Literature DB >> 20170853

EOS orthopaedic imaging system to study patellofemoral kinematics: assessment of uncertainty.

C Azmy1, S Guérard, X Bonnet, F Gabrielli, W Skalli.   

Abstract

BACKGROUND: Accurate knowledge of knee joint kinematics, especially patellofemoral joint kinematics,is essential for prosthetic evaluation so as to further improve total knee arthroplasty performances. Improving the evaluation of the functioning of the extensor apparatus appears,in this respect, particularly important in this optimization effort.
OBJECTIVES: The aim of this study was to propose a new experimental setup for the analysis of knee joint kinematics and to validate its relevance in terms of accuracy and uncertainty.The technique developed herein combines 3D reconstruction imaging with the use of a motion capture system.
MATERIAL AND METHODS: Eight pairs of fresh-frozen cadaver specimens with no evidence of previous knee surgery were studied using a new test rig where the femur remains fixed and the tibia is free to rotate. The flexion-extension cycles were executed using computer-controlled traction of the quadriceps tendon combined with an antagonist force applied to the distal part of the tibia. Knee joint kinematics were tracked using an optoelectronic motion capture system after a preliminary stage of data acquisition of bone geometry and markers position. This stage was carried out using a new digital stereophotogrammetric system, EOS, combined with specific 3D reconstruction software that also determined the coordinate system used in the kinematic analysis. The resulting uncertainty was assessed as was its impact on the estimated kinematics.
RESULTS: Test results on eight knees validated the setup designed for the analysis of knee joint kinematics during the flexion-extension cycle. More specifically, the statistical results show that measurement uncertainty for rotations and translations remains below 0.4 and 1.8 mm,respectively, for the tibia and 0.4 and 1.2 mm for the patella (+/- 2 S.D. for all four measurements). DISCUSSION: The combination of 3D imaging and motion capture enables the proposed method to track the real-time motion of any bone segment during knee flexion-extension cycle. In particular,the new test rig introduced in this paper allows in vitro measurements of the patello femoral and tibiofemoral kinematics with a good level of accuracy. Moreover, this personalized experimental analysis can provide a more objective approach to the evaluation of knee implants as well as the validation of the finite-elements-based models of the patellofemoral joint. 2010 Published by Elsevier Masson SAS.

Entities:  

Mesh:

Year:  2010        PMID: 20170853     DOI: 10.1016/j.rcot.2009.12.003

Source DB:  PubMed          Journal:  Orthop Traumatol Surg Res        ISSN: 1877-0568            Impact factor:   2.256


  6 in total

1.  Three-dimensional measurements of the lower extremity in children and adolescents using a low-dose biplanar X-ray device.

Authors:  Ramon Gheno; Eric Nectoux; Bernard Herbaux; Matteo Baldisserotto; Luiz Glock; Anne Cotten; Nathalie Boutry
Journal:  Eur Radiol       Date:  2011-10-20       Impact factor: 5.315

Review 2.  Recent advances in medical imaging: anatomical and clinical applications.

Authors:  Bruno Grignon; Laurence Mainard; Matthieu Delion; Claude Hodez; Guillaume Oldrini
Journal:  Surg Radiol Anat       Date:  2012-05-29       Impact factor: 1.246

3.  Three-dimensional quantitative analysis of the proximal femur and the pelvis in children and adolescents using an upright biplanar slot-scanning X-ray system.

Authors:  Kinga Szuper; Ádám Tibor Schlégl; Eleonóra Leidecker; Csaba Vermes; Szabolcs Somoskeöy; Péter Than
Journal:  Pediatr Radiol       Date:  2014-08-26

4.  Reliability of overcoverage parameters with varying morphologic pincer features: comparison of EOS® and radiography.

Authors:  Shafagh Monazzam; Mandar Agashe; Harish S Hosalkar
Journal:  Clin Orthop Relat Res       Date:  2013-05-09       Impact factor: 4.176

5.  EOS imaging versus current radiography: A health technology assessment study.

Authors:  Alireza Mahboub-Ahari; Sakineh Hajebrahimi; Mahmoud Yusefi; Ashraf Velayati
Journal:  Med J Islam Repub Iran       Date:  2016-02-17

6.  EOS(®) biplanar X-ray imaging: concept, developments, benefits, and limitations.

Authors:  Elias Melhem; Ayman Assi; Rami El Rachkidi; Ismat Ghanem
Journal:  J Child Orthop       Date:  2016-02-16       Impact factor: 1.548

  6 in total