Literature DB >> 16534409

Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in adults and obese adults.

Sekip Altunkan1, Kamuran Oztaş, Erkan Altunkan.   

Abstract

OBJECTIVE: Wrist blood pressure devices are becoming popular for the self-measurement of blood pressure. Despite widespread use of these devices, there is limited published evidence for their accuracy and reliability. In this study, we assessed the accuracy of the Omron 637IT wrist blood pressure device against the mercury sphygmomanometer in adults and obese adults according to the International Protocol criteria.
METHODS: A hundred and thirty-one participants were studied and classified on the basis of the International Protocol range. Blood pressure measurements at the wrist using the Omron 637IT were compared with the results obtained by two trained observers using a mercury sphygmomanometer. Nine sequential blood pressure measurements were taken. Initially, 33 participants with random arm circumferences were selected for the first validation study. Then, the device was evaluated in adults with arm circumferences equal to or higher than 34 cm. During each validation study, 99 measurements were obtained for comparison from the 33 participants. The first phase was performed on 15 participants, and, if the device passed this phase, 18 more participants were selected.
RESULTS: Mean discrepancies and standard deviations of the sphygmomanometer device were 0.5+/-6.2 mmHg for systolic blood pressure and 0.1+/-3.7 mmHg for diastolic blood pressure in the adult group. The device passed phases 1 and 2 in 33 participants. Mean discrepancies and standard deviations of the sphygmomanometer device were 1.8+/-6.6 mmHg for systolic blood pressure and 1.6+/-4.7 mmHg for diastolic blood pressure in the obese adult group. The device passed phases 1 and 2 in 33 participants in this group.
CONCLUSION: The Omron 637IT passed according to the International Protocol criteria and can be recommended for use in adults and obese adults.

Entities:  

Mesh:

Year:  2006        PMID: 16534409     DOI: 10.1097/01.mbp.0000200483.49540.dc

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  13 in total

1.  Knowledge of blood pressure in a U.K. general public population.

Authors:  J Slark; M S Khan; P Bentley; P Sharma
Journal:  J Hum Hypertens       Date:  2014-01-16       Impact factor: 3.012

2.  Does "hidden undercuffing" occur among obese patients? Effect of arm sizes and other predictors of the difference between wrist and upper arm blood pressures.

Authors:  Hardik Doshi; Alan B Weder; Robert L Bard; Robert D Brook
Journal:  J Clin Hypertens (Greenwich)       Date:  2010-02-01       Impact factor: 3.738

3.  Hypertension among older adults in low- and middle-income countries: prevalence, awareness and control.

Authors:  Peter Lloyd-Sherlock; John Beard; Nadia Minicuci; Shah Ebrahim; Somnath Chatterji
Journal:  Int J Epidemiol       Date:  2014-02-06       Impact factor: 7.196

4.  Psychosocial work factors, blood pressure and psychological strain in male bus operators.

Authors:  Boris Cendales; Sergio Useche; Viviola Gómez
Journal:  Ind Health       Date:  2014-05-28       Impact factor: 2.179

5.  Questionable accuracy of home blood pressure measurements in the obese population - Validation of the Microlife WatchBP O3® and Omron RS6® devices according to the European Society of Hypertension-International Protocol.

Authors:  Alaa Azaki; Reem Diab; Aya Harb; Roland Asmar; Mirna N Chahine
Journal:  Vasc Health Risk Manag       Date:  2017-02-27

6.  Reproducibility of wrist home blood pressure measurement with position sensor and automatic data storage.

Authors:  Sakir Uen; Rolf Fimmers; Miriam Brieger; Georg Nickenig; Thomas Mengden
Journal:  BMC Cardiovasc Disord       Date:  2009-05-27       Impact factor: 2.298

7.  The impact of central blood pressure levels on the relationship between oscillometric and central blood pressure measurements: a multicenter invasive study.

Authors:  Yusuf I Alihanoglu; Mehmet Kayrak; Mehmet S Ulgen; Mehmet Yazici; Mehmet Yazici; Remzi Yilmaz; Kenan Demir; Yildiz Dogan; Murat Sizer; Hakan Ozhan; Fatih Koc; Sait Bodur
Journal:  J Clin Hypertens (Greenwich)       Date:  2013-07-16       Impact factor: 3.738

8.  Socio-demographic inequalities in the prevalence, diagnosis and management of hypertension in India: analysis of nationally-representative survey data.

Authors:  Kath A Moser; Sutapa Agrawal; George Davey Smith; Shah Ebrahim
Journal:  PLoS One       Date:  2014-01-23       Impact factor: 3.240

9.  High out-of-clinic blood pressure is associated with adiposity indicators in leisure physical activity practitioners in Midwest Brazil.

Authors:  Bruna M Giglio; Renata C Fernandes; Ana B Marini; João F Mota; Gustavo D Pimentel
Journal:  Clin Hypertens       Date:  2018-04-23

10.  The QardioArm Blood Pressure App for Self-Measurement in an Obese Population: Validation Study Using the European Society of Hypertension International Protocol Revision 2010.

Authors:  Victoria Mazoteras-Pardo; Ricardo Becerro-De-Bengoa-Vallejo; Marta Elena Losa-Iglesias; Daniel López-López; Patricia Palomo-López; David Rodríguez-Sanz; César Calvo-Lobo
Journal:  JMIR Mhealth Uhealth       Date:  2018-10-25       Impact factor: 4.773

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.