| Literature DB >> 20177464 |
N J Shah1, B N Suhagia, R R Shah, N M Patel.
Abstract
A simple, precise, accurate and rapid high performance thin layer chromatographic method has been developed and validated for the simultaneous estimation of valsartan and hydrochlorothiazide in combined dosage forms. The stationary phase used was precoated silica gel 60F(254). The mobile phase used was a mixture of chloroform: methanol: toluene: glacial acetic acid (6:2:1:0.1 v/v/v/v). The detection of spots were carried out at 260 nm. The method was validated in terms of linearity, accuracy, precision and specificity. The calibration curve was found to be linear between 300 to 800 ng/spot for valsartan and 100 to 600 ng/spot for hydrochlorothiazide. The limit of detection and the limit of quantification for the valsartan were found to be 100 and 300 ng/spot respectively and for hydrochlorothiazide 30 and 100 ng/spot respectively. The proposed method can be successfully used to determine the drug content of marketed formulation.Entities:
Keywords: HPTLC; Valsartan; hydrochlorothiazide; simultaneous estimation
Year: 2009 PMID: 20177464 PMCID: PMC2810056 DOI: 10.4103/0250-474X.51967
Source DB: PubMed Journal: Indian J Pharm Sci ISSN: 0250-474X Impact factor: 0.975
Fig. 1A typical HPTLC chromatogram of valsartan and hydrochlorothiazide.
METHOD VALIDATION PARAMETERS OF PROPOSED METHOD
| Parameters | Values | |
|---|---|---|
| Valsartan | Hydrochlorothiazide | |
| Linearity range (ng/spot) | 300-800 | 100-600 |
| Correlation coefficient (r) | 0.9961 | 0.998 |
| Regression equation (y=mx+c) | ||
| Slope (m) | 3.0779 | 4.1895 |
| Intercept (c) | 43.295 | 2.56 |
| Limit of detection (LOD) | 100 ng/spot | 30 ng/spot |
| Limit of quantification (LOQ) | 300 ng/spot | 100 ng/spot |
| Precision (% CV) | ||
| Repeatability of application (n=5) | 0.98 | 0.83 |
| Repeatability of measurement (n=5) | 0.43 | 0.49 |
ASSAY AND RECOVERY STUDIES OF VALSARTAN AND HYDROCHLOROTHIAZIDE
| Brand name | Label Claim mg / tablet | Total Amount added (mg) | Amount recovered | % Recovery ± SD | % Assay |
|---|---|---|---|---|---|
| Valzarr-H | Valsartan 80 | 40 | 39.83± 0.98 | 99.59±0.98 | |
| 80 | 79.28±1.50 | 99.10±1.50 | 100.78±0.43 | ||
| 120 | 120.70±0.79 | 100.59±0.79 | |||
| Hydrochlorothiazide | 6.25 | 6.27±1.69 | 100.87±1.69 | ||
| 12.5 | 12.50 | 12.32±0.95 | 99.92±0.95 | 99.81±0.85 | |
| 18.75 | 19.13±0.87 | 102.06±0.87 |
Each value is a mean ± standard deviation of three determinations. Valzarr-H is a brand of Torrent Pharmaceutical Ltd.