X Ding1, B Chou, R A Graham, S Cheeti, S Percey, L C Matassa, S A Reuschel, M Meng, S Liu, T Voelker, B L Lum, P J Rudewicz, C E C A Hop.
Abstract
To support clinical development, a solid phase extraction (SPE) liquid chromatographic-tandem mass spectrometry (LC-MS/MS) method for the determination of GDC-0449 concentrations in human plasma has been developed and validated. Samples (200 microl) were extracted using an Oasis MCX 10 mg 96-well SPE plate and the resulting extracts were analyzed using reverse-phase chromatography coupled with a turbo-ionspray interface. The method was validated over calibration curve range 5-5000 ng/mL. Quadratic regression and 1/x(2) weighing were used. Within-run relative standard deviation (%RSD) was within 10.1% and accuracy ranged from 88.6% to 109.0% of nominal. Between-run %RSD was within 8.6% and accuracy ranged from 92.4% to 105.3% of nominal. Extraction recovery of GDC-0449 was between 88.3% and 91.2% as assessed using quality control sample concentrations. GDC-0449 was stable in plasma for 315 days when stored at -70 degrees C and stable in reconstituted sample extracts for 117 h when stored at room temperature. Quantitative matrix effect/ion suppression experiment was performed and no significant matrix ion suppression was observed. This assay allows for the determination of GDC-0449 plasma concentrations over a sufficient time period to determine pharmacokinetic parameters at relevant clinical doses. 2010 Elsevier B.V. All rights reserved.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20172765 DOI: 10.1016/j.jchromb.2010.01.039
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205