Comparison of ICare rebound tonometer with noncontact tonometer in healthy children.

Mitsuyo Kageyama1, Kazuyuki Hirooka, Tetsuya Baba, Fumio Shiraga.   

Abstract

PURPOSE: To evaluate the effectiveness of using a noncontact tonometer (NCT) versus a rebound tonometer (ICare) when measuring the intraocular pressure (IOP) in healthy children.
DESIGN: Prospective observational study.
METHODS: A total of 180 (96 males and 84 females) healthy children, aged 6 months to 15 years, were recruited. IOPs were measured using both the NCT and ICare devices in the absence of anesthesia. A successful outcome was defined as the measurement of the IOP in both eyes. The McNemar test for comparing correlated proportions was used to analyze the ICare and NCT data.
RESULTS: Although the IOP was successfully measured in 160 children (88.9%) when using ICare, we were only able to successfully measure 130 children (72.2%) when using NCT. Below the age of 6 years, measurement of the IOP using ICare was better tolerated as compared with the NCT (McNemar test; P<0.001). The mean differences (95% limits of agreement) for the IOP readings between ICare and NCT in right and left eyes were 0.90±6.40 and 1.18±6.19 mm Hg, respectively.
CONCLUSIONS: IOP measurements performed using ICare are better tolerated in the pediatric population, as compared with measurements using NCT, especially in children below the age of 6 years.

Entities:  

Mesh:

Year:  2011        PMID: 20179620     DOI: 10.1097/IJG.0b013e3181d12dc4

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  12 in total

1.  Age variations in intraocular pressure in a cohort of healthy Austrian school children.

Authors:  W A Dusek; B K Pierscionek; J F McClelland
Journal:  Eye (Lond)       Date:  2012-03-23       Impact factor: 3.775

2.  Intraocular Pressure Measurement in Childhood Glaucoma under Standardized General Anaesthesia: The Prospective EyeBIS Study.

Authors:  Alicja Strzalkowska; Nina Pirlich; Julia V Stingl; Alexander K Schuster; Jasmin Rezapour; Felix M Wagner; Justus Buse; Esther M Hoffmann
Journal:  J Clin Med       Date:  2022-05-18       Impact factor: 4.964

3.  Comprasion of ICare rebound tonometer and Goldmann applanation tonometer in high myopia.

Authors:  Halil Huseyin Cagatay; Metin Ekinci; Zeliha Yazar; Gokcen Gokce; Erdinc Ceylan
Journal:  ScientificWorldJournal       Date:  2014-10-14

4.  Comparison of Goldmann applanation tonometer, Tono-Pen and noncontact tonometer in children.

Authors:  Usha Kaul Raina; Neha Rathie; Anika Gupta; Shantanu Kumar Gupta; Meenakshi Thakar
Journal:  Oman J Ophthalmol       Date:  2016 Jan-Apr

5.  Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients.

Authors:  Shaoying Tan; Marco Yu; Nafees Baig; Linda Hansapinyo; Clement C Tham
Journal:  Sci Rep       Date:  2017-02-06       Impact factor: 4.379

6.  Intraocular pressure monitoring by rebound tonometry in children with myopia.

Authors:  Jenchieh Weng; I-Lun Tsai; Li-Lin Kuo; Ching-Yao Tsai; Lin-Chung Woung; Ya-Chuan Hsiao
Journal:  Taiwan J Ophthalmol       Date:  2017 Jul-Sep

7.  Comparison of Icare Pro Tonometry and Icare One Tonometry Measurements in Healthy Eyes

Authors:  Hüseyin Mayalı; Çağlar Sarıgül; Emin Kurt; Özcan Rasim Kayıkçıoğlu; Süleyman Sami İlker
Journal:  Turk J Ophthalmol       Date:  2019-06-27

8.  Comparison of handheld rebound tonometry with Goldmann applanation tonometry in children with glaucoma: a cohort study.

Authors:  Annegret Hella Dahlmann-Noor; Renata Puertas; Shenille Tabasa-Lim; Ahmed El-Karmouty; Mustafa Kadhim; Nicholas Kloster Wride; Amanda Lewis; Dawn Grosvenor; Poornima Rai; Maria Papadopoulos; John Brookes; Catey Bunce; Peng Tee Khaw
Journal:  BMJ Open       Date:  2013-04-02       Impact factor: 2.692

9.  Intraocular pressure in a cohort of healthy eastern European schoolchildren: variations in method and corneal thickness.

Authors:  Patrycja Krzyżanowska-Berkowska; Magdalena Asejczyk-Widlicka; Barbara Pierscionek
Journal:  BMC Ophthalmol       Date:  2012-12-02       Impact factor: 2.209

10.  New intraocular pressure measurement method using reflected pneumatic pressure from cornea deformed by air puff of ring-type nozzle.

Authors:  Hyung Jin Kim; Yeong Ho Seo; Byeong Hee Kim
Journal:  PLoS One       Date:  2017-12-07       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.