Thermal comparison of Infiniti OZil and Signature Ellips phacoemulsification systems.

Joseph S Schmutz1, Randall J Olson.   

Abstract

PURPOSE: To determine thermal characteristics of Signature Ellips (Abbott Medical Optics) and Infiniti OZil (Alcon, Inc.) transverse ultrasound and compare both with longitudinal ultrasound in clinically relevant scenarios.
DESIGN: Laboratory investigation.
METHODS: Temperature increase over baseline after 60 seconds was measured in water at positions in 90-degree increments around the sleeve near the proximal needle shaft in an artificial chamber for Ellips and OZil on continuous ultrasound with aspiration blocked and unblocked. This was also done with Signature using longitudinal ultrasound, with and without micropulse (6 ms on, 12 ms off), with aspiration blocked and unblocked, and at the OZil sleeve tip on continuous transverse mode with aspiration unblocked.
RESULTS: OZil (8.1 +/- 0.3 C) had greater temperature increase than Ellips (5.2 +/- 0.3 C; P < .0001) with aspiration unblocked and blocked (29.3 +/- 1.0 C vs 12.2 +/- 0.7 C; P < .0001). OZil had uneven distribution of heat around the shaft (30.1 +/- 0.5 C vs 28.5 +/- 0.6 C; P < .0001), whereas Ellips did not (P = .57). OZil was cooler at the tip (6.6 +/- 0.2 C; P < .0001). Friction in a cadaver eye incision only increased these numbers by 10% (OZil, irrigation blocked).
CONCLUSIONS: Metal stress probably creates heat at the proximal needle junction for both transverse methods. Heat generation differences between OZil and Ellips result from the manner in which they create needle motion. Incision burns may occur, especially for OZil, under nonpulsed settings during fragment removal with occlusion when reaching across the anterior chamber such that the proximal needle shaft came near the wound. Copyright 2010 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20202619     DOI: 10.1016/j.ajo.2009.12.006

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  5 in total

1.  Comparison of corneal endothelial changes following phacoemulsification with transversal and torsional phacoemulsification machines.

Authors:  Mustafa Ataş; Süleyman Demircan; Arzu Seyhan Karatepe Haşhaş; Ahmet Gülhan; Gökmen Zararsız
Journal:  Int J Ophthalmol       Date:  2014-10-18       Impact factor: 1.779

2.  Thermal imaging comparison of Signature, Infiniti, and Stellaris phacoemulsification systems.

Authors:  Na Kyung Ryoo; Ji-Won Kwon; Won Ryang Wee; Kevin M Miller; Young Keun Han
Journal:  BMC Ophthalmol       Date:  2013-10-12       Impact factor: 2.209

3.  Wound and surface temperatures in vivo in torsional and longitudinal modalities of ultrasound in coaxial microincisional cataract surgery.

Authors:  Swapna Nair; Ramachandran Unnikrishnan Nair
Journal:  Clin Ophthalmol       Date:  2017-01-27

4.  Comparative analysis of corneal morphological changes after transversal and torsional phacoemulsification through 2.2 mm corneal incision.

Authors:  Ahmed Assaf; Maged Maher Roshdy
Journal:  Clin Ophthalmol       Date:  2013-01-08

5.  Transverse vs torsional ultrasound: prospective randomized contralaterally controlled study comparing two phacoemulsification-system handpieces.

Authors:  Kerry K Assil; Lindsay Harris; Jeannie Cecka
Journal:  Clin Ophthalmol       Date:  2015-08-03
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.