Does "hidden undercuffing" occur among obese patients? Effect of arm sizes and other predictors of the difference between wrist and upper arm blood pressures.

Hardik Doshi1, Alan B Weder, Robert L Bard, Robert D Brook.   

Abstract

Arm size can affect the accuracy of blood pressure (BP) measurement, and "undercuffing" of large upper arms is likely to be a growing problem. Therefore, the authors investigated the relationship between upper arm and wrist readings. Upper arm and wrist circumferences and BP were measured in 261 consecutive patients. Upper arm auscultation and wrist BP was measured in triplicate, rotating measurements every 30 seconds between sites. Upper arm BP was 131.9+/-20.6/71.6+/-12.6 mm Hg in an obese population (body mass index, 30.6+/-6.6 kg/m(2)) with mean upper arm size of 30.7+/-5.1 cm. Wrist BP was higher (2.6+/-9.2 mm Hg and 4.9+/-6.6 mm Hg, respectively, P<.001); however, there was moderate concordance for the Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure (JNC 7) strata (kappa value=0.27-0.71), and the difference was >or=5 mm Hg in 72% of the patients. The authors conclude that there was poor concordance between arm and wrist BP measurement and found no evidence that "hidden undercuffing" was associated with obesity; therefore, they do not support routine use of wrist BP measurements.

Entities:  

Mesh:

Year:  2010        PMID: 20167030      PMCID: PMC8673187          DOI: 10.1111/j.1751-7176.2009.00222.x

Source DB:  PubMed          Journal:  J Clin Hypertens (Greenwich)        ISSN: 1524-6175            Impact factor:   3.738


  19 in total

1.  Blood pressure measuring devices: recommendations of the European Society of Hypertension.

Authors:  E O'Brien; B Waeber; G Parati; J Staessen; M G Myers
Journal:  BMJ       Date:  2001-03-03

2.  Clinical comparison of automatic, noninvasive measurements of blood pressure in the forearm and upper arm.

Authors:  Kathleen Schell; Elisabeth Bradley; Linda Bucher; Maureen Seckel; Denise Lyons; Sandra Wakai; Deborah Bartell; Elizabeth Carson; Melanie Chichester; Teresa Foraker; Kathleen Simpson
Journal:  Am J Crit Care       Date:  2005-05       Impact factor: 2.228

3.  Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in adults and obese adults.

Authors:  Sekip Altunkan; Kamuran Oztaş; Erkan Altunkan
Journal:  Blood Press Monit       Date:  2006-04       Impact factor: 1.444

4.  Wrist blood pressure-measuring devices: a comparative study of accuracy with a standard auscultatory method using a mercury manometer.

Authors:  Sekip Altunkan; Sevil Yildiz; Sabir Azer
Journal:  Blood Press Monit       Date:  2002-10       Impact factor: 1.444

5.  Validation of the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure measuring devices in adults according to the International Protocol.

Authors:  Sekip Altunkan; Nevzat Ilman; Erkan Altunkan
Journal:  Blood Press Monit       Date:  2007-04       Impact factor: 1.444

6.  The changing distribution of arm circumferences in NHANES III and NHANES 2000 and its impact on the utility of the 'standard adult' blood pressure cuff.

Authors:  John W Graves; Kent R Bailey; Sheldon G Sheps
Journal:  Blood Press Monit       Date:  2003-12       Impact factor: 1.444

7.  The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure: the JNC 7 report.

Authors:  Aram V Chobanian; George L Bakris; Henry R Black; William C Cushman; Lee A Green; Joseph L Izzo; Daniel W Jones; Barry J Materson; Suzanne Oparil; Jackson T Wright; Edward J Roccella
Journal:  JAMA       Date:  2003-05-14       Impact factor: 56.272

8.  Can validated wrist devices with position sensors replace arm devices for self-home blood pressure monitoring? A randomized crossover trial using ambulatory monitoring as reference.

Authors:  George S Stergiou; George R Christodoulakis; Efthimia G Nasothimiou; Periklis P Giovas; Petros G Kalogeropoulos
Journal:  Am J Hypertens       Date:  2008-04-24       Impact factor: 2.689

9.  Clinical evaluation of an automated oscillometric blood pressure wrist device.

Authors:  Alexandra C Cuckson; Paul Moran; Paul Seed; Annemarie Reinders; Andrew H Shennan
Journal:  Blood Press Monit       Date:  2004-02       Impact factor: 1.444

10.  Evaluation of the performance of a wrist blood pressure measuring device with a position sensor compared to ambulatory 24-hour blood pressure measurements.

Authors:  Sakir Uen; Burkard Weisser; Paul Wieneke; Hans Vetter; Thomas Mengden
Journal:  Am J Hypertens       Date:  2002-09       Impact factor: 2.689

View more
  2 in total

Review 1.  Which cuff should I use? Indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review.

Authors:  Greg Irving; John Holden; Richard Stevens; Richard J McManus
Journal:  BMJ Open       Date:  2016-11-03       Impact factor: 2.692

2.  Preliminary blood pressure screening in a representative sample of extremely obese Kuwaiti adolescents.

Authors:  Rima Abdul Razzak; Asma Elmteri; Taiba Elkanderi; Fareedah Ishaq; Munera Eljasem; Saja AlDuraie; Jenan Al-Boloushi; Fatma Elbanay; Latifa Eldabos
Journal:  J Obes       Date:  2013-11-18
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.