Evaluation of in vitro measurement accuracy of a three-dimensional ultrasound scanner.

D L King1, D L King1, M Y Shao.   

Abstract

We have constructed a three-dimensional (3-D) ultrasound scanner to make examinations easier and more accurate. The accuracy of the 3-D scanner has been determined by scanning a three-dimensional model. Lengths, angles, and volume of the model were computed and compared to values determined from physical measurement. Distances between image planes were computed with a mean error of 0.4% of true value. Volume was computed with a mean error of 1.6%, or 0.64 mL +/- 0.72 mL (1 SD). We conclude that the 3-D scanner has a system error of less than 0.4%, that it does not introduce significant new errors, and that the principal source of measurement error is poor lateral resolution due to ultrasound beam width. We believe that the 3-D scanner may be a clinically useful instrument for quantitative echographic measurements.

Mesh:

Year:  1991        PMID: 2020053     DOI: 10.7863/jum.1991.10.2.77

Source DB:  PubMed          Journal:  J Ultrasound Med        ISSN: 0278-4297            Impact factor:   2.153


  3 in total

1.  Three-dimensional echocardiographic measurement of left and right ventricular mass and volume: in vitro validation.

Authors:  Michal Hubka; Edward L Bolson; John A McDonald; Roy W Martin; Brad Munt; Florence H Sheehan
Journal:  Int J Cardiovasc Imaging       Date:  2002-04       Impact factor: 2.357

2.  Quantification of mitral valve stenosis by three-dimensional transesophageal echocardiography.

Authors:  I Kupferwasser; S Mohr-Kahaly; T Menzel; M Spiecker; G Dohmen; E Mayer; H Oelert; R Erbel; J Meyer
Journal:  Int J Card Imaging       Date:  1996-12

3.  Three-dimensional ultrasound scanning.

Authors:  Aaron Fenster; Grace Parraga; Jeff Bax
Journal:  Interface Focus       Date:  2011-06-01       Impact factor: 3.906

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.