Anal       Date:  2006-11-02       Impact factor: 3.935

5.  Treatment of aspergillosis: clinical practice guidelines of the Infectious Diseases Society of America.

Authors:  Thomas J Walsh; Elias J Anaissie; David W Denning; Raoul Herbrecht; Dimitrios P Kontoyiannis; Kieren A Marr; Vicki A Morrison; Brahm H Segal; William J Steinbach; David A Stevens; Jo-Anne van Burik; John R Wingard; Thomas F Patterson
Journal:  Clin Infect Dis       Date:  2008-02-01       Impact factor: 9.079

Review 6.  Standards of laboratory practice: antidepressant drug monitoring. National Academy of Clinical Biochemistry.

Authors:  M W Linder; P E Keck
Journal:  Clin Chem       Date:  1998-05       Impact factor: 8.327

7.  Alternatives to least squares linear regression analysis for computation of standard curves for quantitation by high performance liquid chromatography: applications to clinical pharmacology.

Authors:  G K Szabo; H K Browne; A Ajami; E G Josephs
Journal:  J Clin Pharmacol       Date:  1994-03       Impact factor: 3.126

8.  Therapeutic drug monitoring of antidepressants in haemodialysis patients.

Authors:  Stefan Unterecker; Philip Müller; Christian Jacob; Peter Riederer; Bruno Pfuhlmann
Journal:  Clin Drug Investig       Date:  2012-08-01       Impact factor: 2.859

9.  Pharmacokinetics of busulfan: correlation with veno-occlusive disease in patients undergoing bone marrow transplantation.

Authors:  L B Grochow; R J Jones; R B Brundrett; H G Braine; T L Chen; R Saral; G W Santos; O M Colvin
Journal:  Cancer Chemother Pharmacol       Date:  1989       Impact factor: 3.333

10.  Clinical practice guidelines for the management of patients with histoplasmosis: 2007 update by the Infectious Diseases Society of America.

Authors:  L Joseph Wheat; Alison G Freifeld; Martin B Kleiman; John W Baddley; David S McKinsey; James E Loyd; Carol A Kauffman
Journal:  Clin Infect Dis       Date:  2007-08-27       Impact factor: 9.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.