Literature DB >> 20206045

Evaluation of a new femoral fixation device in a simulated anterior cruciate ligament reconstruction.

Andrew Speirs1, David Simon, Peter Lapner.   

Abstract

PURPOSE: The purpose of this in vitro biomechanical study was to determine the cyclic elongation and failure properties of a new anterior cruciate ligament (ACL) reconstruction device and compare the results with several devices that are currently available.
METHODS: We performed 10 ACL reconstructions in 4 groups using fresh porcine femurs and doubled lateral extensor of the toes tendons. Manufacturer guidelines were followed for fixation by use of either of 2 cortical suspension devices (XO Button [ConMed Linvatec, Largo, FL] and EndoButton CL [Smith & Nephew, Andover, MA]), a bio-interference screw (BioScrew; ConMed Linvatec), or a corticocancellous fixation device (Pinn-ACL; ConMed Linvatec). Reconstructions were subjected to cyclic loading to 150 N for 2,000 cycles, followed by static failure tests.
RESULTS: The two cortical suspension devices performed similarly to one another: the XO Button device had a significantly lower elongation amplitude than the EndoButton (P < .05). There were no significant differences in longer-term creep performance or static strength or stiffness. Compared with an interference screw, the XO Button had significantly less creep and higher failure load (P < .05). The corticocancellous device had the lowest creep and cyclic elongation amplitude and the highest strength and stiffness of the devices tested.
CONCLUSIONS: In this in vitro evaluation, reconstructions with the XO Button and EndoButton exhibited very similar biomechanical performance, and our hypothesis was not supported: the XO Button did not limit creep more than the EndoButton. CLINICAL RELEVANCE: The results of this preclinical in vitro testing suggest that the new device is expected to provide clinical results similar to those of the EndoButton, a well-established device for ACL reconstruction. Copyright 2010 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20206045     DOI: 10.1016/j.arthro.2009.08.016

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  3 in total

1.  Single-Bundle Anterior Cruciate Ligament Reconstruction with Semitendinosus Tendon Using the PINN-ACL CrossPin System: Minimum 4-Year Follow-up.

Authors:  Hee-Soo Kyung; Seung-Gil Baek; Byoung-Joo Lee; Chang-Hwa Lee
Journal:  Knee Surg Relat Res       Date:  2015-03-02

2.  Isokinetic muscle strength and knee function associated with double femoral pin fixation and fixation with interference screw in anterior cruciate ligament reconstruction.

Authors:  Tiago Lazzaretti Fernandes; Thiago Rocha Protta; Felipe Fregni; Raul Bolliger Neto; André Pedrinelli; Gilberto Luis Camanho; Arnaldo José Hernandez
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-06-28       Impact factor: 4.342

3.  ACL Reconstruction with Autologous Hamstring Tendon: Comparison of Short Term Clinical Results between Rigid-fix and PINN-ACL Cross Pin.

Authors:  Seung-Suk Seo; Chang-Wan Kim; Tae-Seok Nam; Sang-Yeong Choi
Journal:  Knee Surg Relat Res       Date:  2011-11-30
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.